THE USE OF THE EXRADIN W1 PLASTIC SCINTILLATOR FOR MEASUREMENTS IN EXTERNAL RADIOTHERAPY

. 2019 Dec 31 ; 186 (2-3) : 351-356.

Jazyk angličtina Země Anglie, Velká Británie Médium print

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid31769483

The Exradin W1 plastic scintillator (Standard Imaging) was examined for measurement suitability in high-energy photon beams, high-energy electron beams and tomotherapy in terms of dose to water for reference fields. For photon beams, pulse repetition rate dependence, calibration stability, noise from the photodiode enclosure, detector motion during treatment, output factors, off-axis doses and percentage depth doses were tested. For electron and tomotherapy beams, energy non-dependence was verified. All features make the detector suitable in small and non-standards fields, and for electron beams.