The change of cardiostimulation device programming due to detection of electromagnetic interference

Jan Morava, Aleš Richter

. 2020 ; 50 (2) : 65-68.

Jazyk angličtina Země Česko

Typ dokumentu práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/bmc21004807

The study deals with the interdisciplinary topic of the electromagnetic compatibility of the cardiac implantable electronic devices that are used in patients with a defect of heart conduction system. We are focusing on the detection of disturbing signals on electrodes of cardiostimulation device and its interpretation. The detection of electromagnetic interference (EMI) is related to electrode choice, device placement, its configuration and programming. The aim of the study is the analysis of the pacemaker response in the presence of an external source of the disturbance fields. We point to possible risks of its interaction and discuss mechanisms that can influence the pacemaker sensitivity to EMI. Due to improper signal detection, the device programming changes can occur. We present an experiment of the exposure of the cardiostimulation system to a low-frequency harmonic interference signals and finally we analyse similar clinical episode and discuss proper functioning of the pacemaker.

000      
00000naa a2200000 a 4500
001      
bmc21004807
003      
CZ-PrNML
005      
20210526150539.0
007      
ta
008      
210215s2020 xr a f 000 0|eng||
009      
AR
024    7_
$a 10.14311/CTJ.2020.2.04 $2 doi
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xr
100    1_
$a Morava, Jan $u Faculty of Mechatronics, Informatics and Interdisciplinary Studies, Technical University of Liberec, Liberec; Department of Cardiology, Regional Hospital Liberec, Liberec; Faculty of Health Studies, Technical University of Liberec, Liberec $7 xx0257591
245    14
$a The change of cardiostimulation device programming due to detection of electromagnetic interference / $c Jan Morava, Aleš Richter
520    9_
$a The study deals with the interdisciplinary topic of the electromagnetic compatibility of the cardiac implantable electronic devices that are used in patients with a defect of heart conduction system. We are focusing on the detection of disturbing signals on electrodes of cardiostimulation device and its interpretation. The detection of electromagnetic interference (EMI) is related to electrode choice, device placement, its configuration and programming. The aim of the study is the analysis of the pacemaker response in the presence of an external source of the disturbance fields. We point to possible risks of its interaction and discuss mechanisms that can influence the pacemaker sensitivity to EMI. Due to improper signal detection, the device programming changes can occur. We present an experiment of the exposure of the cardiostimulation system to a low-frequency harmonic interference signals and finally we analyse similar clinical episode and discuss proper functioning of the pacemaker.
650    _2
$a lidé $7 D006801
650    _2
$a kardiostimulace umělá $7 D002304
650    12
$a kardiostimulátor $7 D010138
650    _2
$a elektrofyziologie $7 D004594
650    12
$a elektromagnetické jevy $7 D055590
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Richter, Aleš, $u Faculty of Mechatronics, Informatics and Interdisciplinary Studies, Technical University of Liberec, Liberec; Faculty of Health Studies, Technical University of Liberec, Liberec $d 1957- $7 xx0000450
773    0_
$w MED00011033 $t Lékař a technika $x 0301-5491 $g Roč. 50, č. 2 (2020), s. 65-68
856    41
$u https://ojs.cvut.cz/ojs/index.php/CTJ/article/view/6879/5843 $y plný text volně přístupný
910    __
$a ABA008 $b B 1367 $c 1071 b $y p $z 0
990    __
$a 20210215 $b ABA008
991    __
$a 20210526150537 $b ABA008
999    __
$a ok $b bmc $g 1624868 $s 1125124
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2020 $b 50 $c 2 $d 65-68 $i 0301-5491 $m Lékař a technika $x MED00011033 $y 125690
LZP    __
$c NLK109 $d 20210526 $b NLK111 $a Meditorial-20210215

Citační ukazatele

Nahrávání dat...

Možnosti archivace

Nahrávání dat...