DETERMINATION OF TOLERANCE LEVELS IN RADIOTHERAPY DOSIMETRY BASED ON STATISTICAL INTERFERENCE

I. Koniarová, L. Kotík,

. 2019 ; 186 (2-3) : 367-372. [pub] 2019Dec31

Language English Country Great Britain

Document type Journal Article

The most important dosimetry quantity that is determined at radiotherapy centers is the absorbed dose to water for external beams. Fixed tolerances for absorbed doses measured under reference conditions with an ionization chamber for high-energy photon and electron beams are usually 2 and 3%, respectively, regardless of uncertainties of the input variables and other conditions during evaluation. In reality, this agreement should be evaluated considering the uncertainties of the input variables because they affect the size of the random deviations of the measurements from their true values. The aim of this work was to develop a new approach to evaluate the agreement between measured and reported values based on statistical interference rather than to use fixed tolerance levels. The proposed method considers different scenarios that can occur during the evaluation of agreement. Because the method is described in general, it can be used in all similar situations when partial uncertainties can be established.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc20025363
003      
CZ-PrNML
005      
20201222153844.0
007      
ta
008      
201125s2019 xxk f 000 0|eng||
009      
AR
024    7_
$a 10.1093/rpd/ncz233 $2 doi
035    __
$a (PubMed)31712816
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xxk
100    1_
$a Koniarová, Irena $u National Radiation Protection Institute, Bartoškova 28, 14000 Prague, Czech Republic.
245    10
$a DETERMINATION OF TOLERANCE LEVELS IN RADIOTHERAPY DOSIMETRY BASED ON STATISTICAL INTERFERENCE / $c I. Koniarová, L. Kotík,
520    9_
$a The most important dosimetry quantity that is determined at radiotherapy centers is the absorbed dose to water for external beams. Fixed tolerances for absorbed doses measured under reference conditions with an ionization chamber for high-energy photon and electron beams are usually 2 and 3%, respectively, regardless of uncertainties of the input variables and other conditions during evaluation. In reality, this agreement should be evaluated considering the uncertainties of the input variables because they affect the size of the random deviations of the measurements from their true values. The aim of this work was to develop a new approach to evaluate the agreement between measured and reported values based on statistical interference rather than to use fixed tolerance levels. The proposed method considers different scenarios that can occur during the evaluation of agreement. Because the method is described in general, it can be used in all similar situations when partial uncertainties can be established.
650    _2
$a kalibrace $7 D002138
650    _2
$a lidé $7 D006801
650    12
$a statistické modely $7 D015233
650    _2
$a nádory $x radioterapie $7 D009369
650    _2
$a celková dávka radioterapie $7 D011879
650    _2
$a plánování radioterapie pomocí počítače $x metody $x normy $7 D011880
650    _2
$a vysokoenergetická radioterapie $x normy $7 D011882
650    _2
$a radioterapie s modulovanou intenzitou $x metody $x normy $7 D050397
650    _2
$a nejistota $7 D035501
655    _2
$a časopisecké články $7 D016428
700    1_
$a Kotík, Lukáš $u National Radiation Protection Institute, Bartoškova 28, 14000 Prague, Czech Republic.
773    0_
$w MED00004029 $t Radiation protection dosimetry $x 1742-3406 $g Roč. 186, č. 2-3 (2019), s. 367-372
856    41
$u https://pubmed.ncbi.nlm.nih.gov/31712816 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20201125 $b ABA008
991    __
$a 20201222153840 $b ABA008
999    __
$a ok $b bmc $g 1599508 $s 1116049
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2019 $b 186 $c 2-3 $d 367-372 $e 2019Dec31 $i 1742-3406 $m Radiation protection dosimetry $n Radiat Prot Dosimetry $x MED00004029