• This record comes from PubMed

Phone camera detection of glucose blood level based on magnetic particles entrapped inside bubble wrap

. 2016 Dec 18 ; 37 (Suppl1) : 132-138.

Language English Country Sweden Media print

Document type Journal Article

Links

PubMed 28263541
PII: NEL370916A16
Knihovny.cz E-resources

OBJECTIVES: Glucose is an important diagnostic biochemical marker of diabetes but also for organophosphates, carbamates, acetaminophens or salicylates poisoning. Hence, innovation of accurate and fast detection assay is still one of priorities in biomedical research. METHODS: Glucose sensor based on magnetic particles (MPs) with immobilized enzymes glucose oxidase (GOx) and horseradish peroxidase (HRP) was developed and the GOx catalyzed reaction was visualized by a smart-phone-integrated camera. RESULTS: Exponential decay concentration curve with correlation coefficient 0.997 and with limit of detection 0.4 mmol/l was achieved. Interfering and matrix substances were measured due to possibility of assay influencing and no effect of the tested substances was observed. Spiked plasma samples were also measured and no influence of plasma matrix on the assay was proved. CONCLUSIONS: The presented assay showed complying results with reference method (standard spectrophotometry based on enzymes glucose oxidase and peroxidase inside plastic cuvettes) with linear dependence and correlation coefficient 0.999 in concentration range between 0 and 4 mmol/l. On the grounds of measured results, method was considered as highly specific, accurate and fast assay for detection of glucose.

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...