Comparative analysis of clinical outcomes obtained with a new diffractive multifocal toric intraocular lens implanted through two types of corneal incision
Language English Country United States Media print-electronic
Document type Comparative Study, Journal Article, Multicenter Study, Research Support, Non-U.S. Gov't
- MeSH
- Astigmatism physiopathology MeSH
- Adult MeSH
- Phacoemulsification methods MeSH
- Lens Implantation, Intraocular * MeSH
- Middle Aged MeSH
- Humans MeSH
- Microsurgery methods MeSH
- Young Adult MeSH
- Follow-Up Studies MeSH
- Lenses, Intraocular * MeSH
- Prosthesis Design MeSH
- Refraction, Ocular physiology MeSH
- Retrospective Studies MeSH
- Cornea surgery MeSH
- Corneal Topography MeSH
- Treatment Outcome MeSH
- Visual Acuity physiology MeSH
- Check Tag
- Adult MeSH
- Middle Aged MeSH
- Humans MeSH
- Young Adult MeSH
- Male MeSH
- Female MeSH
- Publication type
- Journal Article MeSH
- Multicenter Study MeSH
- Research Support, Non-U.S. Gov't MeSH
- Comparative Study MeSH
PURPOSE: To analyze and compare the clinical outcomes obtained after cataract surgery with the implantation of a new multifocal toric intraocular lens (IOL) using two different types of corneal incision. METHODS: Retrospective study including 64 eyes of 35 patients who underwent surgery with implantation of the AT LISA 909M multifocal toric IOL (Carl Zeiss Meditec) in 2 different ophthalmologic centers using different criteria for corneal incision size: sub-1.8 mm (micro-incision surgery [MICS] group) and 2.2 mm (mini-incision group). Visual, refractive, and corneal topographic outcomes were evaluated during 6-month follow-up. Additionally, refractive astigmatic changes were analyzed using the Alpins vectorial method. RESULTS: Significant reductions of refractive sphere and cylinder were observed postoperatively (P<.03), with associated visual improvements for near and distance (P<.01) in both groups. Mean postoperative magnitudes of difference vector, torque, and magnitude of error in the overall sample were 0.52±0.36 diopters (D), 0.36 ± 0.36 D, and 0.08 ± 0.38 D, respectively. A mean overcorrection of 4% in refractive astigmatism was found. Mean angle of error was 0.37± 5.50° and -4.51±13.16° for the MICS and mini-incision groups, respectively (P=.09). Significant positive correlations were found between the magnitudes of torque and difference vector (r=0.78, P<.01) as well as between the magnitude of torque and absolute angle of error (r=0.76, P<.01). CONCLUSIONS: Implantation of the AT LISA toric IOL using corneal incisions <2.2 mm provides excellent predictability for astigmatic correction.