Something wrong with this record ?
Zkoumání tahové síly pedikulárního šroubu
[Investigation of pedicle screw pull-out strength]
Pezowicz C., Filipiak J.
Language Czech Country Czech Republic
- MeSH
- Lumbar Vertebrae surgery MeSH
- Biomechanical Phenomena MeSH
- Adult MeSH
- Financing, Organized MeSH
- Thoracic Vertebrae surgery MeSH
- Bone Screws standards utilization MeSH
- Humans MeSH
- Orthopedic Fixation Devices utilization MeSH
- Orthopedic Procedures methods utilization MeSH
- Tensile Strength MeSH
- Back Injuries surgery therapy MeSH
- Radiography methods utilization MeSH
- Statistics as Topic MeSH
- Check Tag
- Adult MeSH
- Humans MeSH
This study was presented problems with usage of transpedicular spine fixators.The purpose of this work was defined the value of pull-out strength of transpedicular screw from vertebrae. Impact of geometrical parameters of transpedicular screw on value of pull-out strength was analysed. Investigation was carried out on cadaver of lumbar and thoracic vertebrae. The results of investigation showed significance of geometrical parameters of transpedicular screws on durability of screw-vertebrae connection. This study proved also that very important for good performance of the vertebrae and pedicle screw unit is adequate match the screw diameter up with pedicle diameter.
Investigation of pedicle screw pull-out strength
Lit.: 10
- 000
- 00000naa 2200000 a 4500
- 001
- bmc10008562
- 003
- CZ-PrNML
- 005
- 20111210162113.0
- 008
- 100413s2003 xr e cze||
- 009
- AR
- 040 __
- $a ABA008 $b cze $c ABA008 $d ABA008 $e AACR2
- 041 0_
- $a cze $b eng
- 044 __
- $a xr
- 100 1_
- $a Pezowicz, C.
- 245 10
- $a Zkoumání tahové síly pedikulárního šroubu / $c Pezowicz C., Filipiak J.
- 246 11
- $a Investigation of pedicle screw pull-out strength
- 314 __
- $a Division of Biomedical Engineering and Experimental Mechanics, Wroclaw
- 504 __
- $a Lit.: 10
- 520 9_
- $a This study was presented problems with usage of transpedicular spine fixators.The purpose of this work was defined the value of pull-out strength of transpedicular screw from vertebrae. Impact of geometrical parameters of transpedicular screw on value of pull-out strength was analysed. Investigation was carried out on cadaver of lumbar and thoracic vertebrae. The results of investigation showed significance of geometrical parameters of transpedicular screws on durability of screw-vertebrae connection. This study proved also that very important for good performance of the vertebrae and pedicle screw unit is adequate match the screw diameter up with pedicle diameter.
- 650 _2
- $a poranění zad $x chirurgie $x terapie $7 D019567
- 650 _2
- $a ortopedické výkony $x metody $x využití $7 D019637
- 650 _2
- $a ortopedické fixační pomůcky $x využití $7 D009984
- 650 _2
- $a kostní šrouby $x normy $x využití $7 D001863
- 650 _2
- $a biomechanika $7 D001696
- 650 _2
- $a pevnost v tahu $7 D013718
- 650 _2
- $a bederní obratle $x chirurgie $7 D008159
- 650 _2
- $a hrudní obratle $x chirurgie $7 D013904
- 650 _2
- $a radiografie $x metody $x využití $7 D011859
- 650 _2
- $a statistika jako téma $7 D013223
- 650 _2
- $a dospělí $7 D000328
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a financování organizované $7 D005381
- 700 1_
- $a Filipiak, J.
- 773 0_
- $w MED00011216 $t Pohybové ústrojí $g Roč. 10, č. 1-2 (2003), s. 61-68 $x 1212-4575
- 856 41
- $u http://www.pojivo.cz/pu/PU_12_2003.pdf $y plný text volně přístupný
- 910 __
- $a ABA008 $b A 4059 $c 509 a $y 8
- 990 __
- $a 20100413085328 $b ABA008
- 991 __
- $a 20100813115239 $b ABA008
- 999 __
- $a ok $b bmc $g 722442 $s 585549
- BAS __
- $a 3
- BMC __
- $a 2003 $b 10 $c 1-2 $d 61-68 $i 1212-4575 $m Pohybové ústrojí. Pokroky ve výzkumu, diagnostice a terapii $x MED00011216
- LZP __
- $a 2010-19/vtal