Optimized convective volume in online hemodiafiltration
František Lopot, Vratislav Fabián
Jazyk angličtina Země Česko
- MeSH
- chronické selhání ledvin terapie MeSH
- dialýza ledvin metody MeSH
- hemodiafiltrace * mortalita škodlivé účinky MeSH
- lidé MeSH
- povrch těla MeSH
- randomizované kontrolované studie jako téma MeSH
- Check Tag
- lidé MeSH
Hemodiafiltration (HDF) adds convective elimination of middle molecules (MM), proportional to filtered volume (Vconv) on the top of diffusion-based elimination of small molecules (SM) by conventional hemodialysis (HD). Studies, both observational and randomized controlled ones, performed so far generally indicated positive impact of higher Vconv on all-cause mortality in HDF patients, although the magnitude of Vconv at which HDF becomes apparently superior to HD differed widely among the studies. Also the issue of a suitable anthropometric parameter by which the Vconv should be normalized has not yet been solved. Data from the ESHOL study seems to indicate that patient´s body surface area (BSA) could well be used for this—mortality was decreasing with increasing Vconv/BSA with a bottom plateau starting at about 15 L/m2 . We have therefore devised a formula and a graph for determination of Vconv which fulfils the requirement Vconv/BSA = 15. Assuming maximum feasible and safe filtration fraction QF/QB = 0.3, the Vconv actually defines the necessary blood flow (QB) to reach Vconv/BSA = 15 as QB = 15·BSA/(0.3·t) (t – session time). It is also possible to check what QB is needed in terms of sufficient diffusion-based transport (target Kt/V) and compare both QB values to see if adequate combined elimination of both SM and MM can be achieved at the same time and under what conditions, respectively.
Department of Medicine General University Hospital Prague Prague Strahov Czech Republic
Department of Physics Faculty of Electrical Engineering Czech Technical University Prague
Institute of Biophysics and Medical Informatics 1st Faculty of Medicine Charles University Prague
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21014493
- 003
- CZ-PrNML
- 005
- 20210616200359.0
- 007
- ta
- 008
- 210510s2020 xr d f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.14311/CTJ.2020.3.01 $2 doi
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Lopot, František, $u Department of Medicine, General University Hospital in Prague, Prague-Strahov, Czech Republic; Institute of Biophysics and Medical Informatics, First Faculty of Medicine, Charles University, Prague $d 1950- $7 nlk20010095505
- 245 10
- $a Optimized convective volume in online hemodiafiltration / $c František Lopot, Vratislav Fabián
- 520 9_
- $a Hemodiafiltration (HDF) adds convective elimination of middle molecules (MM), proportional to filtered volume (Vconv) on the top of diffusion-based elimination of small molecules (SM) by conventional hemodialysis (HD). Studies, both observational and randomized controlled ones, performed so far generally indicated positive impact of higher Vconv on all-cause mortality in HDF patients, although the magnitude of Vconv at which HDF becomes apparently superior to HD differed widely among the studies. Also the issue of a suitable anthropometric parameter by which the Vconv should be normalized has not yet been solved. Data from the ESHOL study seems to indicate that patient´s body surface area (BSA) could well be used for this—mortality was decreasing with increasing Vconv/BSA with a bottom plateau starting at about 15 L/m2 . We have therefore devised a formula and a graph for determination of Vconv which fulfils the requirement Vconv/BSA = 15. Assuming maximum feasible and safe filtration fraction QF/QB = 0.3, the Vconv actually defines the necessary blood flow (QB) to reach Vconv/BSA = 15 as QB = 15·BSA/(0.3·t) (t – session time). It is also possible to check what QB is needed in terms of sufficient diffusion-based transport (target Kt/V) and compare both QB values to see if adequate combined elimination of both SM and MM can be achieved at the same time and under what conditions, respectively.
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a randomizované kontrolované studie jako téma $7 D016032
- 650 12
- $a hemodiafiltrace $x mortalita $x škodlivé účinky $7 D017583
- 650 _2
- $a chronické selhání ledvin $x terapie $7 D007676
- 650 _2
- $a dialýza ledvin $x metody $7 D006435
- 650 _2
- $a povrch těla $7 D001830
- 700 1_
- $a Fabián, Vratislav $u Department of Physics, Faculty of Electrical Engineering, Czech Technical University in Prague $7 xx0087412
- 773 0_
- $w MED00011033 $t Lékař a technika $x 0301-5491 $g Roč. 50, č. 3 (2020), s. 77-84
- 856 41
- $u https://www.prolekare.cz/casopisy/lekar-a-technika/2020-3-31/optimized-convective-volume-in-online-hemodiafiltration-126788 $y plný text volně dostupný
- 910 __
- $a ABA008 $b B 1367 $c 1071 b $y p $z 0
- 990 __
- $a 20210510 $b ABA008
- 991 __
- $a 20210616200357 $b ABA008
- 999 __
- $a ok $b bmc $g 1655140 $s 1134885
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2020 $b 50 $c 3 $d 77-84 $i 0301-5491 $m Lékař a technika $x MED00011033 $y 126788
- LZP __
- $c NLK109 $d 20210616 $b NLK111 $a Meditorial-20210510