Use of EX-PRESS® implant in glaucoma surgery - retrospective study
Fialová V., Váša M.
Jazyk angličtina Země Česko
- Klíčová slova
- EX-PRESS implantát,
- MeSH
- drenážní implantáty glaukomové MeSH
- filtrující operace * metody MeSH
- glaukom s otevřeným úhlem * chirurgie terapie MeSH
- glaukom chirurgie terapie MeSH
- lidé středního věku MeSH
- lidé MeSH
- nitrooční tlak MeSH
- oftalmologické chirurgické výkony metody přístrojové vybavení škodlivé účinky MeSH
- retrospektivní studie MeSH
- senioři nad 80 let MeSH
- senioři MeSH
- statistika jako téma MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři nad 80 let MeSH
- senioři MeSH
- ženské pohlaví MeSH
Purpose: The objective of the study is to evaluate of primary open-angle glaucoma (POAG) with the use of EX-PRESS drainage implant. Evaluated was the decrease of intraocular pressure (IOP), visus stabilization, perimeter Humphrey - related finding (T 30-2), Heidelberg Retina Tomograph (HRT) and possibility of reducing the local drug therapy. Patients and methods: Retrospective data analysis was performed in 40 eyes with POAG in 28 subjects (14 female- and 14 male patients) average-aged 69.5 years. In all 40 eyes, surgery was performed by one surgeon within the years 2011-2017. Indications for EX-PRESS implantation in our study were the POAG with decompensated IOP, decompensated chronic secondary glaucoma or failure previously anti-glaucoma surgical operations. Within the preoperative period, in all cases the progression was observed on the perimeter T 30-2 or on the HRT. Before and after surgery, all 40 eyes were evaluated for the following factors: the IOP, visus, pachymetry, therapy by anti-glaucomatic drugs and regular inspections of the perimeter and HRT. The average post-surgery following-up time in our total of patients was 3 years and 8 months. The identified data of our total of patients were statistically processed using the Wilcoxon Signed-Rank Test. Results: The average pre-surgery IOP was 21.4 mm Hg, 6 months post-surgery 11.2 mm Hg, and 13.2 mm Hg at the last inspection. Pre-surgery anti-glaucoma therapy: monotherapy in 2 eyes, dual therapy in 25 eyes, triple therapy in 13 eyes. Post-operatively without the need of therapy were 19 eyes, the need for monotherapy in 7 eyes, dual therapy in 13-eyes and triple therapy in 1 eye. During the last inspection of the perimeter or HRT within the postoperative period, we identified stationary finding in 39 eyes and only a mild progression in 1 eye. Peroperatively, we have not identified any serious complications. Within the postoperative period, we observed choroid ablation in 10 eyes as recovered within 6-7 days. As a more serious complication, we noted endophthalmitis in 1 eye on the background of generalized lichen planus complicating skin disease. In 5 eyes within the late postoperative period we found occlusion of lateral orifices in the EX-PRESS implant by the fibrotic tissue, the 4 eyes developed cataracts, and the eyeball hypotonia persists in 1 eye at the level of 5 torr without affecting the visus of the operated eye. Conclusion: It outflows from the above results that the use of EX-PRESS implant in the surgery of glaucoma is an effective and safe method with a minimal number of complications.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc19000027
- 003
- CZ-PrNML
- 005
- 20210601023019.0
- 007
- cr|cn|
- 008
- 190104s2018 xr acd fs 000 0|eng||
- 009
- AR
- 024 0_
- $a 10.31348/2018/1/2-1-2018 $2 doi
- 040 __
- $a ABA008 $d ABA008 $e AACR2 $b cze
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Fialová, Veronika $7 xx0230999 $u Department of Ophthalmology, University Hospital Hradec Králové
- 245 10
- $a Use of EX-PRESS® implant in glaucoma surgery - retrospective study / $c Fialová V., Váša M.
