O-Arm Navigated Frameless and Fiducial-Less Deep Brain Stimulation
D. Krahulík, M. Nevrlý, P. Otruba, J. Bardoň, L. Hrabálek, D. Pohlodek, P. Kaňovský, J. Valošek,
Jazyk angličtina Země Švýcarsko
Typ dokumentu časopisecké články
NLK
Directory of Open Access Journals
od 2011
PubMed Central
od 2011
Europe PubMed Central
od 2011
ProQuest Central
od 2011-01-01
Open Access Digital Library
od 2011-01-01
Open Access Digital Library
od 2011-01-01
ROAD: Directory of Open Access Scholarly Resources
od 2011
PubMed
32992610
DOI
10.3390/brainsci10100683
Knihovny.cz E-zdroje
- Publikační typ
- časopisecké články MeSH
OBJECT: Deep brain stimulation (DBS) is a very useful procedure for the treatment of idiopathic Parkinson's disease (PD), essential tremor, and dystonia. The authors evaluated the accuracy of the new method used in their center for the placing of DBS electrodes. Electrodes are placed using the intraoperative O-armTM (Medtronic)-controlled frameless and fiducial-less system, NexframeTM (Medtronic). Accuracy was evaluated prospectively in eleven consecutive PD patients (22 electrodes). METHODS: Eleven adult patients with PD were implanted using the Nexframe system without fiducials and with the intraoperative O-arm (Medtronic) system and StealthStationTM S8 navigation (Medtronic). The implantation of DBS leads was performed using multiple-cell microelectrode recording, and intraoperative test stimulation to determine thresholds for stimulation-induced adverse effects. The accuracy was checked in three different steps: (1) using the intraoperative O-arm image and its fusion with preoperative planning, (2) using multiple-cell microelectrode recording and counting the number of microelectrodes with the signal of the subthalamic nucleus (STN) and finally, (3) total error was calculated according to a postoperative CT control image fused to preoperative planning. RESULTS: The total error of the procedure was 1.79 mm; the radial error and the vector error were 171 mm and 163 mm. CONCLUSIONS: Implantation of DBS electrodes using an O-arm navigated frameless and fiducial-less system is a very useful and technically feasible procedure with excellent patient toleration with experienced Nexframe users. The accuracy of the method was confirmed at all three steps, and it is comparable to other published results.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21002048
- 003
- CZ-PrNML
- 005
- 20220718085921.0
- 007
- ta
- 008
- 210105s2020 sz f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.3390/brainsci10100683 $2 doi
- 035 __
- $a (PubMed)32992610
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a sz
- 100 1_
- $a Krahulík, David $u Department of Neurosurgery, Faculty of Medicine and Dentistry, Palacký University and University Hospital Olomouc, 77900 Olomouc, Czech Republic.
- 245 10
- $a O-Arm Navigated Frameless and Fiducial-Less Deep Brain Stimulation. / $c D. Krahulík, M. Nevrlý, P. Otruba, J. Bardoň, L. Hrabálek, D. Pohlodek, P. Kaňovský, J. Valošek,
- 520 9_
- $a OBJECT: Deep brain stimulation (DBS) is a very useful procedure for the treatment of idiopathic Parkinson's disease (PD), essential tremor, and dystonia. The authors evaluated the accuracy of the new method used in their center for the placing of DBS electrodes. Electrodes are placed using the intraoperative O-armTM (Medtronic)-controlled frameless and fiducial-less system, NexframeTM (Medtronic). Accuracy was evaluated prospectively in eleven consecutive PD patients (22 electrodes). METHODS: Eleven adult patients with PD were implanted using the Nexframe system without fiducials and with the intraoperative O-arm (Medtronic) system and StealthStationTM S8 navigation (Medtronic). The implantation of DBS leads was performed using multiple-cell microelectrode recording, and intraoperative test stimulation to determine thresholds for stimulation-induced adverse effects. The accuracy was checked in three different steps: (1) using the intraoperative O-arm image and its fusion with preoperative planning, (2) using multiple-cell microelectrode recording and counting the number of microelectrodes with the signal of the subthalamic nucleus (STN) and finally, (3) total error was calculated according to a postoperative CT control image fused to preoperative planning. RESULTS: The total error of the procedure was 1.79 mm; the radial error and the vector error were 171 mm and 163 mm. CONCLUSIONS: Implantation of DBS electrodes using an O-arm navigated frameless and fiducial-less system is a very useful and technically feasible procedure with excellent patient toleration with experienced Nexframe users. The accuracy of the method was confirmed at all three steps, and it is comparable to other published results.
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Nevrlý, Martin $u Department of Neurology, Faculty of Medicine and Dentistry, Palacký University and University Hospital Olomouc, 77900 Olomouc, Czech Republic.
- 700 1_
- $a Otruba, Pavel $u Department of Neurology, Faculty of Medicine and Dentistry, Palacký University and University Hospital Olomouc, 77900 Olomouc, Czech Republic.
- 700 1_
- $a Bardoň, Jan $u Department of Neurology, Faculty of Medicine and Dentistry, Palacký University and University Hospital Olomouc, 77900 Olomouc, Czech Republic.
- 700 1_
- $a Hrabálek, Lumír $u Department of Neurosurgery, Faculty of Medicine and Dentistry, Palacký University and University Hospital Olomouc, 77900 Olomouc, Czech Republic.
- 700 1_
- $a Pohlodek, Daniel $u Department of Neurosurgery, Faculty of Medicine and Dentistry, Palacký University and University Hospital Olomouc, 77900 Olomouc, Czech Republic.
- 700 1_
- $a Kaňovský, Petr $u Department of Neurology, Faculty of Medicine and Dentistry, Palacký University and University Hospital Olomouc, 77900 Olomouc, Czech Republic.
- 700 1_
- $a Valošek, Jan, $u Department of Neurosurgery, Faculty of Medicine and Dentistry, Palacký University and University Hospital Olomouc, 77900 Olomouc, Czech Republic. Department of Neurology, Faculty of Medicine and Dentistry, Palacký University and University Hospital Olomouc, 77900 Olomouc, Czech Republic. $d 1993- $7 xx0274853
- 773 0_
- $w MED00198694 $t Brain sciences $x 2076-3425 $g Roč. 10, č. 10 (2020)
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/32992610 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20210105 $b ABA008
- 991 __
- $a 20220718085920 $b ABA008
- 999 __
- $a ind $b bmc $g 1614128 $s 1122332
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2020 $b 10 $c 10 $e 20200927 $i 2076-3425 $m Brain sciences $n Brain Sci $x MED00198694
- LZP __
- $a Pubmed-20210105