Examining exposure reciprocity in a resin based composite using high irradiance levels and real-time degree of conversion values
D. Selig, T. Haenel, B. Hausnerová, B. Moeginger, D. Labrie, B. Sullivan, RB. Price,
Jazyk angličtina Země Anglie, Velká Británie
Typ dokumentu časopisecké články, práce podpořená grantem
- MeSH
- dentální adheziva tuhnoucí světlem * MeSH
- polymerizace MeSH
- složené pryskyřice chemie účinky záření MeSH
- spektroskopie infračervená s Fourierovou transformací MeSH
- stomatologické polymerizační lampy MeSH
- testování materiálů MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
OBJECTIVE: Exposure reciprocity suggests that, as long as the same radiant exposure is delivered, different combinations of irradiance and exposure time will achieve the same degree of resin polymerization. This study examined the validity of exposure reciprocity using real time degree of conversion results from one commercial flowable dental resin. Additionally a new fitting function to describe the polymerization kinetics is proposed. METHODS: A Plasma Arc Light Curing Unit (LCU) was used to deliver 0.75, 1.2, 1.5, 3.7 or 7.5 W/cm(2) to 2mm thick samples of Tetric EvoFlow (Ivoclar Vivadent). The irradiances and radiant exposures received by the resin were determined using an integrating sphere connected to a fiber-optic spectrometer. The degree of conversion (DC) was recorded at a rate of 8.5 measurements a second at the bottom of the resin using attenuated total reflectance Fourier Transform mid-infrared spectroscopy (FT-MIR). Five specimens were exposed at each irradiance level. The DC reached after 170s and after 5, 10 and 15 J/cm(2) had been delivered was compared using analysis of variance and Fisher's PLSD post hoc multiple comparison tests (alpha=0.05). RESULTS: The same DC values were not reached after the same radiant exposures of 5, 10 and 15 J/cm(2) had been delivered at an irradiance of 3.7 and 7.5 W/cm(2). Thus exposure reciprocity was not supported for Tetric EvoFlow (p<0.05). SIGNIFICANCE: For Tetric EvoFlow, there was no significant difference in the DC when 5, 10 and 15J/cm(2) were delivered at irradiance levels of 0.75, 1.2 and 1.5 W/cm(2). The optimum combination of irradiance and exposure time for this commercial dental resin may be close to 1.5 W/cm(2) for 12s.
Bonn Rhein Sieg University of Applied Sciences Department of Natural Sciences Rheinbach Germany
Dalhousie University Department of Dental Clinical Sciences Faculty of Dentistry Halifax Canada
Dalhousie University Department of Physics and Atmospheric Sciences Halifax Canada
FH Aachen University of Applied Sciences Department of Natural Sciences Jülich Germany
Tomas Bata University in Zlin University Institute Centre of Polymer Systems Zlin Czech Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc17014310
- 003
- CZ-PrNML
- 005
- 20170428111205.0
- 007
- ta
- 008
- 170413s2015 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.dental.2015.02.010 $2 doi
- 035 __
- $a (PubMed)25804190
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Selig, Daniela $u Dalhousie University, Department of Dental Clinical Sciences, Faculty of Dentistry, Halifax, Canada; Bonn-Rhein-Sieg University of Applied Sciences, Department of Natural Sciences, Rheinbach, Germany; FH Aachen University of Applied Sciences, Department of Natural Sciences, Jülich, Germany.
