- Research Support, Non-U.S. Gov't MeSH
The aim of this study is to consider the relevance of in situ measurements of bovine serum film thickness in the optical test device that could be related to the function of the artificial hip joint. It is mainly focussed on the effect of the hydrophobicity or hydrophilicity of the transparent surface and the effect of its geometry. Film thickness measurements were performed using ball-on-disc and lens-on-disc configurations of optical test device as a function of time. Chromatic interferograms were recorded with a high-speed complementary metal-oxide semiconductor digital camera and evaluated with thin film colorimetric interferometry. It was clarified that a chromium layer covering the glass disc has a hydrophobic behaviour which supports the adsorption of proteins contained in the bovine serum solution, thereby a thicker lubricating film is formed. On the contrary, the protein film formation was not observed when the disc was covered with a silica layer having a hydrophilic behaviour. In this case, a very thin lubricating film was formed only due to the hydrodynamic effect. Metal and ceramic balls have no substantial effect on lubricant film formation although their contact surfaces have relatively different wettability. It was confirmed that conformity of contacting surfaces and kinematic conditions has fundamental effect on bovine serum film formation. In the ball-on-disc configuration, the lubricant film is formed predominantly due to protein aggregations, which pass through the contact zone and increase the film thickness. In the more conformal ball-on-lens configuration, the lubricant film is formed predominantly due to hydrodynamic effect, thereby the film thickness is kept constant during measurement.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc14074356
- 003
- CZ-PrNML
- 005
- 20190708090227.0
- 007
- ta
- 008
- 141006s2014 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1177/0954411913517498 $2 doi
- 035 __
- $a (PubMed)24398447
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Vrbka, Martin, $u Faculty of Mechanical Engineering, Brno University of Technology, Brno, Czech Republic. $d 1977- $7 mzk2004248591
- 245 10
- $a In situ measurements of thin films in bovine serum lubricated contacts using optical interferometry / $c M. Vrbka, I. Křupka, M. Hartl, T. Návrat, J. Gallo, A. Galandáková,
- 520 9_
- $a The aim of this study is to consider the relevance of in situ measurements of bovine serum film thickness in the optical test device that could be related to the function of the artificial hip joint. It is mainly focussed on the effect of the hydrophobicity or hydrophilicity of the transparent surface and the effect of its geometry. Film thickness measurements were performed using ball-on-disc and lens-on-disc configurations of optical test device as a function of time. Chromatic interferograms were recorded with a high-speed complementary metal-oxide semiconductor digital camera and evaluated with thin film colorimetric interferometry. It was clarified that a chromium layer covering the glass disc has a hydrophobic behaviour which supports the adsorption of proteins contained in the bovine serum solution, thereby a thicker lubricating film is formed. On the contrary, the protein film formation was not observed when the disc was covered with a silica layer having a hydrophilic behaviour. In this case, a very thin lubricating film was formed only due to the hydrodynamic effect. Metal and ceramic balls have no substantial effect on lubricant film formation although their contact surfaces have relatively different wettability. It was confirmed that conformity of contacting surfaces and kinematic conditions has fundamental effect on bovine serum film formation. In the ball-on-disc configuration, the lubricant film is formed predominantly due to protein aggregations, which pass through the contact zone and increase the film thickness. In the more conformal ball-on-lens configuration, the lubricant film is formed predominantly due to hydrodynamic effect, thereby the film thickness is kept constant during measurement.
- 650 _2
- $a adsorpce $7 D000327
- 650 _2
- $a zvířata $7 D000818
- 650 _2
- $a biomechanika $7 D001696
- 650 _2
- $a skot $7 D002417
- 650 12
- $a kyčelní protézy $7 D006622
- 650 _2
- $a hydrofobní a hydrofilní interakce $7 D057927
- 650 _2
- $a interferometrie $x metody $7 D007368
- 650 _2
- $a lubrikanty $x chemie $7 D054327
- 650 _2
- $a testování materiálů $x metody $7 D008422
- 650 _2
- $a sérum $x chemie $7 D044967
- 650 _2
- $a smáčivost $7 D017155
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Křupka, Ivan, $d 1967- $7 mzk2002157050
- 700 1_
- $a Hartl, Martin, $d 1967- $7 mzk2002157052
- 700 1_
- $a Návrat, Tomáš, $d 1978- $7 mzk2005279266
- 700 1_
- $a Gallo, Jiří $7 xx0019005
- 700 1_
- $a Galandáková, Adéla, $d 1979- $7 stk2008428623
- 773 0_
- $w MED00008931 $t Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine $x 2041-3033 $g Roč. 228, č. 2 (2014), s. 149-158
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/24398447 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20141006 $b ABA008
- 991 __
- $a 20190708090424 $b ABA008
- 999 __
- $a ok $b bmc $g 1042239 $s 873268
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2014 $b 228 $c 2 $d 149-158 $i 2041-3033 $m Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine $n Proc Inst Mech Eng [H] $x MED00008931
- GRA __
- $a NT14267 $p MZ0
- LZP __
- $a Pubmed-20141006