Comparison of the Chemiluminescence Immunoassay LIAISON® with the Radioimmunoassay for Aldosterone and Renin Measurement
J. Uhrová, H. Benáková, Z. Vaníčková, T. Zima
Jazyk angličtina Země Česko
Typ dokumentu časopisecké články
NLK
Directory of Open Access Journals
od 2012
Medline Complete (EBSCOhost)
od 2012-01-01
ROAD: Directory of Open Access Scholarly Resources
od 2013
PubMed
34137684
DOI
10.14712/23362936.2021.9
Knihovny.cz E-zdroje
- MeSH
- aldosteron * MeSH
- hyperaldosteronismus * MeSH
- lidé MeSH
- luminiscence MeSH
- radioimunoanalýza MeSH
- renin MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
Determination of renin plasma levels is useful in the diagnosis of hypertension and in the therapeutic follow-up of hypertensive patients. Plasmatic concentration of renin decreases in patients with hypertension due to a primary hyperaldosteronism, contrary to renovascular hypertension where concentrations of renin and aldosterone are both elevated. Blood samples (serum, EDTA plasma) were analysed using two different chemiluminiscent methods CLIA LIAISON® and radioimmunoassay for aldosterone (IMMUNOTECH Beckman Coulter) and renin (Cisbio Bioassay) measurements were compared. We used both methods to ascertain the correlation between serum vs. EDTA plasma levels of aldosterone (RIA, CLIA) and renin (IRMA, CLIA) and to compare aldosterone to renin ratios for CLIA and for radioimmunoassay: serum aldosterone to plasma renin and plasma aldosterone to plasma renin. We compared serum aldosterone CLIA vs. RIA (rP=0.933, P<0.001) and plasma renin determined using CLIA vs. IRMA (rP=0.965, P=0.062). Furthermore, we used both methods to establish the correlation between the serum vs. plasma levels of aldosterone: RIA (rP=0.980, P<0.001); CLIA (rP=0.994, P=0.353) and serum vs. plasma levels of renin: IRMA (rP=0.948, P<0.001); CLIA (rP=0.921, P=0.011). Aldosterone (serum, plasma) to plasmatic renin ratios for CLIA (rP=0.999, P=0.286) and for radioimmunoassay (rP=0.992, P=0.025). Our data demonstrate that renin and aldosterone concentrations obtained using CLIA correlate with renin and aldosterone concentrations using radioimmunoassay methods. Correlation coefficients of pair results ranged from 0.921 to 0.994. Aldosterone (serum, EDTA plasma) to plasmatic renin ratios are comparable and any of them can be used with no significant differences found.
Citace poskytuje Crossref.org
Literatura
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21017503
- 003
- CZ-PrNML
- 005
- 20210811104038.0
- 007
- ta
- 008
- 210723s2021 xr d f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.14712/23362936.2021.9 $2 doi
- 035 __
- $a (PubMed)34137684
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Uhrová, Jana $7 xx0136255 $u Institute of Medical Biochemistry and Laboratory Diagnostics, First Faculty of Medicine, Charles University and General University Hospital in Prague, Prague, Czech Republic
- 245 10
- $a Comparison of the Chemiluminescence Immunoassay LIAISON® with the Radioimmunoassay for Aldosterone and Renin Measurement / $c J. Uhrová, H. Benáková, Z. Vaníčková, T. Zima
- 504 __
- $a Literatura
- 520 9_
- $a Determination of renin plasma levels is useful in the diagnosis of hypertension and in the therapeutic follow-up of hypertensive patients. Plasmatic concentration of renin decreases in patients with hypertension due to a primary hyperaldosteronism, contrary to renovascular hypertension where concentrations of renin and aldosterone are both elevated. Blood samples (serum, EDTA plasma) were analysed using two different chemiluminiscent methods CLIA LIAISON® and radioimmunoassay for aldosterone (IMMUNOTECH Beckman Coulter) and renin (Cisbio Bioassay) measurements were compared. We used both methods to ascertain the correlation between serum vs. EDTA plasma levels of aldosterone (RIA, CLIA) and renin (IRMA, CLIA) and to compare aldosterone to renin ratios for CLIA and for radioimmunoassay: serum aldosterone to plasma renin and plasma aldosterone to plasma renin. We compared serum aldosterone CLIA vs. RIA (rP=0.933, P<0.001) and plasma renin determined using CLIA vs. IRMA (rP=0.965, P=0.062). Furthermore, we used both methods to establish the correlation between the serum vs. plasma levels of aldosterone: RIA (rP=0.980, P<0.001); CLIA (rP=0.994, P=0.353) and serum vs. plasma levels of renin: IRMA (rP=0.948, P<0.001); CLIA (rP=0.921, P=0.011). Aldosterone (serum, plasma) to plasmatic renin ratios for CLIA (rP=0.999, P=0.286) and for radioimmunoassay (rP=0.992, P=0.025). Our data demonstrate that renin and aldosterone concentrations obtained using CLIA correlate with renin and aldosterone concentrations using radioimmunoassay methods. Correlation coefficients of pair results ranged from 0.921 to 0.994. Aldosterone (serum, EDTA plasma) to plasmatic renin ratios are comparable and any of them can be used with no significant differences found.
- 650 12
- $a aldosteron $7 D000450
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a hyperaldosteronismus $7 D006929
- 650 _2
- $a luminiscence $7 D049449
- 650 _2
- $a radioimunoanalýza $7 D011863
- 650 _2
- $a renin $7 D012083
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Benáková, Hana, $d 1954- $7 xx0070521 $u Institute of Medical Biochemistry and Laboratory Diagnostics, First Faculty of Medicine, Charles University and General University Hospital in Prague, Prague, Czech Republic
- 700 1_
- $a Vaníčková, Zdislava $7 xx0262843 $u Institute of Medical Biochemistry and Laboratory Diagnostics, First Faculty of Medicine, Charles University and General University Hospital in Prague, Prague, Czech Republic
- 700 1_
- $a Zima, Tomáš, $d 1966- $7 jn20000620440 $u Institute of Medical Biochemistry and Laboratory Diagnostics, First Faculty of Medicine, Charles University and General University Hospital in Prague, Prague, Czech Republic
- 773 0_
- $w MED00013414 $t Prague medical report $x 1214-6994 $g Roč. 122, č. 2 (2021), s. 80-95
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/34137684 $y Pubmed
- 910 __
- $a ABA008 $b A 7 $c 1071 $y p $z 0
- 990 __
- $a 20210723 $b ABA008
- 991 __
- $a 20210803100355 $b ABA008
- 999 __
- $a ok $b bmc $g 1682315 $s 1137945
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 122 $c 2 $d 80-95 $i 1214-6994 $m Prague Medical Report $n Prague Med. Rep. $x MED00013414
- LZP __
- $b NLK118 $a Pubmed-20210723