Nejvíce citovaný článek - PubMed ID 24670710
Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry
AIMS: To report 5-year outcomes of EFFORTLESS registry patients with early generation subcutaneous implantable cardioverter-defibrillator (S-ICD) devices. METHODS AND RESULTS: Kaplan-Meier, trend and multivariable analyses were performed for mortality and late (years 2-5) complications, appropriate shock (AS) and inappropriate shock (IAS) rates. Nine hundred and eighty-four of 994 enrolled patients with diverse diagnoses (28% female, 48 ± 17 years, body mass index 27 ± 6 kg/m2, ejection fraction 43 ± 18%) underwent S-ICD implantation. Median follow-up was 5.1 years (interquartile range 4.7-5.5 years). All-cause mortality was 9.3% (95% confidence interval 7.2-11.3%) at 5 years; 703 patients remained in follow-up on study completion, 171 withdrew including 87 (8.8%) with device explanted, and 65 (6.6%) lost to follow-up. Of the explants, only 20 (2.0%) patients needed a transvenous device for pacing indications. First and final shock efficacy for discrete ventricular arrhythmias was consistent at 90% and 98%, respectively, with storm episode final shock efficacy at 95.2%. Time to therapy remained unaltered. Overall 1- and 5-year complication rates were 8.9% and 15.2%, respectively. Early complications did not predict later complications. There were no structural lead failures. Inappropriate shock rates at 1 and 5 years were 8.7% and 16.9%, respectively. Self-terminating inappropriately sensed episodes predicted late IAS. Predictors of late AS included self-terminating appropriately sensed episodes and earlier AS. CONCLUSION: In this diverse S-ICD registry population, spontaneous shock efficacy was consistently high over 5 years. Very few patients underwent S-ICD replacement with a transvenous device for pacing indications. Treated and self-terminating arrhythmic episodes predict future shock events, which should encourage more personalized device optimization.
- Klíčová slova
- Implantable cardioverter-defibrillator, Primary prevention, Secondary prevention, Subcutaneous ICD, Sudden death,
- MeSH
- defibrilátory implantabilní * škodlivé účinky MeSH
- lidé MeSH
- náhlá srdeční smrt epidemiologie etiologie prevence a kontrola MeSH
- následné studie MeSH
- srdeční arytmie MeSH
- tepový objem MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- mužské pohlaví MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
AIMS: The subcutaneous implantable cardioverter defibrillator (S-ICD) was introduced to overcome complications related to transvenous leads. Adoption of the S-ICD requires implanters to learn a new implantation technique. The aim of this study was to assess the learning curve for S-ICD implanters with respect to implant-related complications, procedure time, and inappropriate shocks (IASs). METHODS AND RESULTS: In a pooled cohort from two clinical S-ICD databases, the IDE Trial and the EFFORTLESS Registry, complications, IASs at 180 days follow-up and implant procedure duration were assessed. Patients were grouped in quartiles based on experience of the implanter and Kaplan-Meier estimates of complication and IAS rates were calculated. A total of 882 patients implanted in 61 centres by 107 implanters with a median of 4 implants (IQR 1,8) were analysed. There were a total of 59 patients with complications and 48 patients with IAS. The complication rate decreased significantly from 9.8% in Quartile 1 (least experience) to 5.4% in Quartile 4 (most experience) (P = 0.02) and non-significantly for IAS from 7.9 to 4.8% (P = 0.10). Multivariable analysis demonstrated a hazard ratio of 0.78 (P = 0.045) for complications and 1.01 (P = 0.958) for IAS. Dual-zone programming increased with experience of the individual implanter (P < 0.001), which reduced IAS significantly in the multivariable model (HR 0.44, P = 0.01). Procedure time decreased from 75 to 65 min (P < 0.001). The complication rate and procedure time stabilized after Quartile 2 (>13 implants). CONCLUSION: There is a short and significant learning curve associated with physicians adopting the S-ICD. Performance stabilizes after 13 implants.
- Klíčová slova
- Complications, Implantable cardioverter-defibrillator, Inappropriate shock, Learning curve, S-ICD, Subcutaneous implantable cardioverter-defibrillator,
- MeSH
- databáze faktografické MeSH
- defibrilátory implantabilní * MeSH
- dospělí MeSH
- elektrická defibrilace * MeSH
- implantace protézy výchova MeSH
- Kaplanův-Meierův odhad MeSH
- kardiochirurgické výkony výchova MeSH
- křivka učení * MeSH
- lidé středního věku MeSH
- lidé MeSH
- multivariační analýza MeSH
- následné studie MeSH
- proporcionální rizikové modely MeSH
- registrace MeSH
- rizikové faktory MeSH
- senioři MeSH
- výsledek terapie MeSH
- Check Tag
- dospělí MeSH
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- pozorovací studie MeSH