The Effect of Chlorhexidine Disinfectant Gels with Anti-Discoloration Systems on Color and Mechanical Properties of PMMA Resin for Dental Applications

. 2021 May 29 ; 13 (11) : . [epub] 20210529

Status PubMed-not-MEDLINE Language English Country Switzerland Media electronic

Document type Journal Article

Chlorhexidine (CHX)-based dental hygiene products are widely used by dental patients. As these products may have long-term contact with denture poly(methyl methacrylate) (PMMA) resin, anti-discoloration systems (ADSs) were included in them to prevent discoloration of the natural teeth and dental materials. Purpose: The aim of this study was to evaluate the effect of two newly designed CHX-containing gels with ADSs and two commercial products with ADSs (Curasept 0.5% and Curasept 1%) in preventing staining and to analyze the mechanical properties of heat-curing PMMA denture base resin. Materials and methods: Twenty-five discs (five for each test group) of PMMA dental resin with a thickness of 1 mm and a diameter of 20 mm were polymerized according to the manufacturer's instructions and stored in distillate water at a temperature of 37 °C. The surface of the specimens was covered with two commercially available gels-Curasept 1% and Curasept 0.5%, or two experimental gel formulations containing 1% CHX. PMMA specimens stored in distilled water were used as control. The initial values of color and Brinell hardness of the specimens were measured immediately after specimen preparation. The changes in color and Brinell hardness, as well as water sorption, and solubility of the specimens were measured after one year of conditioning. Statistical analysis of the obtained data was performed using one-way analysis of variance and Dunn-Bonferroni post hoc tests. Results: In the group of specimens covered with gel 1 with citric acid or Curasept 0.5%, the color change was clinically acceptable (ΔE* < 2.7). In the specimens stored in contact with gel 2 with polyvinylpyrrolidone (PVP) and Curasept 1%, the ΔE* values were 3.6 and 3.67, respectively. In the control group, the level of hardness decreased significantly from 150 to 140 during the experiment. In addition, a statistically significant decrease in hardness was observed in specimens stored with Curasept 1% and gel 2 with PVP. Specimens stored in contact with Curasept 0.5% and gel 1 with citric acid also showed a lower hardness, but the change was not statistically significant. The sorption of all the groups of PMMA specimens ranged from 22.83 to 24.47 µg/mm3, with no significant differences found between them. All the PMMA specimens stored in contact with the tested CHX gels exhibited a significantly higher solubility (6.84 ± 7.91 µg/mm3) compared to the control group (6.74 µg/mm3), with the highest solubility noted for specimens stored with Curasept 1%. Conclusions: The results showed that CHX used in the gel form with ADSs at a concentration of 0.5% and the experimental gel containing 1% CHX with citric acid caused limited changes to the color and mechanical properties of the PMMA denture base resin during the study period. These gels may be safely used by dental patients for oral hygiene regimen even for prolonged periods of time. ADSs contained in these gels seem to be effective in preventing CHX discoloration.

See more in PubMed

Bogucki Z.A. Clinical aspects of the use of dental adhesive materials in patients with chronic xerostomia. Gerontology. 2013;30:162–166. doi: 10.1111/j.1741-2358.2012.00659.x. PubMed DOI

Sharif-Abdullah S.S.B., Chong M.C., Surdinar-Kaur S.S., Kamaruzzaman S.B., Ng K.H. The effect of chlorhexidine reducing oral colonisation in geriatric patients: A randomised controlled trial. Singap. Med. J. 2016;57:262–266. doi: 10.11622/smedj.2016091. PubMed DOI PMC

Vergara-Buenaventura A., Castro-Ruiz C. Use of mouthwashes against COVID-19 in dentistry. Br. J. Oral Maxillofac. Surg. 2020;58:924–927. doi: 10.1016/j.bjoms.2020.08.016. PubMed DOI PMC

Li Y., Ren B., Peng X., Hu T., Li J., Gong T., Tang B., Xu X., Zhou X. Saliva is a non-negligible factor in the spread of COVID-19. Mol. Oral Microbiol. 2020;35:141–145. doi: 10.1111/omi.12289. PubMed DOI PMC

Badran Z., Gaudin A., Struillou X., Amador G., Soueidan A. Periodontal pockets: A potential reservoir for SARS-CoV-2? Med. Hypotheses. 2020;143:109907. doi: 10.1016/j.mehy.2020.109907. PubMed DOI PMC

Perihan O., Gulsen C. The Effects of disinfectant solutions on the viscoelastic properties of acrylic resins. Cumhur. Dent. J. 2014;17:205–214.

