Zkoumání tahové síly pedikulárního šroubu
[Investigation of pedicle screw pull-out strength]

Pezowicz C., Filipiak J.

Jazyk čeština Země Česko

Perzistentní odkaz   https://www.medvik.cz/link/bmc10008562

This study was presented problems with usage of transpedicular spine fixators.The purpose of this work was defined the value of pull-out strength of transpedicular screw from vertebrae. Impact of geometrical parameters of transpedicular screw on value of pull-out strength was analysed. Investigation was carried out on cadaver of lumbar and thoracic vertebrae. The results of investigation showed significance of geometrical parameters of transpedicular screws on durability of screw-vertebrae connection. This study proved also that very important for good performance of the vertebrae and pedicle screw unit is adequate match the screw diameter up with pedicle diameter.

Investigation of pedicle screw pull-out strength

Bibliografie atd.

Lit.: 10

000      
00000naa 2200000 a 4500
001      
bmc10008562
003      
CZ-PrNML
005      
20111210162113.0
008      
100413s2003 xr e cze||
009      
AR
040    __
$a ABA008 $b cze $c ABA008 $d ABA008 $e AACR2
041    0_
$a cze $b eng
044    __
$a xr
100    1_
$a Pezowicz, C.
245    10
$a Zkoumání tahové síly pedikulárního šroubu / $c Pezowicz C., Filipiak J.
246    11
$a Investigation of pedicle screw pull-out strength
314    __
$a Division of Biomedical Engineering and Experimental Mechanics, Wroclaw
504    __
$a Lit.: 10
520    9_
$a This study was presented problems with usage of transpedicular spine fixators.The purpose of this work was defined the value of pull-out strength of transpedicular screw from vertebrae. Impact of geometrical parameters of transpedicular screw on value of pull-out strength was analysed. Investigation was carried out on cadaver of lumbar and thoracic vertebrae. The results of investigation showed significance of geometrical parameters of transpedicular screws on durability of screw-vertebrae connection. This study proved also that very important for good performance of the vertebrae and pedicle screw unit is adequate match the screw diameter up with pedicle diameter.
650    _2
$a poranění zad $x chirurgie $x terapie $7 D019567
650    _2
$a ortopedické výkony $x metody $x využití $7 D019637
650    _2
$a ortopedické fixační pomůcky $x využití $7 D009984
650    _2
$a kostní šrouby $x normy $x využití $7 D001863
650    _2
$a biomechanika $7 D001696
650    _2
$a pevnost v tahu $7 D013718
650    _2
$a bederní obratle $x chirurgie $7 D008159
650    _2
$a hrudní obratle $x chirurgie $7 D013904
650    _2
$a radiografie $x metody $x využití $7 D011859
650    _2
$a statistika jako téma $7 D013223
650    _2
$a dospělí $7 D000328
650    _2
$a lidé $7 D006801
650    _2
$a financování organizované $7 D005381
700    1_
$a Filipiak, J.
773    0_
$w MED00011216 $t Pohybové ústrojí $g Roč. 10, č. 1-2 (2003), s. 61-68 $x 1212-4575
856    41
$u http://www.pojivo.cz/pu/PU_12_2003.pdf $y plný text volně přístupný
910    __
$a ABA008 $b A 4059 $c 509 a $y 8
990    __
$a 20100413085328 $b ABA008
991    __
$a 20100813115239 $b ABA008
999    __
$a ok $b bmc $g 722442 $s 585549
BAS    __
$a 3
BMC    __
$a 2003 $b 10 $c 1-2 $d 61-68 $i 1212-4575 $m Pohybové ústrojí. Pokroky ve výzkumu, diagnostice a terapii $x MED00011216
LZP    __
$a 2010-19/vtal