A Multicenter Evaluation of a Nongel Mechanical Separator Plasma Blood Collection Tube for Testing of Selected Therapeutic Drugs
S. Morosyuk, J. Berube, R. Christenson, AHB. Wu, D. Uettwiller-Geiger, V. Palicka, R. Prusa, T. Zima, K. Shaw, E. Plokhoy, AJ. Ahuja, AK. Stankovic
Jazyk angličtina Země Velká Británie
Typ dokumentu časopisecké články, multicentrická studie, pozorovací studie, práce podpořená grantem
PubMed
32603441
DOI
10.1093/jalm/jfaa033
Knihovny.cz E-zdroje
- MeSH
- lidé MeSH
- monitorování léčiv přístrojové vybavení MeSH
- odběr vzorku krve přístrojové vybavení MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- pozorovací studie MeSH
- práce podpořená grantem MeSH
BACKGROUND: Some therapeutic drugs are unstable during sample storage in gel tubes. BD Vacutainer® Barricor™ Plasma Blood Collection Tube with nongel separator was compared with plasma gel tubes, BD Vacutainer PST™, PST II, and BD Vacutainer Serum Tube for acetaminophen, salicylate, digoxin, carbamazepine, phenytoin, valproic acid, and vancomycin during sample storage for up to 7 days. METHODS: Seven hospital sites enrolled 705 participants who were taking at least one selected drug. The study tubes were collected and tested at initial time (0 h), after 48 h of storage at room temperature and on day 7 (after additional 5 days of refrigerated storage). The performance of BD Barricor tube was evaluated for each drug by comparing BD Barricor samples with samples from the other tubes at 0 h from the same participant; stability was evaluated by comparing test results from the same tube at 0 h, 48 h, and 7 days. RESULTS: At 0 h, BD Barricor showed clinically equivalent results for selected therapeutic drugs compared with the other tubes, except phenytoin in BD PST. Phenytoin samples ≥20 µg/mL in BD PST had 10-12% lower values than samples in BD Barricor. During sample storage, all selected drugs remained stable for 7 days in BD Barricor and in serum aliquots. In BD PST, all drugs remained stable except phenytoin and carbamazepine and in BD PST II except for phenytoin. CONCLUSION: The BD Barricor Tube is effective for the collection and storage of plasma blood samples for therapeutic drug monitoring without sample aliquoting.
General University Hospital Prague Czech Republic
John T Mather Memorial Hospital Port Jefferson NY
School of Medicine University of Maryland MD USA
University Hospital Hradec Krάlόve Czech Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21026659
- 003
- CZ-PrNML
- 005
- 20211026132721.0
- 007
- ta
- 008
- 211013s2020 xxk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1093/jalm/jfaa033 $2 doi
- 035 __
- $a (PubMed)32603441
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxk
- 100 1_
- $a Morosyuk, Svetlana $u BD, Franklin Lakes, NJ
- 245 12
- $a A Multicenter Evaluation of a Nongel Mechanical Separator Plasma Blood Collection Tube for Testing of Selected Therapeutic Drugs / $c S. Morosyuk, J. Berube, R. Christenson, AHB. Wu, D. Uettwiller-Geiger, V. Palicka, R. Prusa, T. Zima, K. Shaw, E. Plokhoy, AJ. Ahuja, AK. Stankovic
- 520 9_
- $a BACKGROUND: Some therapeutic drugs are unstable during sample storage in gel tubes. BD Vacutainer® Barricor™ Plasma Blood Collection Tube with nongel separator was compared with plasma gel tubes, BD Vacutainer PST™, PST II, and BD Vacutainer Serum Tube for acetaminophen, salicylate, digoxin, carbamazepine, phenytoin, valproic acid, and vancomycin during sample storage for up to 7 days. METHODS: Seven hospital sites enrolled 705 participants who were taking at least one selected drug. The study tubes were collected and tested at initial time (0 h), after 48 h of storage at room temperature and on day 7 (after additional 5 days of refrigerated storage). The performance of BD Barricor tube was evaluated for each drug by comparing BD Barricor samples with samples from the other tubes at 0 h from the same participant; stability was evaluated by comparing test results from the same tube at 0 h, 48 h, and 7 days. RESULTS: At 0 h, BD Barricor showed clinically equivalent results for selected therapeutic drugs compared with the other tubes, except phenytoin in BD PST. Phenytoin samples ≥20 µg/mL in BD PST had 10-12% lower values than samples in BD Barricor. During sample storage, all selected drugs remained stable for 7 days in BD Barricor and in serum aliquots. In BD PST, all drugs remained stable except phenytoin and carbamazepine and in BD PST II except for phenytoin. CONCLUSION: The BD Barricor Tube is effective for the collection and storage of plasma blood samples for therapeutic drug monitoring without sample aliquoting.
- 650 _2
- $a odběr vzorku krve $x přístrojové vybavení $7 D001800
- 650 _2
- $a monitorování léčiv $x přístrojové vybavení $7 D016903
- 650 _2
- $a lidé $7 D006801
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a multicentrická studie $7 D016448
- 655 _2
- $a pozorovací studie $7 D064888
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Berube, Julie $u BD, Franklin Lakes, NJ
- 700 1_
- $a Christenson, Robert $u School of Medicine, University of Maryland, MD, USA
- 700 1_
- $a Wu, Alan H B $u University of California, San Francisco, San Francisco, CA
- 700 1_
- $a Uettwiller-Geiger, Denise $u John T. Mather Memorial Hospital, Port Jefferson, NY
- 700 1_
- $a Palicka, Vladimir $u University Hospital, Hradec Krάlόve, Czech Republic
- 700 1_
- $a Prusa, Richard $u University Hospital Motol, Prague, Czech Republic
- 700 1_
- $a Zima, Tomas $u General University Hospital, Prague, Czech Republic
- 700 1_
- $a Shaw, Kristin $u School of Medicine, University of Maryland, MD, USA
- 700 1_
- $a Plokhoy, Elizabeth $u BD, Franklin Lakes, NJ
- 700 1_
- $a Ahuja, Aparna J $u BD, Franklin Lakes, NJ
- 700 1_
- $a Stankovic, Ana K $u BD, Franklin Lakes, NJ
- 773 0_
- $w MED00208302 $t The journal of applied laboratory medicine $x 2576-9456 $g Roč. 5, č. 4 (2020), s. 671-685
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/32603441 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20211013 $b ABA008
- 991 __
- $a 20211026132727 $b ABA008
- 999 __
- $a ok $b bmc $g 1715406 $s 1147166
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2020 $b 5 $c 4 $d 671-685 $e 20200701 $i 2576-9456 $m The journal of applied laboratory medicine $n J. appl. lab. med. $x MED00208302
- LZP __
- $a Pubmed-20211013