Moffa E.B., Giampaolo E.T., Izumida F.E., Pavarina A.C., Machado A.L., Vergani C.E. Colour stability of relined dentures after chemical disinfection. A randomised clinical trial. J. Dent. 2011;39(Suppl. S3):65–71. doi: 10.1016/j.jdent.2011.10.008. PubMed DOI

Drăghici R.A.L.U.C.A., Preoteasa C.T., Popa L.Ă., Murariu-Măgureanu C.Ă., Tancu A.M., Meghea D., Cristea A.D., Preoteasa E. In vitro spectrophotometric evaluation of acrylic teeth staining related to dietary and oral antiseptic agents. FARM. 2017;65:630–634.

Sipahi C., Beyzadeoglu M., Demirtas S., Ozen J. Effect of different mucosal and acrylic resin surface treatments in a denture retention model for patients with radiotherapy-induced xerostomia. Int. J. Prosthodont. 2007;20:405–408. PubMed

Zafar M.S. Prosthodontic Applications of Polymethyl Methacrylate (PMMA): An Update. Polymers. 2020;12:2299. doi: 10.3390/polym12102299. PubMed DOI PMC

Campaner L.M., Silveira M.P.M., de Andrade G.S., Borges A.L.S., Bottino M.A., Dal Piva A.M.d.O., Lo Giudice R., Ausiello P., Tribst J.P.M. Influence of Polymeric Restorative Materials on the Stress Distribution in Posterior Fixed Partial Dentures: 3D Finite Element Analysis. Polymers. 2021;13:758. doi: 10.3390/polym13050758. PubMed DOI PMC

Raszewski Z., Nowakowska-Toporowska A., Nowakowska D., Więckiewicz W. Update on acrylic resins used in dentistry. Mini Rev. Med. Chem. 2021 doi: 10.2174/1389557521666210226151214. Epub ahead of print. PubMed DOI

Karpiński T.M., Szkaradkiewicz A.K. Chlorhexidine-pharmaco-biological activity and application. Eur. Rev. Med. Pharmacol. Sci. 2015;19:1321–1326. PubMed

Brookes Z.L., Bescos R., Belfield L.A., Ali K., Roberts A. Current uses of chlorhexidine for management of oral disease: A narrative review. J. Dent. 2020;103:103497. doi: 10.1016/j.jdent.2020.103497. PubMed DOI PMC

Opstrup M.S., Jemec G.B.E., Garvey L.H. Chlorhexidine Allergy on the Rise and often Overlooked. Curr. Allergy Asthma Rep. 2019;14:23. doi: 10.1007/s11882-019-0858-2. PubMed DOI

Guerra F., Pasqualotto D., Rinaldo F., Mazur M., Corridore D., Nofroni I., Ottolenghi L., Nardi G.M. Therapeutic efficacy of chlorhexidine-based mouthwashes and its adverse events: Performance-related evaluation of mouthwashes added with Anti-Discoloration System and cetylpyridinium chloride. Int. J. Dent. Hyg. 2019;17:229–236. doi: 10.1111/idh.12371. PubMed DOI

Jones C.G. Chlorhexidine: Is it still the gold standard? Periodontology. 1997;15:55–62. doi: 10.1111/j.1600-0757.1997.tb00105.x. PubMed DOI

Patel A., Sethuraman R., Prajapati P., Patel J., Naveen Y.G. A comparative analysis of staining characteristics of mouthrinses on provisional acrylic resin: An in vitro study. J. Interdiscip. Dent. 2013;3:167–173.

Pacheco F.G. Ph.D. Thesis. Faculdade de Medicina Dentária, Universidade de Lisboa; Lisbon, Portugal: 2016. Insights on the Thermal Aging of Acrylic Reline Resins.

Silva P.M.B.D., Acosta E.J.T.R., Jacobina M., Pinto L.D.R., Porto V.C. Effect of repeated immersion solution cycles on the color stability of denture tooth acrylic resins. J. Appl. Oral Sci. 2011;19:623–627. doi: 10.1590/S1678-77572011000600013. PubMed DOI PMC

Goiato M.C., Dos Santos D.M., Baptista G.T., Moreno A., Andreotti A.M., Bannwart L.C., Dekon S.F. Effect of thermal cycling and disinfection on color stability of denture base acrylic resin. Gerodontology. 2013;30:276–282. doi: 10.1111/j.1741-2358.2012.00676.x. PubMed DOI

Altıncı P., Durkaya P. Effects of thermocycling and various drinks on the color stability of heat-polymerized acrylic resin. J. Istanb. Univ. Fac. Dent. 2016;50:15–20. doi: 10.17096/jiufd.28587. PubMed DOI PMC

