Srovnání otěrových charakteristik polyetylénu in vivo u náhrad kolena s kovovou a keramickou femorální komponentou
[Comparison of in vivo characteristics of polyethylene wear particles produced by a metal and a ceramic femoral component in total knee replacement]

D. Veigl, P. Vavřík, D. Pokorný, M. Šlouf, E. Pavlova, I. Landor

Language Czech Country Czech Republic

Document type Comparative Study

Digital library NLK
Source

E-resources

NLK Free Medical Journals from 2006

PURPOSE OF THE STUDY The aim of the study was to evaluate in vivo and compare, in terms of the quality and number of ultra high-molecular polyethylene (UHMWPE) wear particles, total knee replacements of identical construction differing only in the material used for femoral component production, i.e., CoCrMo alloy or ZrO2 ceramics. MATERIAL AND METHODS Samples of peri-prosthetic granuloma tissue were collected in two patients with total knee replacement suffering from implant migration, who were matched in relevant characteristics. The primary knee replacement in Patient 1 with a CoCrMo femoral component was done 7.2 years and in Patient 2 with a ZrO2 implant 6.8 years before this assessment. The polyethylene wear-induced granuloma was analysed by the MORF method enabling us to assess the shape and size of wear debris and the IRc method for assessment of particle concentration. RESULTS In the granuloma tissue samples of Patient 1, on the average, particles were 0.30 mm in size and their relative volume was 0.19. In the Patient 2 tissue samples, the average size of particles was 0.33 mm and their relative volume was 0.26. There was no significant difference in either particle morphology or their concentration in the granuloma tissue between the two patients. DISCUSSION One of the options of how to reduce the production of polyethylene wear particles is to improve the tribological properties of contacting surfaces in total knee replacement by substituting a cobalt-chrome femoral component with a zirconia ceramic femoral component. The previous in vitro testing carried out with a mechanical simulator under conditions approaching a real weight-bearing in the human body did show a nearly three-fold decrease in the number of UHMWPE wear particles in zirconia components. The evaluation of granuloma tissue induced by the activity of a real prosthetic joint for nearly seven years, however, did not reveal any great difference in either quality or quantity of polyethylene debris between the two replacements. The difference of surface roughness between CoCrMo (Ra = 0.05) and ZrO2 (Ra = 0.02) components did not play any role in in vivo conditions. CONCLUSIONS In accordance with a previous clinical study, this evaluation of the quality and quantity of UHMWPE wear particles produced by a ceramic femoral component in vivo failed to demonstrate any advantage of zirconia ceramic components over the cobalt-chrome femoral components so far used.

Comparison of in vivo characteristics of polyethylene wear particles produced by a metal and a ceramic femoral component in total knee replacement

Bibliography, etc.

Lit.: 49

000      
00000naa 2200000 a 4500
001      
bmc11004180
003      
CZ-PrNML
005      
20111210202555.0
008      
110304s2011 xr e cze||
009      
AR
024    7_
$2 doi $a 10.55095/achot2011/008
040    __
$a ABA008 $b cze $c ABA008 $d ABA008 $e AACR2
041    0_
$a cze $b eng
044    __
$a xr
100    1_
$a Veigl, David $7 xx0143524
245    10
$a Srovnání otěrových charakteristik polyetylénu in vivo u náhrad kolena s kovovou a keramickou femorální komponentou / $c D. Veigl, P. Vavřík, D. Pokorný, M. Šlouf, E. Pavlova, I. Landor
246    11
$a Comparison of in vivo characteristics of polyethylene wear particles produced by a metal and a ceramic femoral component in total knee replacement
314    __
$a I. ortopedická klinika 1. LF UK a FN Motol, Praha
504    __
$a Lit.: 49
520    9_
$a PURPOSE OF THE STUDY The aim of the study was to evaluate in vivo and compare, in terms of the quality and number of ultra high-molecular polyethylene (UHMWPE) wear particles, total knee replacements of identical construction differing only in the material used for femoral component production, i.e., CoCrMo alloy or ZrO2 ceramics. MATERIAL AND METHODS Samples of peri-prosthetic granuloma tissue were collected in two patients with total knee replacement suffering from implant migration, who were matched in relevant characteristics. The primary knee replacement in Patient 1 with a CoCrMo femoral component was done 7.2 years and in Patient 2 with a ZrO2 implant 6.8 years before this assessment. The polyethylene wear-induced granuloma was analysed by the MORF method enabling us to assess the shape and size of wear debris and the IRc method for assessment of particle concentration. RESULTS In the granuloma tissue samples of Patient 1, on the average, particles were 0.30 mm in size and their relative volume was 0.19. In the Patient 2 tissue samples, the average size of particles was 0.33 mm and their relative volume was 0.26. There was no significant difference in either particle morphology or their concentration in the granuloma tissue between the two patients. DISCUSSION One of the options of how to reduce the production of polyethylene wear particles is to improve the tribological properties of contacting surfaces in total knee replacement by substituting a cobalt-chrome femoral component with a zirconia ceramic femoral component. The previous in vitro testing carried out with a mechanical simulator under conditions approaching a real weight-bearing in the human body did show a nearly three-fold decrease in the number of UHMWPE wear particles in zirconia components. The evaluation of granuloma tissue induced by the activity of a real prosthetic joint for nearly seven years, however, did not reveal any great difference in either quality or quantity of polyethylene debris between the two replacements. The difference of surface roughness between CoCrMo (Ra = 0.05) and ZrO2 (Ra = 0.02) components did not play any role in in vivo conditions. CONCLUSIONS In accordance with a previous clinical study, this evaluation of the quality and quantity of UHMWPE wear particles produced by a ceramic femoral component in vivo failed to demonstrate any advantage of zirconia ceramic components over the cobalt-chrome femoral components so far used.
650    _2
$a senioři $7 D000368
650    _2
$a totální endoprotéza kolene $7 D019645
650    _2
$a keramika $7 D002516
650    _2
$a slitiny chromu $7 D002858
650    _2
$a analýza selhání vybavení $7 D019544
650    _2
$a ženské pohlaví $7 D005260
650    _2
$a reakce na cizí těleso $x etiologie $x patologie $7 D005549
650    _2
$a lidé $7 D006801
650    _2
$a protézy kolene $7 D007720
650    _2
$a lidé středního věku $7 D008875
650    _2
$a velikost částic $7 D010316
650    _2
$a polyethyleny $7 D011095
650    _2
$a selhání protézy $7 D011475
650    _2
$a zirkonium $7 D015040
655    _2
$a srovnávací studie $7 D003160
700    1_
$a Vavřík, Pavel, $d 1951- $7 nlk19990073980
700    1_
$a Pokorný, David, $d 1968- $7 jn20001103568
700    1_
$a Šlouf, Miroslav, $d 1973- $7 xx0107048
700    1_
$a Pavlova, Ewa $7 xx0140861
700    1_
$a Landor, Ivan, $d 1956- $7 xx0062301
773    0_
$w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae čechoslovaca $g Roč. 78, č. 1 (2011), s. 49-55 $x 0001-5415
910    __
$a ABA008 $b A 8 $c 507 $y 2
990    __
$a 20110304091944 $b ABA008
991    __
$a 20110726095526 $b ABA008
999    __
$a ok $b bmc $g 831520 $s 696210
BAS    __
$a 3
BMC    __
$a 2011 $b 78 $c 1 $d 49-55 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $x MED00011021
LZP    __
$a 2011-15/vtme

Find record