- 245 10
- $a Examining exposure reciprocity in a resin based composite using high irradiance levels and real-time degree of conversion values / $c D. Selig, T. Haenel, B. Hausnerová, B. Moeginger, D. Labrie, B. Sullivan, RB. Price,
- 520 9_
- $a OBJECTIVE: Exposure reciprocity suggests that, as long as the same radiant exposure is delivered, different combinations of irradiance and exposure time will achieve the same degree of resin polymerization. This study examined the validity of exposure reciprocity using real time degree of conversion results from one commercial flowable dental resin. Additionally a new fitting function to describe the polymerization kinetics is proposed. METHODS: A Plasma Arc Light Curing Unit (LCU) was used to deliver 0.75, 1.2, 1.5, 3.7 or 7.5 W/cm(2) to 2mm thick samples of Tetric EvoFlow (Ivoclar Vivadent). The irradiances and radiant exposures received by the resin were determined using an integrating sphere connected to a fiber-optic spectrometer. The degree of conversion (DC) was recorded at a rate of 8.5 measurements a second at the bottom of the resin using attenuated total reflectance Fourier Transform mid-infrared spectroscopy (FT-MIR). Five specimens were exposed at each irradiance level. The DC reached after 170s and after 5, 10 and 15 J/cm(2) had been delivered was compared using analysis of variance and Fisher's PLSD post hoc multiple comparison tests (alpha=0.05). RESULTS: The same DC values were not reached after the same radiant exposures of 5, 10 and 15 J/cm(2) had been delivered at an irradiance of 3.7 and 7.5 W/cm(2). Thus exposure reciprocity was not supported for Tetric EvoFlow (p<0.05). SIGNIFICANCE: For Tetric EvoFlow, there was no significant difference in the DC when 5, 10 and 15J/cm(2) were delivered at irradiance levels of 0.75, 1.2 and 1.5 W/cm(2). The optimum combination of irradiance and exposure time for this commercial dental resin may be close to 1.5 W/cm(2) for 12s.
- 650 _2
- $a složené pryskyřice $x chemie $x účinky záření $7 D003188
- 650 _2
- $a stomatologické polymerizační lampy $7 D055117
- 650 12
- $a dentální adheziva tuhnoucí světlem $7 D055115
- 650 _2
- $a testování materiálů $7 D008422
- 650 _2
- $a polymerizace $7 D058105
- 650 _2
- $a spektroskopie infračervená s Fourierovou transformací $7 D017550
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Haenel, Thomas $u Dalhousie University, Department of Dental Clinical Sciences, Faculty of Dentistry, Halifax, Canada; Bonn-Rhein-Sieg University of Applied Sciences, Department of Natural Sciences, Rheinbach, Germany; Tomas Bata University in Zlin, Faculty of Technology, Department of Production Engineering, Zlin, Czech Republic; Tomas Bata University in Zlin, University Institute, Centre of Polymer Systems, Zlin, Czech Republic.
- 700 1_
- $a Hausnerová, Berenika $u Tomas Bata University in Zlin, Faculty of Technology, Department of Production Engineering, Zlin, Czech Republic; Tomas Bata University in Zlin, University Institute, Centre of Polymer Systems, Zlin, Czech Republic.
- 700 1_
- $a Moeginger, Bernhard $u Bonn-Rhein-Sieg University of Applied Sciences, Department of Natural Sciences, Rheinbach, Germany.
- 700 1_
- $a Labrie, Daniel $u Dalhousie University, Department of Physics and Atmospheric Sciences, Halifax, Canada.
- 700 1_
- $a Sullivan, Braden $u Dalhousie University, Department of Dental Clinical Sciences, Faculty of Dentistry, Halifax, Canada.
- 700 1_
- $a Price, Richard B T $u Dalhousie University, Department of Dental Clinical Sciences, Faculty of Dentistry, Halifax, Canada. Electronic address: rbprice@dal.ca.
- 773 0_
- $w MED00001334 $t Dental materials official publication of the Academy of Dental Materials $x 1879-0097 $g Roč. 31, č. 5 (2015), s. 583-93
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/25804190 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20170413 $b ABA008
- 991 __
- $a 20170428111526 $b ABA008
- 999 __
- $a ok $b bmc $g 1200775 $s 975088
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2015 $b 31 $c 5 $d 583-93 $e 20150321 $i 1879-0097 $m Dental materials $n Dent Mater $x MED00001334
- LZP __
- $a Pubmed-20170413