Evaluation of Astigmatism-Correcting Efficiency and Rotational Stability after Cataract Surgery with a Double-Loop Haptic Toric Intraocular Lens: A 1-Year Follow-Up
LV. Nováček, M. Němcová, K. Tyx, K. Lahodová, L. Rejmont, P. Rozsíval, P. Studený
Jazyk angličtina Země Švýcarsko
Typ dokumentu časopisecké články
NLK
Directory of Open Access Journals
od 2017
PubMed Central
od 2016
ProQuest Central
od 2017-01-01
Health & Medicine (ProQuest)
od 2017-01-01
Karger Open Access
od 2016-01-01
ROAD: Directory of Open Access Scholarly Resources
od 2014
PubMed
33791315
DOI
10.1159/000513894
Knihovny.cz E-zdroje
- Publikační typ
- časopisecké články MeSH
Objectives: The aim of this study was to assess the clinical outcomes, predictability of results, efficiency of astigmatism correction, and rotational stability of the Bi-Flex 677TAY (Medicontur Medical Engineering Ltd., Zsámbék, Hungary) monofocal toric intraocular lens (IOL) designed for cataract patients with astigmatism. Methods: The IOLs were implanted either mono- or binocularly, following routine cataract surgery. Visual and refractive outcomes, as well as off-axis rotation were assessed throughout a 1-year follow-up period. All clinical data for this work were collected retrospectively. Vector analysis based on the Alpins method was performed to assess the efficiency of astigmatism correction. Results: No complications or adverse events occurred during surgery or the follow-up period. IOL implantation brought 88% of eyes into the ±0.50 D, and 100% into the ± 1.00 D range compared to the target spherical equivalent refraction, emmetropia. Astigmatism correction brought similar results: 94% of eyes had a residual cylindrical error of not higher than ±0.50 D, and 97% were within ±1.00 D. Vector analysis resulted in a correction index of 0.96 and a difference vector of 0.17. Both refractive and visual outcomes showed long-term stability. During the 12-month follow-up period, no eyes had a rotation of >5°. Absolute rotation after 1 year was 1.42 ± 1.89° (median = 0°), while signed rotation was 1.06 ± 2.12° (median = 0°). Conclusion: The Bi-Flex 677TAY monofocal toric IOL, designed by Medicontur Medical Engineering Ltd., represents an efficient and safe solution for cataract patients with astigmatism. Clinical and refractive outcomes are predictable, and rotational stability ensures long-term visual comfort.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21018250
- 003
- CZ-PrNML
- 005
- 20240620085305.0
- 007
- ta
- 008
- 210726s2021 sz f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1159/000513894 $2 doi
- 035 __
- $a (PubMed)33791315
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a sz
- 100 1_
- $a Nováček, Ladislav Viktor $u Department of Ophthalmology, Institute of Aviation Medicine Prague, Prague, Czechia $u Department of Ophthalmology, 1st Faculty of Medicine, Charles University and the Military University Hospital Prague, Prague, Czechia
- 245 10
- $a Evaluation of Astigmatism-Correcting Efficiency and Rotational Stability after Cataract Surgery with a Double-Loop Haptic Toric Intraocular Lens: A 1-Year Follow-Up / $c LV. Nováček, M. Němcová, K. Tyx, K. Lahodová, L. Rejmont, P. Rozsíval, P. Studený
- 520 9_
- $a Objectives: The aim of this study was to assess the clinical outcomes, predictability of results, efficiency of astigmatism correction, and rotational stability of the Bi-Flex 677TAY (Medicontur Medical Engineering Ltd., Zsámbék, Hungary) monofocal toric intraocular lens (IOL) designed for cataract patients with astigmatism. Methods: The IOLs were implanted either mono- or binocularly, following routine cataract surgery. Visual and refractive outcomes, as well as off-axis rotation were assessed throughout a 1-year follow-up period. All clinical data for this work were collected retrospectively. Vector analysis based on the Alpins method was performed to assess the efficiency of astigmatism correction. Results: No complications or adverse events occurred during surgery or the follow-up period. IOL implantation brought 88% of eyes into the ±0.50 D, and 100% into the ± 1.00 D range compared to the target spherical equivalent refraction, emmetropia. Astigmatism correction brought similar results: 94% of eyes had a residual cylindrical error of not higher than ±0.50 D, and 97% were within ±1.00 D. Vector analysis resulted in a correction index of 0.96 and a difference vector of 0.17. Both refractive and visual outcomes showed long-term stability. During the 12-month follow-up period, no eyes had a rotation of >5°. Absolute rotation after 1 year was 1.42 ± 1.89° (median = 0°), while signed rotation was 1.06 ± 2.12° (median = 0°). Conclusion: The Bi-Flex 677TAY monofocal toric IOL, designed by Medicontur Medical Engineering Ltd., represents an efficient and safe solution for cataract patients with astigmatism. Clinical and refractive outcomes are predictable, and rotational stability ensures long-term visual comfort.
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Němcová, Marie $u Department of Ophthalmology, Institute of Aviation Medicine Prague, Prague, Czechia
- 700 1_
- $a Tyx, Kateřina $u Department of Ophthalmology, Institute of Aviation Medicine Prague, Prague, Czechia
- 700 1_
- $a Lahodová, Kristýna $u Department of Ophthalmology, Institute of Aviation Medicine Prague, Prague, Czechia $7 xx0318842
- 700 1_
- $a Rejmont, Leoš $u Department of Ophthalmology, 1st Faculty of Medicine, Charles University and the Military University Hospital Prague, Prague, Czechia
- 700 1_
- $a Rozsíval, Pavel $u Department of Ophthalmology, Institute of Aviation Medicine Prague, Prague, Czechia $u Department of Ophthalmology Charles University Prague, Faculty of Medicine in Hradec Králové, Prague, Czechia
- 700 1_
- $a Studený, Pavel $u Department of Ophthalmology Královské Vinohrady University Hospital, and 3rd Faculty of Medicine, Charles University, Prague, Czechia
- 773 0_
- $w MED00207625 $t Biomedicine hub $x 2296-6870 $g Roč. 6, č. 1 (2021), s. 30-41
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/33791315 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20210726 $b ABA008
- 991 __
- $a 20240620085306 $b ABA008
- 999 __
- $a ind $b bmc $g 1676592 $s 1138694
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 6 $c 1 $d 30-41 $e 20210217 $i 2296-6870 $m Biomedicine hub $n Biomed Hub $x MED00207625
- LZP __
- $a Pubmed-20210726