Injections from sub-μL sample volumes in commercial capillary electrophoresis
A. Šlampová, P. Kubáň,
Jazyk angličtina Země Nizozemsko
Typ dokumentu časopisecké články
- MeSH
- elektroforéza kapilární přístrojové vybavení metody MeSH
- elektrolyty chemie MeSH
- hmotnostní spektrometrie MeSH
- nortriptylin analýza izolace a purifikace MeSH
- papaverin analýza MeSH
- roztoky chemie MeSH
- spektrofotometrie ultrafialová MeSH
- Publikační typ
- časopisecké články MeSH
A simple sample injection procedure compatible with commercial capillary electrophoresis (CE) instrumentation was developed, which enables handling sample volumes as little as 250nL for analytical applications where sample volume availability is of concern. Single-use micro-sampling inserts were prepared by thermal modification of polypropylene micropipette tips and the inserts were accommodated in standard CE vials in CE autosampler carousel. To ensure direct contact of separation capillary injection end with sample solution and to avoid possible damage to the capillary, a soft compression spring was placed at the bottom of the vial underneath the micro-sampling insert. Injections from sub-μL samples were carried out in conventional as well as in short-end injection mode, were compatible with standard i.d./o.d. (25-100μm/365μm) fused silica capillaries and with various background electrolyte solutions and detection modes. Excellent repeatability of replicate injections from 250nL to 3μL was achieved based on RSD values of quantitative analytical measures (peak heights ≤2.4% and peak areas ≤3.7%) for CE-UV-vis, CE-ESI-MS and CE-contactless conductivity detection of model basic drugs. The achieved RSD values were comparable with those for replicate injections of the drugs from standard CE vials. The reported concept of injections from micro-sampling inserts was further demonstrated useful in evaluation of micro-electromembrane extraction (μ-EME) of model basic drugs. Sub-μL volumes of operational solutions resulted in reduced lengths of μ-EME phases and improved extraction recoveries (66-91%) were achieved.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc17023155
- 003
- CZ-PrNML
- 005
- 20170720123127.0
- 007
- ta
- 008
- 170720s2017 ne f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.chroma.2017.03.064 $2 doi
- 035 __
- $a (PubMed)28381364
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ne
- 100 1_
- $a Šlampová, Andrea $u Institute of Analytical Chemistry of the Czech Academy of Sciences, v. v. i., Veveří 97, CZ-60200 Brno, Czech Republic.
- 245 10
- $a Injections from sub-μL sample volumes in commercial capillary electrophoresis / $c A. Šlampová, P. Kubáň,
- 520 9_
- $a A simple sample injection procedure compatible with commercial capillary electrophoresis (CE) instrumentation was developed, which enables handling sample volumes as little as 250nL for analytical applications where sample volume availability is of concern. Single-use micro-sampling inserts were prepared by thermal modification of polypropylene micropipette tips and the inserts were accommodated in standard CE vials in CE autosampler carousel. To ensure direct contact of separation capillary injection end with sample solution and to avoid possible damage to the capillary, a soft compression spring was placed at the bottom of the vial underneath the micro-sampling insert. Injections from sub-μL samples were carried out in conventional as well as in short-end injection mode, were compatible with standard i.d./o.d. (25-100μm/365μm) fused silica capillaries and with various background electrolyte solutions and detection modes. Excellent repeatability of replicate injections from 250nL to 3μL was achieved based on RSD values of quantitative analytical measures (peak heights ≤2.4% and peak areas ≤3.7%) for CE-UV-vis, CE-ESI-MS and CE-contactless conductivity detection of model basic drugs. The achieved RSD values were comparable with those for replicate injections of the drugs from standard CE vials. The reported concept of injections from micro-sampling inserts was further demonstrated useful in evaluation of micro-electromembrane extraction (μ-EME) of model basic drugs. Sub-μL volumes of operational solutions resulted in reduced lengths of μ-EME phases and improved extraction recoveries (66-91%) were achieved.
- 650 _2
- $a elektrolyty $x chemie $7 D004573
- 650 _2
- $a elektroforéza kapilární $x přístrojové vybavení $x metody $7 D019075
- 650 _2
- $a hmotnostní spektrometrie $7 D013058
- 650 _2
- $a nortriptylin $x analýza $x izolace a purifikace $7 D009661
- 650 _2
- $a papaverin $x analýza $7 D010208
- 650 _2
- $a roztoky $x chemie $7 D012996
- 650 _2
- $a spektrofotometrie ultrafialová $7 D013056
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Kubáň, Pavel $u Institute of Analytical Chemistry of the Czech Academy of Sciences, v. v. i., Veveří 97, CZ-60200 Brno, Czech Republic. Electronic address: kuban@iach.cz.
- 773 0_
- $w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1497, č. - (2017), s. 164-171
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/28381364 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20170720 $b ABA008
- 991 __
- $a 20170720123620 $b ABA008
- 999 __
- $a ok $b bmc $g 1238836 $s 984068
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2017 $b 1497 $c - $d 164-171 $e 20170327 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
- LZP __
- $a Pubmed-20170720