Je něco špatně v tomto záznamu ?
Micro-SPE in pipette tips as a tool for analysis of small-molecule drugs in serum
H. Vlčková, V. Pilařová, O. Novák, P. Solich, L. Nováková,
Jazyk angličtina Země Velká Británie
Typ dokumentu časopisecké články, validační studie
PubMed
28617093
DOI
10.4155/bio-2017-0033
Knihovny.cz E-zdroje
- MeSH
- design vybavení MeSH
- léčivé přípravky krev MeSH
- lidé MeSH
- limita detekce MeSH
- mikroextrakce na pevné fázi přístrojové vybavení metody MeSH
- tandemová hmotnostní spektrometrie metody MeSH
- vysokoúčinná kapalinová chromatografie metody MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- validační studie MeSH
AIM: Micro-SPE in pipette tips (μ-SPE-PT) with particle sorbent has never been used in small-molecule drug analysis. Methodology & results: μ-SPE-PT was used for the extraction of statins from biological materials followed by UHPLC-MS/MS. The commercial and homemade μ-SPE-PT tips filled with particle sorbent were compared. While the homemade tips enabled direct serum sample loading into the sorbent, protein precipitation (PP) had to be implemented before μ-SPE-PT procedure using commercial tips. Three μ-SPE-PT methods were developed and validated: method A: μ-SPE-PT with homemade tips; method B: PP + μ-SPE-PT with homemade tips; and method C: PP + μ-SPE-PT with commercial tips. Method A enabled a simple high-throughput approach (48 samples in 90 min) compared with methods B and C that required three-times longer time. However, PP increased the recoveries of protein-bound analytes and extracts purity in methods B and C. The matrix effects without internal standards correction for method C were significantly higher than those for the methods A and B. CONCLUSION: Compared with commercial tips, homemade tips filled with particles were found to be more suitable for drug analysis. Commercial tips tested in this study were found challenging but the conditions under which they could be applicable were also defined.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc18016545
- 003
- CZ-PrNML
- 005
- 20180521101151.0
- 007
- ta
- 008
- 180515s2017 xxk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.4155/bio-2017-0033 $2 doi
- 035 __
- $a (PubMed)28617093
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxk
- 100 1_
- $a Vlčková, Hana $u Department of Analytical Chemistry, Faculty of Pharmacy, Charles University, Heyrovského 1203, 50005 Hradec Králové, Czech Republic.
- 245 10
- $a Micro-SPE in pipette tips as a tool for analysis of small-molecule drugs in serum / $c H. Vlčková, V. Pilařová, O. Novák, P. Solich, L. Nováková,
- 520 9_
- $a AIM: Micro-SPE in pipette tips (μ-SPE-PT) with particle sorbent has never been used in small-molecule drug analysis. Methodology & results: μ-SPE-PT was used for the extraction of statins from biological materials followed by UHPLC-MS/MS. The commercial and homemade μ-SPE-PT tips filled with particle sorbent were compared. While the homemade tips enabled direct serum sample loading into the sorbent, protein precipitation (PP) had to be implemented before μ-SPE-PT procedure using commercial tips. Three μ-SPE-PT methods were developed and validated: method A: μ-SPE-PT with homemade tips; method B: PP + μ-SPE-PT with homemade tips; and method C: PP + μ-SPE-PT with commercial tips. Method A enabled a simple high-throughput approach (48 samples in 90 min) compared with methods B and C that required three-times longer time. However, PP increased the recoveries of protein-bound analytes and extracts purity in methods B and C. The matrix effects without internal standards correction for method C were significantly higher than those for the methods A and B. CONCLUSION: Compared with commercial tips, homemade tips filled with particles were found to be more suitable for drug analysis. Commercial tips tested in this study were found challenging but the conditions under which they could be applicable were also defined.
- 650 _2
- $a vysokoúčinná kapalinová chromatografie $x metody $7 D002851
- 650 _2
- $a design vybavení $7 D004867
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a limita detekce $7 D057230
- 650 _2
- $a léčivé přípravky $x krev $7 D004364
- 650 _2
- $a mikroextrakce na pevné fázi $x přístrojové vybavení $x metody $7 D052617
- 650 _2
- $a tandemová hmotnostní spektrometrie $x metody $7 D053719
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a validační studie $7 D023361
- 700 1_
- $a Pilařová, Veronika $u Department of Analytical Chemistry, Faculty of Pharmacy, Charles University, Heyrovského 1203, 50005 Hradec Králové, Czech Republic.
- 700 1_
- $a Novák, Ondřej $u Laboratory of Growth Regulators, Centre of the Region Hana for Biotechnological & Agricultural Research, Faculty of Science, Palacký University & Institute of Experimental Botany AS CR, Šlechtitelů 11, 783 71 Olomouc, Czech Republic.
- 700 1_
- $a Solich, Petr $u Department of Analytical Chemistry, Faculty of Pharmacy, Charles University, Heyrovského 1203, 50005 Hradec Králové, Czech Republic.
- 700 1_
- $a Nováková, Lucie $u Department of Analytical Chemistry, Faculty of Pharmacy, Charles University, Heyrovského 1203, 50005 Hradec Králové, Czech Republic.
- 773 0_
- $w MED00184533 $t Bioanalysis $x 1757-6199 $g Roč. 9, č. 11 (2017), s. 887-901
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/28617093 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20180515 $b ABA008
- 991 __
- $a 20180521101332 $b ABA008
- 999 __
- $a ok $b bmc $g 1300169 $s 1013385
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2017 $b 9 $c 11 $d 887-901 $e 20170615 $i 1757-6199 $m Bioanalysis $n Bioanal. $x MED00184533
- LZP __
- $a Pubmed-20180515