Early fracture of a plasma cup ceramic liner: a case report and surface analysis
J. Gallo, T. Stewart, R. Novotny, J. Dusza, D. Galusek
Jazyk angličtina Země Česko
Typ dokumentu kazuistiky, časopisecké články, práce podpořená grantem
NLK
Directory of Open Access Journals
od 2001
Free Medical Journals
od 1998
Medline Complete (EBSCOhost)
od 2007-06-01
ROAD: Directory of Open Access Scholarly Resources
od 2001
PubMed
18345275
DOI
10.5507/bp.2007.058
Knihovny.cz E-zdroje
- MeSH
- analýza selhání vybavení MeSH
- dospělí MeSH
- kyčelní protézy škodlivé účinky MeSH
- lidé MeSH
- mikroskopie elektronová rastrovací MeSH
- náhrada kyčelního kloubu MeSH
- oxid hlinitý MeSH
- reoperace MeSH
- selhání protézy MeSH
- Check Tag
- dospělí MeSH
- lidé MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- kazuistiky MeSH
- práce podpořená grantem MeSH
BACKGROUND: Ceramic surfaces offer excellent tribological properties and for this reason are frequently used in younger patients with osteoarthritis of the hip. However, they may fracture easily owing to low tolerance to tensile and impact stresses in particular. We report a case of early fracture of a currently available ceramic liner in a 44-year-old woman with revision total hip arthroplasty. METHODS AND RESULTS: Scanning electron microscopy and macroscopic fractographic examination were done to elucidate the cause of failure. These showed the fracture started most probably at the inner surface of the cup. Several mechanisms of fracture initiation are discussed in the paper. CONCLUSION: This is the first report on early fracture of the Plasma cup ceramic liner in white man. The fracture occurred intraoperatively or early postoperatively while the diagnosis was determined much later despite the patient had experienced sensations of subluxations long time before.
Department of Microscopy Methods Faculty of Medicine and Dentistry Palacky University Czech Republic
Institute of Material Research Slovak Academy of Sciences Kosice
Joint Glass Centre of the IIC SAS TnUAD and RONA Trencin Slovak Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc13003591
- 003
- CZ-PrNML
- 005
- 20130225152716.0
- 007
- ta
- 008
- 130128s2007 xr of f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.5507/bp.2007.058 $2 doi
- 035 __
- $a (PubMed)18345275
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Gallo, Jiří $7 xx0019005 $u Department of Orthopaedics, Teaching Hospital, Faculty of Medicine and Dentistry, Palacky University, Olomouc
- 245 10
- $a Early fracture of a plasma cup ceramic liner: a case report and surface analysis / $c J. Gallo, T. Stewart, R. Novotny, J. Dusza, D. Galusek
- 520 9_
- $a BACKGROUND: Ceramic surfaces offer excellent tribological properties and for this reason are frequently used in younger patients with osteoarthritis of the hip. However, they may fracture easily owing to low tolerance to tensile and impact stresses in particular. We report a case of early fracture of a currently available ceramic liner in a 44-year-old woman with revision total hip arthroplasty. METHODS AND RESULTS: Scanning electron microscopy and macroscopic fractographic examination were done to elucidate the cause of failure. These showed the fracture started most probably at the inner surface of the cup. Several mechanisms of fracture initiation are discussed in the paper. CONCLUSION: This is the first report on early fracture of the Plasma cup ceramic liner in white man. The fracture occurred intraoperatively or early postoperatively while the diagnosis was determined much later despite the patient had experienced sensations of subluxations long time before.
- 650 _2
- $a dospělí $7 D000328
- 650 _2
- $a oxid hlinitý $7 D000537
- 650 _2
- $a náhrada kyčelního kloubu $7 D019644
- 650 _2
- $a analýza selhání vybavení $7 D019544
- 650 _2
- $a ženské pohlaví $7 D005260
- 650 _2
- $a kyčelní protézy $x škodlivé účinky $7 D006622
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a mikroskopie elektronová rastrovací $7 D008855
- 650 _2
- $a selhání protézy $7 D011475
- 650 _2
- $a reoperace $7 D012086
- 655 _2
- $a kazuistiky $7 D002363
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Stewart, Todd $u Institute of Medical and Biological Engineering, School of Mechanical Engineering University of Leeds, United Kingdom
- 700 1_
- $a Novotný, Radko $7 xx0061155 $u Department of Microscopy Methods, Faculty of Medicine and Dentistry, Palacky University, Czech Republic
- 700 1_
- $a Dusza, Ján. $7 xx0314970 $u Institute of Material Research, Slovak Academy of Sciences, Kosice
- 700 1_
- $a Galusek, Dušan. $7 _AN071435 $u Joint Glass Centre of the IIC SAS, TnUAD and RONA, Trencin, Slovak Republic
- 773 0_
- $w MED00012606 $t Biomedical papers of the Medical Faculty of the University Palacký, Olomouc, Czech Republic $x 1213-8118 $g Roč. 151, č. 2 (2007), s. 341-346
- 910 __
- $a ABA008 $b A 1502 $c sign $y 3 $z 0
- 990 __
- $a 20130128 $b ABA008
- 991 __
- $a 20130225152914 $b ABA008
- 999 __
- $a ok $b bmc $g 966247 $s 801786
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2007 $b 151 $c 2 $d 341-346 $i 1213-8118 $m Biomedical papers of the Medical Faculty of the University Palacký, Olomouc Czech Republic $n Biomed. Pap. Fac. Med. Palacký Univ. Olomouc Czech Repub. (Print) $x MED00012606
- LZP __
- $b NLK111 $a Pubmed-20130128