- 520 3_
- $a Purpose: The objective of the study is to evaluate of primary open-angle glaucoma (POAG) with the use of EX-PRESS drainage implant. Evaluated was the decrease of intraocular pressure (IOP), visus stabilization, perimeter Humphrey - related finding (T 30-2), Heidelberg Retina Tomograph (HRT) and possibility of reducing the local drug therapy. Patients and methods: Retrospective data analysis was performed in 40 eyes with POAG in 28 subjects (14 female- and 14 male patients) average-aged 69.5 years. In all 40 eyes, surgery was performed by one surgeon within the years 2011-2017. Indications for EX-PRESS implantation in our study were the POAG with decompensated IOP, decompensated chronic secondary glaucoma or failure previously anti-glaucoma surgical operations. Within the preoperative period, in all cases the progression was observed on the perimeter T 30-2 or on the HRT. Before and after surgery, all 40 eyes were evaluated for the following factors: the IOP, visus, pachymetry, therapy by anti-glaucomatic drugs and regular inspections of the perimeter and HRT. The average post-surgery following-up time in our total of patients was 3 years and 8 months. The identified data of our total of patients were statistically processed using the Wilcoxon Signed-Rank Test. Results: The average pre-surgery IOP was 21.4 mm Hg, 6 months post-surgery 11.2 mm Hg, and 13.2 mm Hg at the last inspection. Pre-surgery anti-glaucoma therapy: monotherapy in 2 eyes, dual therapy in 25 eyes, triple therapy in 13 eyes. Post-operatively without the need of therapy were 19 eyes, the need for monotherapy in 7 eyes, dual therapy in 13-eyes and triple therapy in 1 eye. During the last inspection of the perimeter or HRT within the postoperative period, we identified stationary finding in 39 eyes and only a mild progression in 1 eye. Peroperatively, we have not identified any serious complications. Within the postoperative period, we observed choroid ablation in 10 eyes as recovered within 6-7 days. As a more serious complication, we noted endophthalmitis in 1 eye on the background of generalized lichen planus complicating skin disease. In 5 eyes within the late postoperative period we found occlusion of lateral orifices in the EX-PRESS implant by the fibrotic tissue, the 4 eyes developed cataracts, and the eyeball hypotonia persists in 1 eye at the level of 5 torr without affecting the visus of the operated eye. Conclusion: It outflows from the above results that the use of EX-PRESS implant in the surgery of glaucoma is an effective and safe method with a minimal number of complications.
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a lidé středního věku $7 D008875
- 650 _2
- $a mužské pohlaví $7 D008297
- 650 _2
- $a senioři $7 D000368
- 650 _2
- $a senioři nad 80 let $7 D000369
- 650 _2
- $a ženské pohlaví $7 D005260
- 650 12
- $a glaukom s otevřeným úhlem $x chirurgie $x terapie $7 D005902
- 650 12
- $a filtrující operace $x metody $7 D018463
- 650 _2
- $a drenážní implantáty glaukomové $7 D020327
- 650 _2
- $a retrospektivní studie $7 D012189
- 650 _2
- $a výsledek terapie $7 D016896
- 650 _2
- $a glaukom $x chirurgie $x terapie $7 D005901
- 650 _2
- $a oftalmologické chirurgické výkony $x metody $x přístrojové vybavení $x škodlivé účinky $7 D013508
- 650 _2
- $a nitrooční tlak $7 D007429
- 650 _2
- $a statistika jako téma $7 D013223
- 653 00
- $a EX-PRESS implantát
- 700 1_
- $a Váša, Martin. $7 xx0231013 $u VISUS spol. s. r. o., Náchod
- 773 0_
- $t Czech and Slovak Ophthalmology $g Roč. 5, č. 1 (2018), s. 9-14 $w MED00194413
- 856 41
- $u http://www.cs-ophthalmology.cz/en/journal/2018/1/2-1-2018 $y plný text volně přístupný
- 910 __
- $a ABA008 $b online $y p $z 0
- 990 __
- $a 20190104132740 $b ABA008
- 991 __
- $a 20210601023018 $b ABA008
- 999 __
- $a ok $b bmc $g 1363056 $s 1038139
- BAS __
- $a 3 $a 4
- BMC __
- $a 2018 $b 5 $c 1 $d 9-14 $m Czech and Slovak Ophthalmology $x MED00194413
- LZP __
- $c NLK189 $d 20210531 $a NLK 2019-01/pk