Je něco špatně v tomto záznamu ?
Portable, lightweight, low power, ion chromatographic system with open tubular capillary columns
IK. Kiplagat, P. Kubán, P. Pelcová, V. Kubán,
Jazyk angličtina Země Nizozemsko
Typ dokumentu časopisecké články, práce podpořená grantem
- MeSH
- chromatografie iontoměničová ekonomika přístrojové vybavení MeSH
- kationty analýza MeSH
- sníh chemie MeSH
- voda analýza MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
Basic operation principles of a lightweight, low power, low cost, portable ion chromatograph utilizing open tubular ion chromatography in capillary columns coated with multi-layer polymeric stationary phases are demonstrated. A minimalistic configuration of a portable IC instrument was developed that does not require any chromatographic eluent delivery system, nor sample injection device as it uses gravity-based eluent flow and hydrodynamic sample injection adopted from capillary electrophoresis. As a detection device, an inexpensive commercially available capacitance sensor is used that has been shown to be a suitable substitute for contactless conductivity detection in capillary separation systems. The built-in temperature sensor allows for baseline drift correction typically encountered in conductivity/capacitance measurements without thermostating device. The whole instrument does not require any power supply for its operation, except the detection and data acquisition part that is provided by a USB port of a Netbook computer. It is extremely lightweight, its total weight including the Netbook computer is less than 2.5kg and it can be continuously operated for more than 8h. Several parameters of the instrument, such as detection cell design, eluent delivery systems and data treatment were optimized as well as the composition of eluent for non-suppressed ion chromatographic analysis of common inorganic cations (Na(+), NH(4)(+), K(+), Cs(+), Ca(2+), Mg(2+), transition metals). Low conductivity eluents based on weakly complexing organic acids such as tartaric, oxalic or pyridine-2,6-dicarboxylic acids were used with contactless capacitance detection for simultaneous separation of mono- and divalent cations. Separation of Na(+) and NH(4)(+) cations was optimized by addition of 18-crown-6 to the eluent. The best separation of 6 metal cations commonly present in various environmental samples was accomplished in less than 30min using a 1.75mM pyridine-2,6-dicarboxylic acid and 3mM 18-crown-6 eluent with excellent repeatability (below 2%) and detection limits in the low micromolar range. The analysis of field samples is demonstrated; the concentrations of common inorganic cations in river water, mineral water and snow samples were determined.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc12025633
- 003
- CZ-PrNML
- 005
- 20121207115843.0
- 007
- ta
- 008
- 120817e20100609ne f 000 0#eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.chroma.2010.06.017 $2 doi
- 035 __
- $a (PubMed)20580367
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ne
- 100 1_
- $a Kiplagat, Isaac K $u Department of Chemistry and Biochemistry, Mendel University in Brno, Zemedelská 1, CZ-613 00 Brno, Czech Republic.
- 245 10
- $a Portable, lightweight, low power, ion chromatographic system with open tubular capillary columns / $c IK. Kiplagat, P. Kubán, P. Pelcová, V. Kubán,
- 520 9_
- $a Basic operation principles of a lightweight, low power, low cost, portable ion chromatograph utilizing open tubular ion chromatography in capillary columns coated with multi-layer polymeric stationary phases are demonstrated. A minimalistic configuration of a portable IC instrument was developed that does not require any chromatographic eluent delivery system, nor sample injection device as it uses gravity-based eluent flow and hydrodynamic sample injection adopted from capillary electrophoresis. As a detection device, an inexpensive commercially available capacitance sensor is used that has been shown to be a suitable substitute for contactless conductivity detection in capillary separation systems. The built-in temperature sensor allows for baseline drift correction typically encountered in conductivity/capacitance measurements without thermostating device. The whole instrument does not require any power supply for its operation, except the detection and data acquisition part that is provided by a USB port of a Netbook computer. It is extremely lightweight, its total weight including the Netbook computer is less than 2.5kg and it can be continuously operated for more than 8h. Several parameters of the instrument, such as detection cell design, eluent delivery systems and data treatment were optimized as well as the composition of eluent for non-suppressed ion chromatographic analysis of common inorganic cations (Na(+), NH(4)(+), K(+), Cs(+), Ca(2+), Mg(2+), transition metals). Low conductivity eluents based on weakly complexing organic acids such as tartaric, oxalic or pyridine-2,6-dicarboxylic acids were used with contactless capacitance detection for simultaneous separation of mono- and divalent cations. Separation of Na(+) and NH(4)(+) cations was optimized by addition of 18-crown-6 to the eluent. The best separation of 6 metal cations commonly present in various environmental samples was accomplished in less than 30min using a 1.75mM pyridine-2,6-dicarboxylic acid and 3mM 18-crown-6 eluent with excellent repeatability (below 2%) and detection limits in the low micromolar range. The analysis of field samples is demonstrated; the concentrations of common inorganic cations in river water, mineral water and snow samples were determined.
- 650 _2
- $a kationty $x analýza $7 D002412
- 650 _2
- $a chromatografie iontoměničová $x ekonomika $x přístrojové vybavení $7 D002852
- 650 _2
- $a sníh $x chemie $7 D012914
- 650 _2
- $a voda $x analýza $7 D014867
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Kubán, Petr
- 700 1_
- $a Pelcová, Pavlína
- 700 1_
- $a Kubán, Vlastimil
- 773 0_
- $w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1217, č. 31 (20100609), s. 5116-23
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/20580367 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y m
- 990 __
- $a 20120817 $b ABA008
- 991 __
- $a 20121207115917 $b ABA008
- 999 __
- $a ok $b bmc $g 947675 $s 782979
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2010 $b 1217 $c 31 $d 5116-23 $e 20100609 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
- LZP __
- $a Pubmed-20120817/10/03