• This record comes from PubMed

Finite element analysis of 6 large PMMA skull reconstructions: A multi-criteria evaluation approach

. 2017 ; 12 (6) : e0179325. [epub] 20170613

Language English Country United States Media electronic-ecollection

Document type Evaluation Study, Journal Article

In this study 6 pre-operative designs for PMMA based reconstructions of cranial defects were evaluated for their mechanical robustness using finite element modeling. Clinical experience and engineering principles were employed to create multiple plan options, which were subsequently computationally analyzed for mechanically relevant parameters under 50N loads: stress, strain and deformation in various components of the assembly. The factors assessed were: defect size, location and shape. The major variable in the cranioplasty assembly design was the arrangement of the fixation plates. An additional study variable introduced was the location of the 50N load within the implant area. It was found that in smaller defects, it was simpler to design a symmetric distribution of plates and under limited variability in load location it was possible to design an optimal for expected loads. However, for very large defects with complex shapes, the variability in the load locations introduces complications to the intuitive design of the optimal assembly. The study shows that it can be beneficial to incorporate multi design computational analyses to decide upon the most optimal plan for a clinical case.

See more in PubMed

Farré-Guasch E, Wolff J, Helder MN, Schulten EA, Forouzanfar T, Klein-Nulend J. Application of Additive Manufacturing in Oral and Maxillofacial Surgery. J Oral Maxillofac Surg. 2015;12: 2408–2418 PubMed

Huang GJ, Zhong S, Susarla SM, Swanson EW, Huang J, Gordon CR. Craniofacial reconstruction with poly(methyl methacrylate) customized cranial implants. The Journal of craniofacial surgery. 2015;26(1):64–70. doi: 10.1097/SCS.0000000000001315 PubMed DOI

Elkins CW, Cameron JE. Cranioplasty with acrylic plates. J Neurosurg. 1946;3:199–205. doi: 10.3171/jns.1946.3.3.0199 PubMed DOI

Kim BJ, Hong KS, Park KJ, Park DH, Chung YG, Kang SH. Customized cranioplasty implants using three-dimensional printers and polymethyl-methacrylate casting. J Korean Neurosurg Soc. 2012;52(6):541–546. doi: 10.3340/jkns.2012.52.6.541 PubMed DOI PMC

Spanio S, Ashammakhi N, Ilomaki J, Valimaa T, Rainer C, Waris T, et al. Use of new bioabsorbable tacks and a tackshooter in cranial bone osteofixation saves operative time. The Journal of craniofacial surgery. 2002;13(5):693–696. PubMed

Ridwan-Pramana A, Wolff J, Raziei A, Ashton-James CE, Forouzanfar T. Porous polyethylene implants in facial reconstruction: Outcome and complications. J Craniomaxillofac Surg. 2015;43(8): 1330–1334. doi: 10.1016/j.jcms.2015.06.022 PubMed DOI

Jung S, Cho W, Paik J, Yang S. Long-term surgical outcomes of porous polyethylene orbital implants: a review of 314 cases. Br J Ophthalmol. 2012;96(4): 494–498. doi: 10.1136/bjophthalmol-2011-300132 PubMed DOI PMC

Lienau J, Schell H, Duda GN, Seebeck P, Muchow S, Bail HJ. Initial vascularization and tissue differentiation are influenced by fixation stability. J Orthop Res. 2005;23(3): 639–645. doi: 10.1016/j.orthres.2004.09.006 PubMed DOI

Ridwan-Pramana A, Marcian P, Borak L, Narra N, Forouzanfar T, Wolff J. Structural and mechanical implications of PMMA implant shape and interface geometry in cranioplasty—A finite element study. Journal of cranio-maxillo-facial surgery: official publication of the European Association for Cranio-Maxillo-Facial Surgery. 2016;44(1):34–44. PubMed

El Halabi F, Rodriguez JF, Rebolledo L, Hurtos E, Doblare M. Mechanical characterization and numerical simulation of polyether-ether-ketone (PEEK) cranial implants. J Mech Behav Biomed Mater. 2011;4(8): 1819–1832. doi: 10.1016/j.jmbbm.2011.05.039 PubMed DOI

Marcian P, Konecný O, Borak L, Valasek J, Rehak K, Krpalek D, Florian Z. On the Level of Computational Models in Biomechanics Depending on Gained Data from Ct/Mri and Micro- Ct. MENDEL. 2011;1, 455–462.

Marcian P, Borak L, Valasek J, Kaiser J, Florian Z, Wolff J. Finite element analysis of dental implant loading on atrophic and non-atrophic cancellous and cortical mandibular bone—a feasibility study. J Biomech. 2014;47(16): 3830–3836. doi: 10.1016/j.jbiomech.2014.10.019 PubMed DOI

Yoganandan N, Pintar FA, Zhang J, Baisden JL. Physical properties of the human head: mass, center of gravity and moment of inertia. J Biomech. 2009;42(9): 1177–1192. doi: 10.1016/j.jbiomech.2009.03.029 PubMed DOI

Wang M, Qu X, Cao M, Wang D, Zhang C. Biomechanical three-dimensional finite element analysis of prostheses retained with/without zygoma implants in maxillectomy patients. J Biomech. 2013;46(6): 1155–1561. doi: 10.1016/j.jbiomech.2013.01.004 PubMed DOI

Freedman M, Ring M, Stassen LF. Effect of alveolar bone support on zygomatic implants: a finite element analysis study. Int J Oral Maxillofac Surg. 2013;42(5), 671–676. doi: 10.1016/j.ijom.2012.12.006 PubMed DOI

Motherway JA, Verschueren P, Van der Perre G, Vander Sloten J, Gilchrist MD. The mechanical properties of cranial bone: the effect of loading rate and cranial sampling position. J Biomech. 2009;42(13): 2129–2135. doi: 10.1016/j.jbiomech.2009.05.030 PubMed DOI

Preusser T, Rumpf M, Sauter S, Schwen LO, 3D Composite Finite Elements for Elliptic Boundary Value Problems with Discontinuous Coefficients. SIAM J Sci Comput. 2011;33: 2115–2143.

Gómez FJ, Elices M, Fracture of components with V-shaped notches. Eng Frac Mechanics. 2003;70: 1913–1927.

Pellier J, Geringer J, Forest B. Fretting-corrosion between 316L SS and PMMA: Influence of ionic strength, protein and electrochemical conditions on material wear, Application to orthopaedic implants. Wear. 2011;271: 1563–1571.

Kolios A, Mytilinou V, Lozano-Minguez E, Salonitis K. A Comparative Study of Multiple-Criteria Decision-Making Methods under Stochastic Inputs. Energies. 2016;9(7), 566.

Frost HM. A 2003 update of bone physiology and Wolff's Law for clinicians. Angle Orthod. 2004;74(1): 3–15. doi: 10.1043/0003-3219(2004)074<0003:AUOBPA>2.0.CO;2 PubMed DOI

Tsouknidas A, Maropoulos S, Savvakis S, Michailidis N. FEM assisted evaluation of PMMA and Ti6Al4V as materials for cranioplasty resulting mechanical behaviour and the neurocranial protection. Biomed Mater Eng. 2011;21(3): 139–147. doi: 10.3233/BME-2011-0663 PubMed DOI