Optimal reconstruction matrix and PET image filtration for point-spread function and time-of-flight reconstruction - A phantom study
J. Ptáček, P. Karhan, P. Fiala,
Jazyk angličtina Země Itálie
Typ dokumentu časopisecké články
- MeSH
- fantomy radiodiagnostické * MeSH
- lidé MeSH
- počítačové zpracování obrazu * MeSH
- pozitronová emisní tomografie * MeSH
- reprodukovatelnost výsledků MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
INTRODUCTION: The aim of this study was to determine the optimal image matrix and half-width of the Gaussian filter after iterative reconstruction of the PET image with point-spread function (PSF) and time-of-flight (TOF) correction, based on measuring the recovery coefficient (RC) curves. The measured RC curves were compared to those from an older system which does not use PSF and TOF corrections. MATERIALS AND METHODS: The measurements were carried out on a NEMA IEC Body Phantom. We measured the RC curves based on SUVmaxand SUVA50in source spheres with different diameters. The change in noise level for different reconstruction parameter settings and the relation between RC curves and the administered activity were also evaluated. RESULTS: With an increasing size of image matrix and reduction in the half-width of the post-reconstruction Gaussian filter, there was a significant increase in image noise and overestimation of the SUV. The local increase in SUV, observed for certain filtrations and objects with a diameter below 13mm, was caused by PSF correction. The decrease in administered activity, while maintaining the same conditions of acquisition and reconstruction, also led to overestimation of readings of the SUV and additionally to deterioration in reproducibility. CONCLUSION: This study proposes a suitable size for the image matrix and filtering for displaying PET and SUV measurements. The benefits were demonstrated as improved image parameters for the newer instrument, these even being found using relatively strong filtration of the reconstructed images.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc18010501
- 003
- CZ-PrNML
- 005
- 20180419144053.0
- 007
- ta
- 008
- 180404s2017 it f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.ejmp.2017.06.002 $2 doi
- 035 __
- $a (PubMed)28601381
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a it
- 100 1_
- $a Ptáček, Jaroslav $u Department of Medical Physics and Radiation Protection, University Hospital Olomouc, Olomouc, Czech Republic; Clinic of Nuclear Medicine, Faculty of Medicine and Dentistry, Palacký University Olomouc, Olomouc, Czech Republic. Electronic address: jaroslav.ptacek@fnol.cz.
- 245 10
- $a Optimal reconstruction matrix and PET image filtration for point-spread function and time-of-flight reconstruction - A phantom study / $c J. Ptáček, P. Karhan, P. Fiala,
- 520 9_
- $a INTRODUCTION: The aim of this study was to determine the optimal image matrix and half-width of the Gaussian filter after iterative reconstruction of the PET image with point-spread function (PSF) and time-of-flight (TOF) correction, based on measuring the recovery coefficient (RC) curves. The measured RC curves were compared to those from an older system which does not use PSF and TOF corrections. MATERIALS AND METHODS: The measurements were carried out on a NEMA IEC Body Phantom. We measured the RC curves based on SUVmaxand SUVA50in source spheres with different diameters. The change in noise level for different reconstruction parameter settings and the relation between RC curves and the administered activity were also evaluated. RESULTS: With an increasing size of image matrix and reduction in the half-width of the post-reconstruction Gaussian filter, there was a significant increase in image noise and overestimation of the SUV. The local increase in SUV, observed for certain filtrations and objects with a diameter below 13mm, was caused by PSF correction. The decrease in administered activity, while maintaining the same conditions of acquisition and reconstruction, also led to overestimation of readings of the SUV and additionally to deterioration in reproducibility. CONCLUSION: This study proposes a suitable size for the image matrix and filtering for displaying PET and SUV measurements. The benefits were demonstrated as improved image parameters for the newer instrument, these even being found using relatively strong filtration of the reconstructed images.
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a počítačové zpracování obrazu $7 D007091
- 650 12
- $a fantomy radiodiagnostické $7 D019047
- 650 12
- $a pozitronová emisní tomografie $7 D049268
- 650 _2
- $a reprodukovatelnost výsledků $7 D015203
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Karhan, Pavel $u Department of Medical Physics and Radiation Protection, University Hospital Olomouc, Olomouc, Czech Republic; Clinic of Nuclear Medicine, Faculty of Medicine and Dentistry, Palacký University Olomouc, Olomouc, Czech Republic.
- 700 1_
- $a Fiala, Petr $u Department of Medical Physics and Radiation Protection, University Hospital Olomouc, Olomouc, Czech Republic.
- 773 0_
- $w MED00167391 $t Physica medica PM an international journal devoted to the applications of physics to medicine and biology official journal of the Italian Association of Biomedical Physics (AIFB) $x 1724-191X $g Roč. 39, č. - (2017), s. 95-99
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/28601381 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20180404 $b ABA008
- 991 __
- $a 20180419144155 $b ABA008
- 999 __
- $a ok $b bmc $g 1287986 $s 1007313
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2017 $b 39 $c - $d 95-99 $e 20170607 $i 1724-191X $m Physica medica $n Phys Med $x MED00167391
- LZP __
- $a Pubmed-20180404