Percutaneous transcatheter one-step mechanical aortic disc valve prosthesis implantation: a preliminary feasibility study in swine

Sochman J, Peregrin JH, Rocek M, Timmermans HA, Pavcnik D, Rösch J

. 2006 ; 29 (1) : 114-119.

Jazyk angličtina Země Spojené státy americké

Perzistentní odkaz   https://www.medvik.cz/link/bmc07520267
E-zdroje Online

NLK ProQuest Central od 1997-01-01 do Před 1 rokem
Medline Complete (EBSCOhost) od 2003-02-01 do Před 1 rokem
Nursing & Allied Health Database (ProQuest) od 1997-01-01 do Před 1 rokem
Health & Medicine (ProQuest) od 1997-01-01 do Před 1 rokem

PURPOSE: To evaluate the feasibility of one-step implantation of a new type of stent-based mechanical aortic disc valve prosthesis (MADVP) above and across the native aortic valve and its short-term function in swine with both functional and dysfunctional native valves. METHODS: The MADVP consisted of a folding disc valve made of silicone elastomer attached to either a nitinol Z-stent (Z model) or a nitinol cross-braided stent (SX model). Implantation of 10 MADVPs (6 Z and 4 SX models) was attempted in 10 swine: 4 (2 Z and 2 SX models) with a functional native valve and 6 (4 Z and 2 SX models) with aortic regurgitation induced either by intentional valve injury or by MADVP placement across the native valve. MADVP function was observed for up to 3 hr after implantation. RESULTS: MADVP implantation was successful in 9 swine. One animal died of induced massive regurgitation prior to implantation. Four MADVPs implanted above functioning native valves exhibited good function. In 5 swine with regurgitation, MADVP implantation corrected the induced native valve dysfunction and the device's continuous good function was observed in 4 animals. One MADVP (SX model) placed across native valve gradually migrated into the left ventricle. CONCLUSION: The tested MADVP can be implanted above and across the native valve in a one-step procedure and can replace the function of the regurgitating native valve. Further technical development and testing are warranted, preferably with a manufactured MADVP.

000      
00000naa 2200000 a 4500
001      
bmc07520267
003      
CZ-PrNML
005      
20160715103816.0
008      
090330s2006 xxu e eng||
009      
AR
040    __
$a ABA008 $b cze $c ABA008 $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xxu
100    1_
$a Šochman, Jan, $d 1956- $7 xx0088500
245    10
$a Percutaneous transcatheter one-step mechanical aortic disc valve prosthesis implantation: a preliminary feasibility study in swine / $c Sochman J, Peregrin JH, Rocek M, Timmermans HA, Pavcnik D, Rösch J
314    __
$a Intensive Care Unit, Clinic of Cardiology, Institute for Clinical and Experimental Medicine, Vídenská 1958/9, 140 21 Prague 4, Czech Republic. jan.sochman@medicon.cz
520    9_
$a PURPOSE: To evaluate the feasibility of one-step implantation of a new type of stent-based mechanical aortic disc valve prosthesis (MADVP) above and across the native aortic valve and its short-term function in swine with both functional and dysfunctional native valves. METHODS: The MADVP consisted of a folding disc valve made of silicone elastomer attached to either a nitinol Z-stent (Z model) or a nitinol cross-braided stent (SX model). Implantation of 10 MADVPs (6 Z and 4 SX models) was attempted in 10 swine: 4 (2 Z and 2 SX models) with a functional native valve and 6 (4 Z and 2 SX models) with aortic regurgitation induced either by intentional valve injury or by MADVP placement across the native valve. MADVP function was observed for up to 3 hr after implantation. RESULTS: MADVP implantation was successful in 9 swine. One animal died of induced massive regurgitation prior to implantation. Four MADVPs implanted above functioning native valves exhibited good function. In 5 swine with regurgitation, MADVP implantation corrected the induced native valve dysfunction and the device's continuous good function was observed in 4 animals. One MADVP (SX model) placed across native valve gradually migrated into the left ventricle. CONCLUSION: The tested MADVP can be implanted above and across the native valve in a one-step procedure and can replace the function of the regurgitating native valve. Further technical development and testing are warranted, preferably with a manufactured MADVP.
650    _2
$a zvířata $7 D000818
650    _2
$a aortální insuficience $x chirurgie $x radiografie $7 D001022
650    _2
$a modely nemocí na zvířatech $7 D004195
650    _2
$a studie proveditelnosti $7 D005240
650    _2
$a srdeční chlopně umělé $7 D006350
650    _2
$a chirurgická náhrada chlopně $x metody $7 D019918
650    _2
$a protézy - design $7 D011474
650    _2
$a stenty $7 D015607
650    _2
$a prasata $7 D013552
650    _2
$a financování organizované $7 D005381
700    1_
$a Peregrin, Jan H., $d 1951- $7 mzk2006331741
700    1_
$a Roček, Miloslav, $d 1958- $7 nlk20010095509
700    1_
$a Timmermans, Hans A.
700    1_
$a Pavcnik, Dušan
700    1_
$a Rösch, Josef, $d 1925-2016 $7 nlk19990073810
773    0_
$w MED00001056 $t Cardiovascular and interventional radiology $g Roč. 29, č. 1 (2006), s. 114-119 $x 0174-1551
910    __
$a ABA008 $b x $y 9 $z 0
990    __
$a 20090312170439 $b ABA008
991    __
$a 20160715104021 $b ABA008
999    __
$a ok $b bmc $g 638070 $s 490868
BAS    __
$a 3
BMC    __
$a 2006 $b 29 $c 1 $d 114-119 $i 0174-1551 $m Cardiovascular and interventional radiology $x MED00001056
LZP    __
$a 2009-B1/vtme

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...