Characterization and comparison of mixed-mode and reversed-phase columns; interaction abilities and applicability for peptide separation
Z. Kadlecová, P. Kozlík, E. Tesařová, M. Gilar, K. Kalíková
Language English Country Netherlands
Document type Journal Article
- MeSH
- Chromatography, Reverse-Phase instrumentation MeSH
- Hydrophobic and Hydrophilic Interactions MeSH
- Hydrogen-Ion Concentration MeSH
- Peptides isolation & purification MeSH
- Buffers MeSH
- Publication type
- Journal Article MeSH
In this work, two mixed-mode columns from a different manufacturers and one marketed as a reversed-phase column were characterized and compared in the terms of their interaction abilities, retentivity, peak symmetry, and applicability for peptide separation. All the tested columns contain octadecyl ligand and positively charged modifier, i.e. pyridyl group for the reversed-phase column XSelect CSH C18, quaternary alkylamine for mixed-mode column Atlantis PREMIER BEH C18 AX, and permanently charged moiety (details not available from the manufacturer) for mixed-mode column Luna Omega PS C18. For detailed characterization and comparison of their interaction potential, several approaches were used. First, a simple Walters test was performed to estimate hydrophobic and silanophilic interactions of the tested columns. The highest values of both parameters were observed for column Atlantis PREMIER BEH C18 AX. To investigate the effect of pH and buffer concentration on retention, mobile phases composed of acetonitrile and buffer (ammonium formate, pH 3.0; ammonium acetate pH 4.7 and pH 6.9) in various concentrations (5mM; 10mM; 15mM and 20mM) were used. The analysis of permanently charged compounds was used to describe the electrostatic interaction abilities of the stationary phases. The most significant contribution of electrostatic interactions to the retention was observed for Atlantis PREMIER BEH C18 AX column in the mobile phase with buffer of pH 3.0. A set of ten dipeptides, three pentapeptides and one octapeptide was used to investigate the effects of pH and buffer concentration on retention and peak symmetry. Each of the tested columns provides the optimal peak shape under different buffer pH and concentration. The gradient separation of the 14 tested peptides was used to verify the application potential of the tested columns for peptide separation. The best separation was achieved within 4 minutes on column Atlantis PREMIER BEH C18 AX.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21018400
- 003
- CZ-PrNML
- 005
- 20210830095954.0
- 007
- ta
- 008
- 210728s2021 ne f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.chroma.2021.462182 $2 doi
- 035 __
- $a (PubMed)33979757
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a ne
- 100 1_
- $a Kadlecová, Zuzana $u Department of Physical and Macromolecular Chemistry, Faculty of Science, Charles University, Hlavova 8, Prague 12843, Czech Republic
- 245 10
- $a Characterization and comparison of mixed-mode and reversed-phase columns; interaction abilities and applicability for peptide separation / $c Z. Kadlecová, P. Kozlík, E. Tesařová, M. Gilar, K. Kalíková
- 520 9_
- $a In this work, two mixed-mode columns from a different manufacturers and one marketed as a reversed-phase column were characterized and compared in the terms of their interaction abilities, retentivity, peak symmetry, and applicability for peptide separation. All the tested columns contain octadecyl ligand and positively charged modifier, i.e. pyridyl group for the reversed-phase column XSelect CSH C18, quaternary alkylamine for mixed-mode column Atlantis PREMIER BEH C18 AX, and permanently charged moiety (details not available from the manufacturer) for mixed-mode column Luna Omega PS C18. For detailed characterization and comparison of their interaction potential, several approaches were used. First, a simple Walters test was performed to estimate hydrophobic and silanophilic interactions of the tested columns. The highest values of both parameters were observed for column Atlantis PREMIER BEH C18 AX. To investigate the effect of pH and buffer concentration on retention, mobile phases composed of acetonitrile and buffer (ammonium formate, pH 3.0; ammonium acetate pH 4.7 and pH 6.9) in various concentrations (5mM; 10mM; 15mM and 20mM) were used. The analysis of permanently charged compounds was used to describe the electrostatic interaction abilities of the stationary phases. The most significant contribution of electrostatic interactions to the retention was observed for Atlantis PREMIER BEH C18 AX column in the mobile phase with buffer of pH 3.0. A set of ten dipeptides, three pentapeptides and one octapeptide was used to investigate the effects of pH and buffer concentration on retention and peak symmetry. Each of the tested columns provides the optimal peak shape under different buffer pH and concentration. The gradient separation of the 14 tested peptides was used to verify the application potential of the tested columns for peptide separation. The best separation was achieved within 4 minutes on column Atlantis PREMIER BEH C18 AX.
- 650 _2
- $a pufry $7 D002021
- 650 _2
- $a chromatografie s reverzní fází $x přístrojové vybavení $7 D056148
- 650 _2
- $a koncentrace vodíkových iontů $7 D006863
- 650 _2
- $a hydrofobní a hydrofilní interakce $7 D057927
- 650 _2
- $a peptidy $x izolace a purifikace $7 D010455
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Kozlík, Petr $u Department of Analytical Chemistry, Faculty of Science, Charles University, Hlavova 8, Prague 12843, Czech Republic
- 700 1_
- $a Tesařová, Eva $u Department of Physical and Macromolecular Chemistry, Faculty of Science, Charles University, Hlavova 8, Prague 12843, Czech Republic
- 700 1_
- $a Gilar, Martin $u Waters Corporation, 34 Maple Street, Milford, MA 01757, USA
- 700 1_
- $a Kalíková, Květa $u Department of Physical and Macromolecular Chemistry, Faculty of Science, Charles University, Hlavova 8, Prague 12843, Czech Republic. Electronic address: kalikova@natur.cuni.cz
- 773 0_
- $w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1648, č. - (2021), s. 462182
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/33979757 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20210728 $b ABA008
- 991 __
- $a 20210830095954 $b ABA008
- 999 __
- $a ok $b bmc $g 1689498 $s 1138844
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 1648 $c - $d 462182 $e 20210427 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
- LZP __
- $a Pubmed-20210728