Reduction of effective dose and organ dose to the eye lens in head MDCT using iterative image reconstruction and automatic tube current modulation

P. Ryska, T. Kvasnicka, J. Jandura, L. Klzo, J. Grepl, J. Zizka

. 2014 ; 158 (2) : 265-272. [pub] 20131003

Language English Country Czech Republic

Document type Journal Article

AIMS: To compare the effective and eye lens radiation dose in helical MDCT brain examinations using automatic tube current modulation in conjunction with either standard filtered back projection (FBP) technique or iterative reconstruction in image space (IRIS). METHODS: Of 400 adult brain MDCT examinations, 200 were performed using FBP and 200 using IRIS with the following parameters: tube voltage 120 kV, rotation period 1 second, pitch factor 0.55, automatic tube current modulation in both transverse and longitudinal planes with reference mAs 300 (FBP) and 200 (IRIS). Doses were calculated from CT dose index and dose length product values utilising ImPACT software; the organ dose to the lens was derived from the actual tube current-time product value applied to the lens. Image quality was assessed by two independent readers blinded to the type of image reconstruction technique. RESULTS: The average effective scan dose was 1.47±0.26 mSv (FBP) and 0.98±0.15 mSv (IRIS), respectively (33.3% decrease). The average organ dose to the eye lens decreased from 40.0±3.3 mGy (FBP) to 26.6±2.0 mGy (IRIS, 33.5% decrease). No significant change in diagnostic image quality was noted between IRIS and FBP scans (P=0.17). CONCLUSION: Iterative reconstruction of cerebral MDCT examinations enables reduction of both effective and organ eye lens dose by one third without signficant loss of image quality.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc15033231
003      
CZ-PrNML
005      
20151030145018.0
007      
ta
008      
151016s2014 xr a f 000 0|eng||
009      
AR
024    7_
$a 10.5507/bp.2013.071 $2 doi
035    __
$a (PubMed)24091879
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xr
100    1_
$a Ryška, Pavel, $u Department of Radiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic $d 1971- $7 xx0079145
245    10
$a Reduction of effective dose and organ dose to the eye lens in head MDCT using iterative image reconstruction and automatic tube current modulation / $c P. Ryska, T. Kvasnicka, J. Jandura, L. Klzo, J. Grepl, J. Zizka
520    9_
$a AIMS: To compare the effective and eye lens radiation dose in helical MDCT brain examinations using automatic tube current modulation in conjunction with either standard filtered back projection (FBP) technique or iterative reconstruction in image space (IRIS). METHODS: Of 400 adult brain MDCT examinations, 200 were performed using FBP and 200 using IRIS with the following parameters: tube voltage 120 kV, rotation period 1 second, pitch factor 0.55, automatic tube current modulation in both transverse and longitudinal planes with reference mAs 300 (FBP) and 200 (IRIS). Doses were calculated from CT dose index and dose length product values utilising ImPACT software; the organ dose to the lens was derived from the actual tube current-time product value applied to the lens. Image quality was assessed by two independent readers blinded to the type of image reconstruction technique. RESULTS: The average effective scan dose was 1.47±0.26 mSv (FBP) and 0.98±0.15 mSv (IRIS), respectively (33.3% decrease). The average organ dose to the eye lens decreased from 40.0±3.3 mGy (FBP) to 26.6±2.0 mGy (IRIS, 33.5% decrease). No significant change in diagnostic image quality was noted between IRIS and FBP scans (P=0.17). CONCLUSION: Iterative reconstruction of cerebral MDCT examinations enables reduction of both effective and organ eye lens dose by one third without signficant loss of image quality.
650    _2
$a dospělí $7 D000328
650    _2
$a senioři nad 80 let $7 D000369
650    _2
$a mozek $x radiografie $7 D001921
650    _2
$a ženské pohlaví $7 D005260
650    _2
$a lidé $7 D006801
650    _2
$a počítačové zpracování obrazu $x metody $7 D007091
650    12
$a oční čočka $7 D007908
650    _2
$a mužské pohlaví $7 D008297
650    _2
$a lidé středního věku $7 D008875
650    _2
$a multidetektorová počítačová tomografie $x metody $7 D061330
650    _2
$a prospektivní studie $7 D011446
650    _2
$a dávka záření $7 D011829
650    _2
$a poměr signál - šum $7 D059629
655    _2
$a časopisecké články $7 D016428
700    1_
$a Kvasnička, Tomáš $7 xx0189651 $u Department of Radiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic
700    1_
$a Jandura, Jiří $7 xx0230622 $u Department of Radiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic
700    1_
$a Klzo, Ludovít $7 xx0081745 $u Department of Radiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic
700    1_
$a Grepl, Jakub $7 xx0230649 $u Department of Radiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic
700    1_
$a Žižka, Jan, $d 1968- $7 skuk0001531 $u Department of Radiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic
773    0_
$w MED00012606 $t Biomedical papers of the Medical Faculty of the University Palacký, Olomouc, Czech Republic $x 1213-8118 $g Roč. 158, č. 2 (2014), s. 265-272
856    41
$u http://biomed.papers.upol.cz/ $y domovská stránka časopisu
910    __
$a ABA008 $b A 1502 $c 958 $y 4 $z 0
990    __
$a 20151016 $b ABA008
991    __
$a 20151029083221 $b ABA008
999    __
$a ok $b bmc $g 1095629 $s 916362
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2014 $b 158 $c 2 $d 265-272 $e 20131003 $i 1213-8118 $m Biomedical papers of the Medical Faculty of the University Palacký, Olomouc Czech Republic $n Biomed. Pap. Fac. Med. Palacký Univ. Olomouc Czech Repub. (Print) $x MED00012606
LZP    __
$b NLK118 $a Pubmed-20151016

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...