Activation of QA devices and phantom materials under clinical scanning proton beams-a gamma spectrometry study
T. Hanušová, K. Johnová, M. Navrátil, J. Valenta, L. Müller,
Jazyk angličtina Země Anglie, Velká Británie
Typ dokumentu časopisecké články, práce podpořená grantem
PubMed
29726411
DOI
10.1088/1361-6560/aac27f
Knihovny.cz E-zdroje
- MeSH
- fantomy radiodiagnostické MeSH
- lidé MeSH
- protonová terapie přístrojové vybavení metody MeSH
- protony * MeSH
- radiometrie přístrojové vybavení MeSH
- spektrometrie gama přístrojové vybavení metody MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
Activation of detectors and phantoms used for commissioning and quality assurance of clinical proton beams may lead to radiation protection issues. Good understanding of the activation nuclide vectors involved is necessary to assess radiation risk for the personnel working with these devices on a daily basis or to fulfill legal requirements regarding transport of radioactive material and its release to the public. 11 devices and material samples were irradiated with a 220 MeV proton pencil beam (PBS, Proton Therapy Center, Prague). This study focuses on devices manufactured by IBA Dosimetry GmbH: MatriXX PT, PPC05, Stingray, Zebra, Lynx, a Blue Phantom rail and samples of RW3, PMMA, titanium, copper and carbon fibre plastic. Monitor units (MU) were monitored during delivery. Gamma spectrometry was then performed for each item using a HPGe detector, with a focus on longer lived gamma emitting radionuclides. Activities were quantified for all found isotopes and compared to relevant legal limits for exemption and clearance of radioactive objects. Activation was found to be significant after long irradiation sessions, as done during commissioning of a proton therapy room. Some of the investigated devices may also cumulate activity in time, depending on the scenario of periodic irradiation in routine clinical practice. However, the levels of activity and resulting beta/gamma doses are more comparable to internationally recommended concentration limits for exemption than to dose limits for radiation workers. Results of this study will help to determine nuclide inventories required by some legal authorities for radiation protection purposes.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc19028445
- 003
- CZ-PrNML
- 005
- 20190822091851.0
- 007
- ta
- 008
- 190813s2018 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1088/1361-6560/aac27f $2 doi
- 035 __
- $a (PubMed)29726411
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Hanušová, Tereza $u Faculty of Nuclear Sciences and Physical Engineering, Department of Dosimetry and Application of Ionizing Radiation, Czech Technical University in Prague, Břehová 7, 115 19 Praha 1, Czechia.
- 245 10
- $a Activation of QA devices and phantom materials under clinical scanning proton beams-a gamma spectrometry study / $c T. Hanušová, K. Johnová, M. Navrátil, J. Valenta, L. Müller,
- 520 9_
- $a Activation of detectors and phantoms used for commissioning and quality assurance of clinical proton beams may lead to radiation protection issues. Good understanding of the activation nuclide vectors involved is necessary to assess radiation risk for the personnel working with these devices on a daily basis or to fulfill legal requirements regarding transport of radioactive material and its release to the public. 11 devices and material samples were irradiated with a 220 MeV proton pencil beam (PBS, Proton Therapy Center, Prague). This study focuses on devices manufactured by IBA Dosimetry GmbH: MatriXX PT, PPC05, Stingray, Zebra, Lynx, a Blue Phantom rail and samples of RW3, PMMA, titanium, copper and carbon fibre plastic. Monitor units (MU) were monitored during delivery. Gamma spectrometry was then performed for each item using a HPGe detector, with a focus on longer lived gamma emitting radionuclides. Activities were quantified for all found isotopes and compared to relevant legal limits for exemption and clearance of radioactive objects. Activation was found to be significant after long irradiation sessions, as done during commissioning of a proton therapy room. Some of the investigated devices may also cumulate activity in time, depending on the scenario of periodic irradiation in routine clinical practice. However, the levels of activity and resulting beta/gamma doses are more comparable to internationally recommended concentration limits for exemption than to dose limits for radiation workers. Results of this study will help to determine nuclide inventories required by some legal authorities for radiation protection purposes.
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a fantomy radiodiagnostické $7 D019047
- 650 _2
- $a protonová terapie $x přístrojové vybavení $x metody $7 D061766
- 650 12
- $a protony $7 D011522
- 650 _2
- $a radiometrie $x přístrojové vybavení $7 D011874
- 650 _2
- $a spektrometrie gama $x přístrojové vybavení $x metody $7 D013051
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Johnová, Kamila
- 700 1_
- $a Navrátil, Matěj
- 700 1_
- $a Valenta, Jiří
- 700 1_
- $a Müller, Lutz
- 773 0_
- $w MED00003821 $t Physics in medicine and biology $x 1361-6560 $g Roč. 63, č. 11 (2018), s. 115014
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/29726411 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20190813 $b ABA008
- 991 __
- $a 20190822092129 $b ABA008
- 999 __
- $a ok $b bmc $g 1433594 $s 1066905
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2018 $b 63 $c 11 $d 115014 $e 20180607 $i 1361-6560 $m Physics in medicine and biology $n Phys Med Biol $x MED00003821
- LZP __
- $a Pubmed-20190813