Detail
Article
Online article
FT
Medvik - BMC
  • Something wrong with this record ?

Respiratory arrhythmia as an encouragement of inappropriate ICD therapy

David Korpas, Jana Haluzikova, Marek Penhaker, Martin Augustynek

. 2016 ; 46 (3) : 69-73.

Language English Country Czech Republic

Document type Case Reports

Implantable cardioverter-defibrillators (ICDs) are active cardiac implants for immediate treatment of ventricular arrhythmias. They provide life-saving therapy, but may also deliver inappropriate therapy. These two cases demonstrate a possibility of ICD therapy detection or induction encouraged by physiological patient breathing. The cases of 64-year-old and 30-year-old women patients are reported. The first one has received some therapy, the second one has not. The devices records did not show any abnormalities in trends or electrical parameters. As a solution, the detection parameters of the devices were optimized, with reference to individual physiologically reachable heart rates of the patients.

Bibliography, etc.

Literatura

000      
00000naa a2200000 a 4500
001      
bmc19015226
003      
CZ-PrNML
005      
20210629233831.0
007      
ta
008      
190428s2016 xr d f 000 0|eng||
009      
AR
040    __
$a ABA008 $d ABA008 $e AACR2 $b cze
041    0_
$a eng
044    __
$a xr
100    1_
$a Korpas, David $7 xx0034427 $u Institute of Nursing, Faculty of Public Policies, Silesian University, Opava, Czech Republic
245    10
$a Respiratory arrhythmia as an encouragement of inappropriate ICD therapy / $c David Korpas, Jana Haluzikova, Marek Penhaker, Martin Augustynek
504    __
$a Literatura
520    9_
$a Implantable cardioverter-defibrillators (ICDs) are active cardiac implants for immediate treatment of ventricular arrhythmias. They provide life-saving therapy, but may also deliver inappropriate therapy. These two cases demonstrate a possibility of ICD therapy detection or induction encouraged by physiological patient breathing. The cases of 64-year-old and 30-year-old women patients are reported. The first one has received some therapy, the second one has not. The devices records did not show any abnormalities in trends or electrical parameters. As a solution, the detection parameters of the devices were optimized, with reference to individual physiologically reachable heart rates of the patients.
650    _2
$a dospělí $7 D000328
650    _2
$a ženské pohlaví $7 D005260
650    _2
$a lidé $7 D006801
650    12
$a elektrická defibrilace $x metody $x škodlivé účinky $7 D004554
650    12
$a defibrilátory implantabilní $x škodlivé účinky $7 D017147
650    _2
$a sinusová arytmie $x terapie $7 D001146
650    _2
$a srdeční arytmie $x terapie $7 D001145
650    _2
$a individualizovaná medicína $7 D057285
650    _2
$a dýchání $7 D012119
650    _2
$a srdeční frekvence $7 D006339
655    _2
$a kazuistiky $7 D002363
700    1_
$a Haluzíková, Jana $7 xx0064196 $u Institute of Nursing, Faculty of Public Policies, Silesian University, Opava, Czech Republic
700    1_
$a Penhaker, Marek, $d 1972- $7 mzk2004246505 $u Department of Cybernetics & Biomedical Engineerign, Faculty of Electrical Engineering & Computer Science, Technical University, Ostrava, Czech Republic
700    1_
$a Augustynek, Martin $7 ola2011639627 $u Department of Cybernetics & Biomedical Engineerign, Faculty of Electrical Engineering & Computer Science, Technical University, Ostrava, Czech Republic
773    0_
$t Lékař a technika $x 0301-5491 $g Roč. 46, č. 3 (2016), s. 69-73 $w MED00011033
856    41
$u https://ojs.cvut.cz/ojs/index.php/CTJ/article/view/4188/4040 $y plný text volně přístupný
910    __
$a ABA008 $b B 1367 $c 1071 b $y p $z 0
990    __
$a 20190428103001 $b ABA008
991    __
$a 20210629233827 $b ABA008
999    __
$a ok $b bmc $g 1398976 $s 1053704
BAS    __
$a 3
BMC    __
$a 2016 $b 46 $c 3 $d 69-73 $i 0301-5491 $m Lékař a technika $n Lék. tech. $x MED00011033
LZP    __
$c NLK183 $d 20210629 $a NLK 2019-15/dk