Evaluation of Important Analytical Parameters of the Peptest Immunoassay that Limit its Use in Diagnosing Gastroesophageal Reflux Disease

J. Dolina, Š. Konečný, P. Ďurč, J. Lačná, M. Greguš, F. Foret, J. Skřičková, M. Doubková, D. Kindlová, E. Pokojová, P. Kubáň,

. 2019 ; 53 (5) : 355-360. [pub] -

Language English Country United States

Document type Evaluation Study, Journal Article

Grant support
NV17-31945A MZ0 CEP Register

GOAL: To evaluate the analytical parameters of a lateral flow (LF) pepsin immunoassay (Peptest) and assess its suitability in the diagnostics of gastroesophageal reflux disease (GERD). BACKGROUND: Peptest is a noninvasive assay to analyze pepsin in saliva, intended for use in GERD diagnostics. Although commercialized, fundamental studies on its performance are missing. The assay therefore requires basic analytical parameter evaluation to assess its suitability in clinical practice. STUDY: Assay reaction's time dependence, reader device repeatability, and individual LF devices and longitudinal pepsin concentration reproducibility in individual subjects was evaluated. Salivary pepsin was analyzed in 32 GERD patients with extraesophageal reflux symptoms and 13 healthy individuals. RESULTS: The assay's signal increase is not completed at the recommend readout time and continues to increase for another 25 minutes. The relative standard deviation of measurement was good when using the same LF device, ranging from 2.3% to 12.9%, but the reproducibility of 10 different individual LF devices was poor. The random error when analyzing the same saliva sample on 10 LF devices was as high as 36 ng/mL and this value is thus suggested as the positivity cut-off. Pepsin concentration in individual subjects during a 10-day period varied significantly. The sensitivity of the Peptest was 36.8% in the group with acid reflux and 23.1% in the group with weakly acid reflux. The specificity was 61.5%. CONCLUSIONS: The Peptest assay's sensitivity and specificity is low, the results are highly variable and it should not be used as a near-patient diagnostic method in primary care.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc20025878
003      
CZ-PrNML
005      
20201222155517.0
007      
ta
008      
201125s2019 xxu f 000 0|eng||
009      
AR
024    7_
$a 10.1097/MCG.0000000000001066 $2 doi
035    __
$a (PubMed)29863588
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xxu
100    1_
$a Dolina, Jiří $u Internal Gastroenterology Department.
245    10
$a Evaluation of Important Analytical Parameters of the Peptest Immunoassay that Limit its Use in Diagnosing Gastroesophageal Reflux Disease / $c J. Dolina, Š. Konečný, P. Ďurč, J. Lačná, M. Greguš, F. Foret, J. Skřičková, M. Doubková, D. Kindlová, E. Pokojová, P. Kubáň,
520    9_
$a GOAL: To evaluate the analytical parameters of a lateral flow (LF) pepsin immunoassay (Peptest) and assess its suitability in the diagnostics of gastroesophageal reflux disease (GERD). BACKGROUND: Peptest is a noninvasive assay to analyze pepsin in saliva, intended for use in GERD diagnostics. Although commercialized, fundamental studies on its performance are missing. The assay therefore requires basic analytical parameter evaluation to assess its suitability in clinical practice. STUDY: Assay reaction's time dependence, reader device repeatability, and individual LF devices and longitudinal pepsin concentration reproducibility in individual subjects was evaluated. Salivary pepsin was analyzed in 32 GERD patients with extraesophageal reflux symptoms and 13 healthy individuals. RESULTS: The assay's signal increase is not completed at the recommend readout time and continues to increase for another 25 minutes. The relative standard deviation of measurement was good when using the same LF device, ranging from 2.3% to 12.9%, but the reproducibility of 10 different individual LF devices was poor. The random error when analyzing the same saliva sample on 10 LF devices was as high as 36 ng/mL and this value is thus suggested as the positivity cut-off. Pepsin concentration in individual subjects during a 10-day period varied significantly. The sensitivity of the Peptest was 36.8% in the group with acid reflux and 23.1% in the group with weakly acid reflux. The specificity was 61.5%. CONCLUSIONS: The Peptest assay's sensitivity and specificity is low, the results are highly variable and it should not be used as a near-patient diagnostic method in primary care.
650    _2
$a mladiství $7 D000293
650    _2
$a dospělí $7 D000328
650    _2
$a senioři $7 D000368
650    _2
$a ženské pohlaví $7 D005260
650    _2
$a gastroezofageální reflux $x diagnóza $7 D005764
650    _2
$a lidé $7 D006801
650    12
$a imunoanalýza $7 D007118
650    _2
$a mužské pohlaví $7 D008297
650    _2
$a lidé středního věku $7 D008875
650    _2
$a pepsin A $x metabolismus $7 D010434
650    _2
$a reprodukovatelnost výsledků $7 D015203
650    _2
$a sliny $x metabolismus $7 D012463
650    _2
$a senzitivita a specificita $7 D012680
650    _2
$a mladý dospělý $7 D055815
655    _2
$a hodnotící studie $7 D023362
655    _2
$a časopisecké články $7 D016428
700    1_
$a Konečný, Štefan $u Internal Gastroenterology Department.
700    1_
$a Ďurč, Pavol $u Department of Bioanalytical Instrumentation, CEITEC Masaryk University. Department of Chemistry, Masaryk University, Brno, Czech Republic.
700    1_
$a Lačná, Júlia $u Department of Bioanalytical Instrumentation, CEITEC Masaryk University. Department of Chemistry, Masaryk University, Brno, Czech Republic.
700    1_
$a Greguš, Michal $u Department of Bioanalytical Instrumentation, CEITEC Masaryk University.
700    1_
$a Foret, František $u Department of Bioanalytical Instrumentation, CEITEC Masaryk University.
700    1_
$a Skřičková, Jana $u Department of Pulmonary Diseases and Tuberculosis, University Hospital Brno and Faculty of Medicine, Masaryk University.
700    1_
$a Doubková, Martina $u Department of Pulmonary Diseases and Tuberculosis, University Hospital Brno and Faculty of Medicine, Masaryk University.
700    1_
$a Kindlová, Dagmar $u Department of Pulmonary Diseases and Tuberculosis, University Hospital Brno and Faculty of Medicine, Masaryk University.
700    1_
$a Pokojová, Eva $u Department of Pulmonary Diseases and Tuberculosis, University Hospital Brno and Faculty of Medicine, Masaryk University.
700    1_
$a Kubáň, Petr $u Department of Bioanalytical Instrumentation, CEITEC Masaryk University.
773    0_
$w MED00002586 $t Journal of clinical gastroenterology $x 1539-2031 $g Roč. 53, č. 5 (2019), s. 355-360
856    41
$u https://pubmed.ncbi.nlm.nih.gov/29863588 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20201125 $b ABA008
991    __
$a 20201222155513 $b ABA008
999    __
$a ok $b bmc $g 1600023 $s 1116564
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2019 $b 53 $c 5 $d 355-360 $e - $i 1539-2031 $m Journal of clinical gastroenterology $n J Clin Gastroenterol $x MED00002586
GRA    __
$a NV17-31945A $p MZ0
LZP    __
$a Pubmed-20201125

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...