Something wrong with this record ?
A modular Geant4 model of Leksell Gamma Knife Perfexion™
J. Pipek, J. Novotný, J. Novotný, P. Kozubíková,
Language English Country England, Great Britain
Document type Journal Article, Research Support, Non-U.S. Gov't
- MeSH
- Phantoms, Imaging * MeSH
- Humans MeSH
- Monte Carlo Method * MeSH
- Neoplasms surgery MeSH
- Radiotherapy Planning, Computer-Assisted MeSH
- Computer Simulation * MeSH
- Radiosurgery methods MeSH
- Models, Statistical * MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
This work presents a Monte Carlo model of Leksell Gamma Knife Perfexion as well as the main parameters of the dose distribution in the standard phantom obtained using this model. The model is developed in the Geant4 simulation toolkit in a modular way which enables its reuse in other Perfexion studies. Large phase space files were created, containing particles that are entering the inner machine cavity after being transported through the collimation system. All 14 output factors of the machine and effective output factors for both the 4 mm (0.830 ± 0.009) and 8 mm (0.921 ± 0.004) collimators were calculated. Dose profiles along the main axes are also included for each collimator size. All results are compared to the values obtained from the treatment planning system, from experiments, and from other Monte Carlo models.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc16010694
- 003
- CZ-PrNML
- 005
- 20160418100437.0
- 007
- ta
- 008
- 160408s2014 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1088/0031-9155/59/24/7609 $2 doi
- 024 7_
- $a 10.1088/0031-9155/59/24/7609 $2 doi
- 035 __
- $a (PubMed)25415510
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Pipek, J $u Czech Technical University in Prague, Faculty of Nuclear Sciences and Physical Engineering, Břehová 7, 115 19 Prague 1, Czech Republic.
- 245 12
- $a A modular Geant4 model of Leksell Gamma Knife Perfexion™ / $c J. Pipek, J. Novotný, J. Novotný, P. Kozubíková,
- 520 9_
- $a This work presents a Monte Carlo model of Leksell Gamma Knife Perfexion as well as the main parameters of the dose distribution in the standard phantom obtained using this model. The model is developed in the Geant4 simulation toolkit in a modular way which enables its reuse in other Perfexion studies. Large phase space files were created, containing particles that are entering the inner machine cavity after being transported through the collimation system. All 14 output factors of the machine and effective output factors for both the 4 mm (0.830 ± 0.009) and 8 mm (0.921 ± 0.004) collimators were calculated. Dose profiles along the main axes are also included for each collimator size. All results are compared to the values obtained from the treatment planning system, from experiments, and from other Monte Carlo models.
- 650 12
- $a počítačová simulace $7 D003198
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a statistické modely $7 D015233
- 650 12
- $a metoda Monte Carlo $7 D009010
- 650 _2
- $a nádory $x chirurgie $7 D009369
- 650 12
- $a fantomy radiodiagnostické $7 D019047
- 650 _2
- $a radiochirurgie $x metody $7 D016634
- 650 _2
- $a plánování radioterapie pomocí počítače $7 D011880
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Novotný, J
- 700 1_
- $a Novotný, J
- 700 1_
- $a Kozubíková, P
- 773 0_
- $w MED00003821 $t Physics in medicine and biology $x 1361-6560 $g Roč. 59, č. 24 (2014), s. 7609-23
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/25415510 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20160408 $b ABA008
- 991 __
- $a 20160418100524 $b ABA008
- 999 __
- $a ok $b bmc $g 1114123 $s 935062
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2014 $b 59 $c 24 $d 7609-23 $i 1361-6560 $m Physics in medicine and biology $n Phys Med Biol $x MED00003821
- LZP __
- $a Pubmed-20160408