Enhanced capabilities of separation in Sequential Injection Chromatography--fused-core particle column and its comparison with narrow-bore monolithic column

P. Chocholouš, L. Kosařová, D. Satínský, H. Sklenářová, P. Solich,

. 2011 ; 85 (2) : 1129-34.

Jazyk angličtina Země Anglie, Velká Británie

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/bmc13015850

In the Sequential Injection Chromatography (SIC) only monolithic columns for chromatographic separations have been used so far. This article presents the first use of fused-core particle packed column in an attempt to extend of the chromatographic capabilities of the SIC system. A new fused-core particle column (2.7 μm) Ascentis(®) Express C18 (Supelco™ Analytical) 30 mm × 4.6 mm brings high separation efficiency within flow rates and pressures comparable to monolithic column Chromolith(®) Performance RP-18e 100-3 (Merck(®)) 100 mm × 3 mm. Both columns matches the conditions of the commercially produced SIC system - SIChrom™ (8-port high-pressure selection valve and medium-pressure Sapphire™ syringe pump with 4 mL reservoir - maximal work pressure 1000 PSI) (FIAlab(®), USA). The system was tested by the separation of four estrogens with similar structure and an internal standard - ethylparaben. The mobile phase composed of acetonitrile/water (40/60 (v/v)) was pumped isocratic at flow rate 0.48 mL min(-1). Spectrophotometric detection was performed at wavelength of 225 nm and injected volume of sample solutions was 10 μL. The chromatographic characteristics of both columns were compared. Obtained results and conclusions have shown that both fused-core particle column and longer narrow shaped monolithic column bring benefits into the SIC method.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc13015850
003      
CZ-PrNML
005      
20130424134929.0
007      
ta
008      
130424s2011 enk f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.talanta.2011.05.027 $2 doi
035    __
$a (PubMed)21726748
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a enk
100    1_
$a Chocholouš, Petr $u Department of Analytical Chemistry, Faculty of Pharmacy, Charles University, Heyrovského 1203, 500 05 Hradec Králové, Czech Republic. petr.chocholous@faf.cuni.cz
245    10
$a Enhanced capabilities of separation in Sequential Injection Chromatography--fused-core particle column and its comparison with narrow-bore monolithic column / $c P. Chocholouš, L. Kosařová, D. Satínský, H. Sklenářová, P. Solich,
520    9_
$a In the Sequential Injection Chromatography (SIC) only monolithic columns for chromatographic separations have been used so far. This article presents the first use of fused-core particle packed column in an attempt to extend of the chromatographic capabilities of the SIC system. A new fused-core particle column (2.7 μm) Ascentis(®) Express C18 (Supelco™ Analytical) 30 mm × 4.6 mm brings high separation efficiency within flow rates and pressures comparable to monolithic column Chromolith(®) Performance RP-18e 100-3 (Merck(®)) 100 mm × 3 mm. Both columns matches the conditions of the commercially produced SIC system - SIChrom™ (8-port high-pressure selection valve and medium-pressure Sapphire™ syringe pump with 4 mL reservoir - maximal work pressure 1000 PSI) (FIAlab(®), USA). The system was tested by the separation of four estrogens with similar structure and an internal standard - ethylparaben. The mobile phase composed of acetonitrile/water (40/60 (v/v)) was pumped isocratic at flow rate 0.48 mL min(-1). Spectrophotometric detection was performed at wavelength of 225 nm and injected volume of sample solutions was 10 μL. The chromatographic characteristics of both columns were compared. Obtained results and conclusions have shown that both fused-core particle column and longer narrow shaped monolithic column bring benefits into the SIC method.
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Kosařová, Lucie $u -
700    1_
$a Satínský, Dalibor $u -
700    1_
$a Sklenářová, Hana $u -
700    1_
$a Solich, Petr $u -
773    0_
$w MED00004484 $t Talanta $x 1873-3573 $g Roč. 85, č. 2 (2011), s. 1129-34
856    41
$u https://pubmed.ncbi.nlm.nih.gov/21726748 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20130424 $b ABA008
991    __
$a 20130424135216 $b ABA008
999    __
$a ind $b bmc $g 979051 $s 814171
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2011 $b 85 $c 2 $d 1129-34 $i 1873-3573 $m Talanta $n Talanta $x MED00004484
LZP    __
$a Pubmed-20130424

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...