Piskin B., Sipahi C., Akin H. Effect of different chemical disinfectants on color stability of acrylic denture teeth. J. Prosthodont. 2014;23:476–483. doi: 10.1111/jopr.12131. PubMed DOI

Baig A.R., Shori D.D., Shenoi P.R., Ali S.N., Shetti S., Godhane A. Mouthrinses affect color stability of composite. J. Conserv. Dent. 2016;19:355–359. doi: 10.4103/0972-0707.186448. PubMed DOI PMC

Gregorius W.C., Kattadiyil M.T., Goodacre C.J., Roggenkamp C.L., Powers J.M., Paravina R.D. Effects of ageing and staining on color of acrylic resin denture teeth. J. Color. Appear. Dent. 2012;40:47–54. doi: 10.1016/j.jdent.2012.09.009. PubMed DOI

Lai Y.L., Lui H.F., Lee S.Y. In vitro color stability, stain resistance, and water sorption of four removable gingival flange materials. J. Prosthet. Dent. 2003;90:293–300. doi: 10.1016/S0022-3913(03)00432-3. PubMed DOI

Neppelenbroek K.H., Kuroishi E., Hotta J., Marques V.R., Moffa E.B., Soares S., Urban V.M. Surface properties of multilayered, acrylic resin artificial teeth after immersion in staining beverages. J. Appl. Oral Sci. 2015;23:376–382. doi: 10.1590/1678-775720150054. PubMed DOI PMC

Goiato M.C., Dos Santos D.M., Baptista G.T., Moreno A., Andreotti A.M., Dekon S.F.D.C. Effect of thermal cycling and disinfection on microhardness of acrylic resin denture base. J. Med. Eng. Technol. 2013;37:203–207. doi: 10.3109/03091902.2013.774444. PubMed DOI

Campanha N.H., Pavarina A.C., Jorge J.H., Vergani C.E., Machado A.L., Giampaolo E.T. The effect of long-term disinfection procedures on hardness property of resin denture teeth. Gerodontology. 2012;29:571–576. doi: 10.1111/j.1741-2358.2011.00520.x. PubMed DOI

Raszewski Z., Nowakowska-Toporowska A., Weżgowiec J., Nowakowska D. Design and characteristics of new experimental chlorhexidine dental gels with anti-staining properties. Adv. Clin. Exp. Med. 2019;28:885–890. doi: 10.17219/acem/94152. PubMed DOI

Marrelli M., Amantea M., Tatullo M. A comparative, randomized, controlled study on clinical efficacy and dental staining reduction of a mouthwash containing Chlorhexidine 0.20% and Anti Discoloration System (ADS) Ann. Stomatol. 2015;6:35–42. PubMed PMC

Van Swaaij B.W., Van der Weijden G.A., Bakker E.W., Graziani F., Slot D.E. Does chlorhexidine mouthwash, with an anti-discoloration system, reduce tooth surface discoloration without losing its efficacy? A systematic review and meta-analysis. Int. J. Dent. Hyg. 2020;18:27–43. doi: 10.1111/idh.12402. PubMed DOI PMC

Bernardi F., Pincelli M.R., Carloni S., Gatto M.R., Montebugnoli L. Chlorhexidine with an Anti Discoloration System. A comparative study. Int. J. Dent. Hyg. 2004;2:122–126. doi: 10.1111/j.1601-5037.2004.00083.x. PubMed DOI

Bevilacqua L., Liani G., Castronovo G., Costantinides F. Clinical and spectrophotometric evaluation after chlorhexidine use in periodontal flap surgery: A prospective randomized clinical trial. Am. J. Dent. 2016;29:75–80. PubMed

Kuehni R.G., Marcus M. An experiment in visual scaling of small color differences. Color Res. Appl. 1979;4:83–91.

Imirzalioglu P., Karacaer O., Yilmaz B., Ozmen MSc I. Color stability of denture acrylic resins and a soft lining material against tea, coffee, and nicotine. J. Prosthodont. 2010;19:118–124. doi: 10.1111/j.1532-849X.2009.00535.x. PubMed DOI

ISO . ISO; Geneva, Switzerland: 2013. 20795-1:2013 Dentistry—Base polymers—Part 1: Denture Base Polymers.

Paravina R.D., Ghinea R., Herrera L.J., Bona A.D., Igiel C., Linninger M., Sakai M., Takahashi H., Tashkandi E., Mar Perez M.D. Color difference thresholds in dentistry. J. Esthet. Restor. Dent. 2015;27(Suppl. S1):S1–S9. doi: 10.1111/jerd.12149. PubMed DOI

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...