Electrical Impedance Tomography Can Be Used to Quantify Lung Hyperinflation during HFOV: The Pilot Study in Pigs

V. Ort, K. Roubik

. 2022 ; 12 (9) : . [pub] 20220828

Jazyk angličtina Země Švýcarsko

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc22023425

Grantová podpora
SGS19/203/OHK4/3T/17 Grant Agency of the Czech Technical University in Prague
SGS20/202/OHK4/3T/17 Grant Agency of the Czech Technical University in Prague

Dynamic hyperinflation is reported as a potential risk during high-frequency oscillatory ventilation (HFOV), and its existence has been documented both by physical models and by CT. The aim of this study is to determine the suitability of electrical impendence tomography (EIT) for the measurement of dynamic lung hyperinflation and hypoinflation during HFOV. Eleven healthy pigs were anaesthetized and ventilated using HFOV. The difference between the airway pressure at the airway opening and alveolar space was measured by EIT and esophageal balloons at three mean airway pressures (12, 18 and 24 cm H2O) and two inspiratory to expiratory time ratios (1:1, 1:2). The I:E ratio was the primary parameter associated with differences between airway and alveolar pressures. All animals showed hyperinflation at a 1:1 ratio (median 1.9 cm H2O) and hypoinflation at a 1:2 (median -4.0 cm H2O) as measured by EIT. EIT measurements had a linear correlation to esophageal balloon measurements (r2 = -0.915, p = 0.0085). EIT measurements were slightly higher than that of the esophageal balloon transducer with the mean difference of 0.57 cm H2O. Presence of a hyperinflation or hypoinflation was also confirmed independently by chest X-ray. We found that dynamic hyperinflation developed during HFOV may be detected and characterized noninvasively by EIT.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc22023425
003      
CZ-PrNML
005      
20221031095226.0
007      
ta
008      
221010s2022 sz f 000 0|eng||
009      
AR
024    7_
$a 10.3390/diagnostics12092081 $2 doi
035    __
$a (PubMed)36140483
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a sz
100    1_
$a Ort, Vaclav $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, nam. Sitna 3105, 272 01 Kladno, Czech Republic $1 0000000316772446
245    10
$a Electrical Impedance Tomography Can Be Used to Quantify Lung Hyperinflation during HFOV: The Pilot Study in Pigs / $c V. Ort, K. Roubik
520    9_
$a Dynamic hyperinflation is reported as a potential risk during high-frequency oscillatory ventilation (HFOV), and its existence has been documented both by physical models and by CT. The aim of this study is to determine the suitability of electrical impendence tomography (EIT) for the measurement of dynamic lung hyperinflation and hypoinflation during HFOV. Eleven healthy pigs were anaesthetized and ventilated using HFOV. The difference between the airway pressure at the airway opening and alveolar space was measured by EIT and esophageal balloons at three mean airway pressures (12, 18 and 24 cm H2O) and two inspiratory to expiratory time ratios (1:1, 1:2). The I:E ratio was the primary parameter associated with differences between airway and alveolar pressures. All animals showed hyperinflation at a 1:1 ratio (median 1.9 cm H2O) and hypoinflation at a 1:2 (median -4.0 cm H2O) as measured by EIT. EIT measurements had a linear correlation to esophageal balloon measurements (r2 = -0.915, p = 0.0085). EIT measurements were slightly higher than that of the esophageal balloon transducer with the mean difference of 0.57 cm H2O. Presence of a hyperinflation or hypoinflation was also confirmed independently by chest X-ray. We found that dynamic hyperinflation developed during HFOV may be detected and characterized noninvasively by EIT.
655    _2
$a časopisecké články $7 D016428
700    1_
$a Roubik, Karel $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, nam. Sitna 3105, 272 01 Kladno, Czech Republic $1 0000000158136355
773    0_
$w MED00195450 $t Diagnostics (Basel, Switzerland) $x 2075-4418 $g Roč. 12, č. 9 (2022)
856    41
$u https://pubmed.ncbi.nlm.nih.gov/36140483 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y - $z 0
990    __
$a 20221010 $b ABA008
991    __
$a 20221031095224 $b ABA008
999    __
$a ind $b bmc $g 1853863 $s 1174713
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2022 $b 12 $c 9 $e 20220828 $i 2075-4418 $m Diagnostics $n Diagnostics $x MED00195450
GRA    __
$a SGS19/203/OHK4/3T/17 $p Grant Agency of the Czech Technical University in Prague
GRA    __
$a SGS20/202/OHK4/3T/17 $p Grant Agency of the Czech Technical University in Prague
LZP    __
$a Pubmed-20221010

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...