High-Resolution Mapping of Ventricular Scar: Evaluation of a Novel Integrated Multielectrode Mapping and Ablation Catheter
Jazyk angličtina Země Spojené státy americké Médium print-electronic
Typ dokumentu srovnávací studie, časopisecké články, Research Support, N.I.H., Extramural, práce podpořená grantem
Grantová podpora
T32 HL007374
NHLBI NIH HHS - United States
R21 HL127650
NHLBI NIH HHS - United States
PubMed
29759516
DOI
10.1016/j.jacep.2016.12.016
PII: S2405-500X(17)30029-4
Knihovny.cz E-zdroje
- Klíčová slova
- electrodes, mapping, microelectrodes, post-infarct, ventricular tachycardia,
- MeSH
- elektrofyziologické techniky kardiologické přístrojové vybavení metody MeSH
- elektrokardiografie MeSH
- endokard patologie patofyziologie MeSH
- infarkt myokardu patofyziologie MeSH
- jizva patologie patofyziologie MeSH
- katetrizační ablace přístrojové vybavení MeSH
- komorová tachykardie patofyziologie MeSH
- magnetická rezonanční spektroskopie metody MeSH
- mapování potenciálů tělesného povrchu přístrojové vybavení MeSH
- mikroelektrody škodlivé účinky normy MeSH
- myokard patologie MeSH
- prasata MeSH
- prospektivní studie MeSH
- srdeční komory diagnostické zobrazování patologie patofyziologie MeSH
- zvířata MeSH
- Check Tag
- mužské pohlaví MeSH
- zvířata MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
- Research Support, N.I.H., Extramural MeSH
- srovnávací studie MeSH
OBJECTIVES: This study sought to evaluate an investigational catheter that incorporates 3 microelectrodes embedded along the circumference of a standard 3.5-mm open-irrigated catheter. BACKGROUND: Mapping resolution is influenced by both electrode size and interelectrode spacing. Multielectrode mapping catheters enhance mapping resolution within scar compared with standard ablation catheters; however, this requires the use of 2 separate catheters for mapping and ablation. METHODS: Six swine with healed infarction and 2 healthy controls underwent mapping of the left ventricle using a THERMOCOOL SMARTTOUCH SF catheter with 3 additional microelectrodes (0.167 mm2) along its circumference (Qdot, Biosense Webster, Diamond Bar, California). Mapping resolution in healthy and scarred tissue was compared between the standard electrodes and microelectrodes using electrogram characteristics, cardiac magnetic resonance, and histology. RESULTS: In healthy myocardium, bipolar voltage amplitude was similar between the standard electrodes and microelectrodes, with a fifth percentile of 1.19 and 1.30 mV, respectively. In healed infarction, the area of low bipolar voltage (defined as <1.5 mV) was smaller with microelectrodes (16.8 cm2 vs. 25.3 cm2; p = 0.033). Specifically, the microelectrodes detected zones of increased bipolar voltage amplitude, with normal electrogram characteristics occurring at the end of or after the QRS, consistent with channels of preserved subendocardium. Identification of surviving subendocardium by the microelectrodes was consistent with cardiac magnetic resonance and histology. The microelectrodes also improved distinction between near-field and far-field electrograms, with more precise identification of scar border zones. CONCLUSIONS: This novel catheter combines high-resolution mapping and radiofrequency ablation with an open-irrigated, tissue contact-sensing technology. It improves scar mapping resolution while limiting the need for and cost associated with the use of a separate mapping catheter.