The study deals with the interdisciplinary topic of the electromagnetic compatibility of the cardiac implantable electronic devices that are used in patients with a defect of heart conduction system. We are focusing on the detection of disturbing signals on electrodes of cardiostimulation device and its interpretation. The detection of electromagnetic interference (EMI) is related to electrode choice, device placement, its configuration and programming. The aim of the study is the analysis of the pacemaker response in the presence of an external source of the disturbance fields. We point to possible risks of its interaction and discuss mechanisms that can influence the pacemaker sensitivity to EMI. Due to improper signal detection, the device programming changes can occur. We present an experiment of the exposure of the cardiostimulation system to a low-frequency harmonic interference signals and finally we analyse similar clinical episode and discuss proper functioning of the pacemaker.
Department of Cardiology Regional Hospital Liberec Liberec
Faculty of Health Studies Technical University of Liberec Liberec
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21004807
- 003
- CZ-PrNML
- 005
- 20210526150539.0
- 007
- ta
- 008
- 210215s2020 xr a f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.14311/CTJ.2020.2.04 $2 doi
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Morava, Jan $u Faculty of Mechatronics, Informatics and Interdisciplinary Studies, Technical University of Liberec, Liberec; Department of Cardiology, Regional Hospital Liberec, Liberec; Faculty of Health Studies, Technical University of Liberec, Liberec $7 xx0257591
- 245 14
- $a The change of cardiostimulation device programming due to detection of electromagnetic interference / $c Jan Morava, Aleš Richter
- 520 9_
- $a The study deals with the interdisciplinary topic of the electromagnetic compatibility of the cardiac implantable electronic devices that are used in patients with a defect of heart conduction system. We are focusing on the detection of disturbing signals on electrodes of cardiostimulation device and its interpretation. The detection of electromagnetic interference (EMI) is related to electrode choice, device placement, its configuration and programming. The aim of the study is the analysis of the pacemaker response in the presence of an external source of the disturbance fields. We point to possible risks of its interaction and discuss mechanisms that can influence the pacemaker sensitivity to EMI. Due to improper signal detection, the device programming changes can occur. We present an experiment of the exposure of the cardiostimulation system to a low-frequency harmonic interference signals and finally we analyse similar clinical episode and discuss proper functioning of the pacemaker.
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a kardiostimulace umělá $7 D002304
- 650 12
- $a kardiostimulátor $7 D010138
- 650 _2
- $a elektrofyziologie $7 D004594
- 650 12
- $a elektromagnetické jevy $7 D055590
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Richter, Aleš, $u Faculty of Mechatronics, Informatics and Interdisciplinary Studies, Technical University of Liberec, Liberec; Faculty of Health Studies, Technical University of Liberec, Liberec $d 1957- $7 xx0000450
- 773 0_
- $w MED00011033 $t Lékař a technika $x 0301-5491 $g Roč. 50, č. 2 (2020), s. 65-68
- 856 41
- $u https://ojs.cvut.cz/ojs/index.php/CTJ/article/view/6879/5843 $y plný text volně přístupný
- 910 __
- $a ABA008 $b B 1367 $c 1071 b $y p $z 0
- 990 __
- $a 20210215 $b ABA008
- 991 __
- $a 20210526150537 $b ABA008
- 999 __
- $a ok $b bmc $g 1624868 $s 1125124
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2020 $b 50 $c 2 $d 65-68 $i 0301-5491 $m Lékař a technika $x MED00011033 $y 125690
- LZP __
- $c NLK109 $d 20210526 $b NLK111 $a Meditorial-20210215