Nejvíce citovaný článek - PubMed ID 35090007
Subcutaneous implantable cardioverter-defibrillators: long-term results of the EFFORTLESS study
BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (S-ICD) has demonstrated safety and efficacy for the treatment of malignant ventricular arrhythmias. However, a limitation of the S-ICD lies in the inability to either pace-terminate ventricular tachycardia or provide prolonged bradycardia pacing support. OBJECTIVE: The rationale and design of a prospective, single-arm, multinational trial of an intercommunicative leadless pacing system integrated with the S-ICD will be presented. METHODS: A technical description of the modular cardiac rhythm management (mCRM) system (EMPOWER leadless pacemaker and EMBLEM S-ICD) and the implantation procedure is provided. MODULAR ATP (Effectiveness of the EMPOWER™ Modular Pacing System and EMBLEM™ Subcutaneous ICD to Communicate Antitachycardia Pacing) is a multicenter, international trial enrolling up to 300 patients at risk of sudden cardiac death at up to 60 centers trial design. The safety endpoint of freedom from major complications related to the mCRM system or implantation procedure at 6 months and 2 years are significantly higher than 86% and 81%, respectively, and all-cause survival is significantly >85% at 2 years. RESULTS: Efficacy endpoints are that at 6 months mCRM communication success is significantly higher than 88% and the percentage of subjects with low and stable thresholds is significantly higher than 80%. Substudies to evaluate rate-responsive features and performance of the pacing module are also described. CONCLUSION: The MODULAR ATP global clinical trial will prospectively test the safety and efficacy of the first intercommunicating leadless pacing system with the S-ICD. This trial will allow for robust validation of device-device communication, pacing performance, rate responsiveness, and system safety.
- Klíčová slova
- Antitachycardia pacing, Defibrillator, Leadless pacemaker, Subcutaneous ICD, Transcatheter pacemaker,
- Publikační typ
- časopisecké články MeSH
BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (S-ICD) is developed to overcome lead-related complications and systemic infections, inherent to transvenous ICD (TV-ICD) therapy. The PRAETORIAN trial demonstrated that the S-ICD is non-inferior to the TV-ICD with regard to the combined primary endpoint of inappropriate shocks and complications. This prespecified secondary analysis evaluates all complications in the PRAETORIAN trial. METHODS AND RESULTS: The PRAETORIAN trial is an international, multicentre, randomized trial in which 849 patients with an indication for ICD therapy were randomized to receive an S- ICD (N = 426) or TV-ICD (N = 423) and followed for a median of 49 months. Endpoints were device-related complications, lead-related complications, systemic infections, and the need for invasive interventions. Thirty-six device-related complications occurred in 31 patients in the S-ICD group of which bleedings were the most frequent. In the TV-ICD group, 49 complications occurred in 44 patients of which lead dysfunction was most frequent (HR: 0.69; P = 0.11). In both groups, half of all complications were within 30 days after implantation. Lead-related complications and systemic infections occurred significantly less in the S-ICD group compared with the TV-ICD group (P < 0.001, P = 0.03, respectively). Significantly more complications required invasive interventions in the TV-ICD group compared with the S-ICD group (8.3% vs. 4.3%, HR: 0.59; P = 0.047). CONCLUSION: This secondary analysis shows that lead-related complications and systemic infections are more prevalent in the TV-ICD group compared with the S-ICD group. In addition, complications in the TV-ICD group were more severe as they required significantly more invasive interventions. This data contributes to shared decision-making in clinical practice.
- Klíčová slova
- Complications, Infections, Invasive interventions, Lead-related complications, Subcutaneous ICD, Transvenous ICD,
- MeSH
- defibrilátory implantabilní * škodlivé účinky MeSH
- lidé MeSH
- náhlá srdeční smrt * MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- práce podpořená grantem MeSH
- randomizované kontrolované studie MeSH