Conductivity detection in capillary zone electrophoresis: inspection by PeakMaster

Jaros M, Soga T, van de Goor T, Gas B.

. 2005 ; 26 (10) : 1948-1953.

Language English Country Germany

E-resources Online

NLK Wiley Online Library (archiv) from 1999-01-01 to 2012-12-31

A simple rule stating that the signal in conductivity detection in capillary zone electrophoresis is proportional to the difference between the analyte mobility and mobility of the background electrolyte (BGE) co-ion is valid only for systems with fully ionized electrolytes. In zone electrophoresis systems with weak electrolytes both conductivity signal and electromigration dispersion of analyte peaks depend on the conductivity and pH effects. This allows optimization of the composition of BGEs to give a good conductivity signal of analytes while still keeping electromigration dispersion near zero, regardless of the injected amount of sample. The demands to achieve minimum electromigration dispersion and high sensitivity in conductivity detection can be accomplished at the same time. PeakMaster software is used for inspection of BGEs commonly used for separation of sugars (carbohydrates, saccharides) at highly alkaline pH. It is shown that the terms direct and indirect conductivity detection are misleading and should not be used.

000      
02422naa 2200361 a 4500
001      
bmc07507033
003      
CZ-PrNML
005      
20111210122803.0
008      
080728s2005 gw e eng||
009      
AR
040    __
$a ABA008 $b cze $c ABA008 $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a gw
100    1_
$a Jaroš, Michal $7 xx0105779
245    10
$a Conductivity detection in capillary zone electrophoresis: inspection by PeakMaster / $c Jaros M, Soga T, van de Goor T, Gas B.
314    __
$a Faculty of Science, Charles University, CZ-128 040 Prague 2, Czech Republic.
520    9_
$a A simple rule stating that the signal in conductivity detection in capillary zone electrophoresis is proportional to the difference between the analyte mobility and mobility of the background electrolyte (BGE) co-ion is valid only for systems with fully ionized electrolytes. In zone electrophoresis systems with weak electrolytes both conductivity signal and electromigration dispersion of analyte peaks depend on the conductivity and pH effects. This allows optimization of the composition of BGEs to give a good conductivity signal of analytes while still keeping electromigration dispersion near zero, regardless of the injected amount of sample. The demands to achieve minimum electromigration dispersion and high sensitivity in conductivity detection can be accomplished at the same time. PeakMaster software is used for inspection of BGEs commonly used for separation of sugars (carbohydrates, saccharides) at highly alkaline pH. It is shown that the terms direct and indirect conductivity detection are misleading and should not be used.
650    _2
$a elektroforéza kapilární $x metody $7 D019075
650    _2
$a elektrická vodivost $7 D004553
650    _2
$a elektrolyty $x chemie $7 D004573
650    _2
$a financování organizované $7 D005381
700    1_
$a Soga, T.
700    1_
$a Goor, T. van de
700    1_
$a Gas, B.
773    0_
$w MED00001508 $t Electrophoresis $g Roč. 26, č. 10 (2005), s. 1948-1953 $x 0173-0835
910    __
$a ABA008 $b x $y 9
990    __
$a 20080721085927 $b ABA008
991    __
$a 20081009154631 $b ABA008
999    __
$a ok $b bmc $g 622641 $s 475074
BAS    __
$a 3
BMC    __
$a 2005 $b 26 $c 10 $d 1948-1953 $i 0173-0835 $m Electrophoresis $x MED00001508
LZP    __
$a 2008-Doreen

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...