A framework for alignment of on-board imagers of medical linear accelerators
M. Tulik, D. Kabat, RA. Kycia, Z. Tabor, A. Woszczyna, Z. Latała,
Language English Country Italy
Document type Journal Article
- MeSH
- Algorithms MeSH
- Particle Accelerators * MeSH
- Diagnostic Imaging instrumentation MeSH
- Calibration MeSH
- Image Processing, Computer-Assisted MeSH
- Radiotherapy, Intensity-Modulated MeSH
- Publication type
- Journal Article MeSH
PURPOSE: To develop an assumption-free methodology for aligning the geometry of on-board imagers with the geometry of medical linear accelerators applied in image-guided radiotherapy (IGRT). MATERIAL: Alignment of the on-board imaging (OBI) system with respect to the accelerator system is achieved using a multi-modular phantom described elsewhere (Tabor et al., 2017), enabling the geometry of the linear accelerator to be specified without any pre-assumptions. METHODS: The placement of two isocentres (of the on-board imager and of the therapeutic system) and of three frames of reference (those of the on-board imager, of the therapeutic system, and of the treatment table) is formulated as an optimization problem. It is solved by analysing the images of fiducial points positioned in 3D space in phantom modules attached to the treatment table and to the collimator of the accelerator. Fiducials are projected onto an imaging plane of unknown characteristics from a virtual source of unknown coordinates. CONCLUSIONS: An analytical framework exploiting projection images of the proposed multi-modular phantom has been developed, enabling precise alignment of the reference frame related to the on-board imager with the reference frame related to the therapeutic system. Within the proposed framework, the necessary corrections of treatment table positioning prior to patient irradiation, are delivered in the treatment table coordinates.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc18033198
- 003
- CZ-PrNML
- 005
- 20181012102136.0
- 007
- ta
- 008
- 181008s2018 it f 000 0|0eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.ejmp.2018.02.017 $2 doi
- 035 __
- $a (PubMed)29609822
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a it
- 100 1_
- $a Tulik, M $u Maria Sklodowska-Curie Memorial Cancer Center and Institute of Oncology Cracow Branch, ul.Garncarska 11, 31-115 Kraków, Poland. Electronic address: mdtulik@gmail.com.
- 245 12
- $a A framework for alignment of on-board imagers of medical linear accelerators / $c M. Tulik, D. Kabat, RA. Kycia, Z. Tabor, A. Woszczyna, Z. Latała,
- 520 9_
- $a PURPOSE: To develop an assumption-free methodology for aligning the geometry of on-board imagers with the geometry of medical linear accelerators applied in image-guided radiotherapy (IGRT). MATERIAL: Alignment of the on-board imaging (OBI) system with respect to the accelerator system is achieved using a multi-modular phantom described elsewhere (Tabor et al., 2017), enabling the geometry of the linear accelerator to be specified without any pre-assumptions. METHODS: The placement of two isocentres (of the on-board imager and of the therapeutic system) and of three frames of reference (those of the on-board imager, of the therapeutic system, and of the treatment table) is formulated as an optimization problem. It is solved by analysing the images of fiducial points positioned in 3D space in phantom modules attached to the treatment table and to the collimator of the accelerator. Fiducials are projected onto an imaging plane of unknown characteristics from a virtual source of unknown coordinates. CONCLUSIONS: An analytical framework exploiting projection images of the proposed multi-modular phantom has been developed, enabling precise alignment of the reference frame related to the on-board imager with the reference frame related to the therapeutic system. Within the proposed framework, the necessary corrections of treatment table positioning prior to patient irradiation, are delivered in the treatment table coordinates.
- 650 _2
- $a algoritmy $7 D000465
- 650 _2
- $a kalibrace $7 D002138
- 650 _2
- $a diagnostické zobrazování $x přístrojové vybavení $7 D003952
- 650 _2
- $a počítačové zpracování obrazu $7 D007091
- 650 12
- $a částice - urychlovače $7 D010315
- 650 _2
- $a radioterapie s modulovanou intenzitou $7 D050397
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Kabat, D $u Maria Sklodowska-Curie Memorial Cancer Center and Institute of Oncology Cracow Branch, ul.Garncarska 11, 31-115 Kraków, Poland.
- 700 1_
- $a Kycia, R A $u Cracow University of Technology, Faculty of Physics, Mathematics and Computer Science, ul. Warszawska 24, 31-155 Kraków, Poland; Department of Mathematics and Statistics, Masaryk University, Brno, Czech Republic.
- 700 1_
- $a Tabor, Z $u Cracow University of Technology, Faculty of Physics, Mathematics and Computer Science, ul. Warszawska 24, 31-155 Kraków, Poland.
- 700 1_
- $a Woszczyna, A $u Cracow University of Technology, Faculty of Physics, Mathematics and Computer Science, ul. Warszawska 24, 31-155 Kraków, Poland.
- 700 1_
- $a Latała, Z $u Cracow University of Technology, Faculty of Mechanical Engineering, al. Jana Pawła II 37, 30-001 Kraków, Poland.
- 773 0_
- $w MED00167391 $t Physica medica PM an international journal devoted to the applications of physics to medicine and biology official journal of the Italian Association of Biomedical Physics (AIFB) $x 1724-191X $g Roč. 47, č. - (2018), s. 80-85
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/29609822 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20181008 $b ABA008
- 991 __
- $a 20181012102628 $b ABA008
- 999 __
- $a ok $b bmc $g 1340842 $s 1030192
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2018 $b 47 $c - $d 80-85 $e 20180305 $i 1724-191X $m Physica medica $n Phys Med $x MED00167391
- LZP __
- $a Pubmed-20181008