The insufficiency of left anterior oblique and the usefulness of right anterior oblique projection for correct localization of a computed tomography-verified right ventricular lead into the midseptum
Jazyk angličtina Země Spojené státy americké Médium print-electronic
Typ dokumentu časopisecké články, práce podpořená grantem
PubMed
23742805
DOI
10.1161/circep.113.000232
PII: CIRCEP.113.000232
Knihovny.cz E-zdroje
- Klíčová slova
- cardiac pacing, fluoroscopy, pacemaker implantation, right ventricular apex, ventricular septum,
- MeSH
- design vybavení MeSH
- elektrokardiografie MeSH
- fluoroskopie MeSH
- kardiostimulace umělá * MeSH
- lidé MeSH
- logistické modely MeSH
- mezikomorová přepážka diagnostické zobrazování MeSH
- multidetektorová počítačová tomografie * MeSH
- multivariační analýza MeSH
- prediktivní hodnota testů MeSH
- prospektivní studie MeSH
- rentgenový obraz - interpretace počítačová * MeSH
- rozdělení chí kvadrát MeSH
- senioři nad 80 let MeSH
- senioři MeSH
- srdeční komory diagnostické zobrazování MeSH
- Check Tag
- lidé MeSH
- mužské pohlaví MeSH
- senioři nad 80 let MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
BACKGROUND: The aim of the study was to verify the correct anchoring location for the tip of the right ventricular lead using cardiac computed tomography and to assess the best fluoroscopic and ECG criteria associated with the correct location of the electrode into the midseptum. METHODS AND RESULTS: Patients indicated to pacemaker implantation were prospectively enrolled. The right ventricular lead was implanted into the midseptum according to standard criteria in left anterior oblique 40 view. The cardiac shadow on the right anterior oblique 30 was divided into 4 quadrants perpendicular to the lateral cardiac silhouette and the position of the lead tip was analyzed. The exact position of the lead tip was assessed using computed tomography. Of 51 patients, the right ventricular lead was anchored midseptum in 21 (41.2%; MS group). In 30 patients (58.8%; non-MS group), the lead was anchored in the adjacent anterior wall. The angle between the lead and horizontal axis on the left anterior oblique was similar in both groups. The non-MS group was associated with shorter distances between the tip and the cardiac contours in the right anterior oblique 30 (96.7% of leads in the non-MS group were in the outer quadrant versus 9.6% in the MS group; P<0.001). The presence of the lead in the middle or inferior quadrants was independently associated with correct midseptum placement with positive predictive value of 94.7%. CONCLUSIONS: Despite the optimal shape of the left anterior oblique, substantial numbers of leads were not anchored in the midseptum. Knowing the right anterior oblique 30 lead position can ensure proper midseptal placement.
Citace poskytuje Crossref.org