On-line monitoring of in-vitro oral bioaccessibility tests as front-end to liquid chromatography for determination of chlorogenic acid isomers in dietary supplements

D. Kremr, DJ. Cocovi-Solberg, P. Bajerová, K. Ventura, M. Miró,

. 2017 ; 166 (-) : 391-398. [pub] 20160101

Language English Country Netherlands

Document type Journal Article

A novel fully automated in-vitro oral dissolution test assay as a front-end to liquid chromatography has been developed and validated for on-line chemical profiling and monitoring of temporal release profiles of three caffeoylquinic acid (CQA) isomers, namely, 3-CQA,4-CQA and 5-CQA, known as chlorogenic acids, in dietary supplements. Tangential-flow filtration is harnessed as a sample processing approach for on-line handling of CQA containing extracts of hard gelatin capsules and introduction of protein-free samples into the liquid chromatograph. Oral bioaccessibility/dissolution test assays were performed at 37.0±0.5°C as per US Pharmacopeia recommendations using pepsin with activity of ca. 749,000 USP units/L in 0.1mol/L HCl as the extraction medium and a paddle apparatus stirred at 50rpm. CQA release rates and steady-state dissolution conditions were determined accurately by fitting the chromatographic datasets, namely, the average cumulative concentrations of bioaccessible pools of every individual isomer monitored during 200min, with temporal resolutions of ≥10min, to a first-order dissolution kinetic model. Distinct solid-to-liquid phase ratios in the mimicry of physiological extraction conditions were assessed. Relative standard deviations for intra-day repeatability and inter-day intermediate precision of 5-CQA within the 5-40µg/mL concentration range were <3.4% and <5.5%, respectively. Trueness of the automatic flow method for determination of 5-CQA released from dietary supplements in gastric fluid surrogate was demonstrated by spike recoveries, spanning from 91.5-104.0%, upon completion of the dissolution process. The proposed hyphenated setup was resorted for evaluating potential differences in dissolution profiles and content of the three most abundant chlorogenic acid isomers in dietary supplements from varied manufacturers.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc18033992
003      
CZ-PrNML
005      
20181008122222.0
007      
ta
008      
181008s2017 ne f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.talanta.2015.12.082 $2 doi
035    __
$a (PubMed)28213251
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a ne
100    1_
$a Kremr, Daniel $u University of Pardubice, Faculty of Chemical Technology, Department of Analytical Chemistry, Studentská 573, 532 10 Pardubice, Czech Republic.
245    10
$a On-line monitoring of in-vitro oral bioaccessibility tests as front-end to liquid chromatography for determination of chlorogenic acid isomers in dietary supplements / $c D. Kremr, DJ. Cocovi-Solberg, P. Bajerová, K. Ventura, M. Miró,
520    9_
$a A novel fully automated in-vitro oral dissolution test assay as a front-end to liquid chromatography has been developed and validated for on-line chemical profiling and monitoring of temporal release profiles of three caffeoylquinic acid (CQA) isomers, namely, 3-CQA,4-CQA and 5-CQA, known as chlorogenic acids, in dietary supplements. Tangential-flow filtration is harnessed as a sample processing approach for on-line handling of CQA containing extracts of hard gelatin capsules and introduction of protein-free samples into the liquid chromatograph. Oral bioaccessibility/dissolution test assays were performed at 37.0±0.5°C as per US Pharmacopeia recommendations using pepsin with activity of ca. 749,000 USP units/L in 0.1mol/L HCl as the extraction medium and a paddle apparatus stirred at 50rpm. CQA release rates and steady-state dissolution conditions were determined accurately by fitting the chromatographic datasets, namely, the average cumulative concentrations of bioaccessible pools of every individual isomer monitored during 200min, with temporal resolutions of ≥10min, to a first-order dissolution kinetic model. Distinct solid-to-liquid phase ratios in the mimicry of physiological extraction conditions were assessed. Relative standard deviations for intra-day repeatability and inter-day intermediate precision of 5-CQA within the 5-40µg/mL concentration range were <3.4% and <5.5%, respectively. Trueness of the automatic flow method for determination of 5-CQA released from dietary supplements in gastric fluid surrogate was demonstrated by spike recoveries, spanning from 91.5-104.0%, upon completion of the dissolution process. The proposed hyphenated setup was resorted for evaluating potential differences in dissolution profiles and content of the three most abundant chlorogenic acid isomers in dietary supplements from varied manufacturers.
650    _2
$a aplikace orální $7 D000284
650    _2
$a biologická dostupnost $7 D001682
650    _2
$a kyselina chlorogenová $x aplikace a dávkování $x analýza $x farmakokinetika $7 D002726
650    _2
$a chromatografie kapalinová $x metody $7 D002853
650    _2
$a káva $x chemie $7 D003069
650    _2
$a potravní doplňky $x analýza $7 D019587
650    _2
$a isomerie $7 D007536
650    _2
$a limita detekce $7 D057230
650    _2
$a on-line systémy $7 D009862
655    _2
$a časopisecké články $7 D016428
700    1_
$a Cocovi-Solberg, David J $u FI-TRACE group, University of the Balearic Islands, Department of Chemistry, Carretera de Valldemossa km 7.5, E-07122 Palma de Mallorca, Spain.
700    1_
$a Bajerová, Petra $u University of Pardubice, Faculty of Chemical Technology, Department of Analytical Chemistry, Studentská 573, 532 10 Pardubice, Czech Republic.
700    1_
$a Ventura, Karel $u University of Pardubice, Faculty of Chemical Technology, Department of Analytical Chemistry, Studentská 573, 532 10 Pardubice, Czech Republic.
700    1_
$a Miró, Manuel $u FI-TRACE group, University of the Balearic Islands, Department of Chemistry, Carretera de Valldemossa km 7.5, E-07122 Palma de Mallorca, Spain. Electronic address: manuel.miro@uib.es.
773    0_
$w MED00004484 $t Talanta $x 1873-3573 $g Roč. 166, č. - (2017), s. 391-398
856    41
$u https://pubmed.ncbi.nlm.nih.gov/28213251 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20181008 $b ABA008
991    __
$a 20181008122708 $b ABA008
999    __
$a ok $b bmc $g 1339680 $s 1030986
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2017 $b 166 $c - $d 391-398 $e 20160101 $i 1873-3573 $m Talanta $n Talanta $x MED00004484
LZP    __
$a Pubmed-20181008

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...