Je něco špatně v tomto záznamu ?
A Lattice-Tip Temperature-Controlled Radiofrequency Ablation Catheter for Wide Thermal Lesions: First-in-Human Experience With Atrial Fibrillation
E. Anter, P. Neužil, G. Rackauskas, P. Peichl, A. Aidietis, J. Kautzner, H. Nakagawa, WM. Jackman, A. Natale, VY. Reddy
Jazyk angličtina Země Spojené státy americké
Typ dokumentu časopisecké články, práce podpořená grantem
- MeSH
- fibrilace síní * chirurgie MeSH
- katetrizační ablace * MeSH
- katétry MeSH
- lidé MeSH
- teplota MeSH
- venae pulmonales * chirurgie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
OBJECTIVES: This study sought to evaluate the safety and acute performance of the lattice tip for the treatment of atrial flutter and fibrillation (AF). BACKGROUND: A novel catheter using an expandable lattice structure with a wide thermal footprint incorporating multiple surface thermocouples/mini-electrodes has been designed for high-resolution mapping and high-current, temperature-controlled radiofrequency ablation (RFA). METHODS: Patients with typical right atrial flutter or AF were prospectively enrolled in a single-arm study at 3 centers. Patients with atrial flutter underwent cavotricuspid isthmus (CTI) ablation. Patients with paroxysmal AF underwent pulmonary vein isolation (PVI) and CTI if desired, and for patients with persistent AF, mitral isthmus and left atrial roof lines were also permitted. Mapping was performed with the lattice (Sphere-9) catheter and a novel compatible electroanatomic mapping system (Prism-1). RFA was performed in a point-by-point fashion (Tmax, 73°C to 80°C; range 2 to 7 s). Patients were followed for 3 months. RESULTS: A total of 71 patients underwent ablation: 65 PVI (38% with persistent AF) and 22 mitral isthmus, 24 roof, and 48 CTI lines. PVI was achieved in 64 of 65 (98.5%) by using the lattice alone and required a mean of 2.7 ± 0.70 RFA min. Mitral block was achieved in 100% by using 11.5 ± 10.7 applications and 1.0 ± 0.92 RFA min; only 1 patient required adjunctive epicardial coronary sinus ablation. Roof line and CTI block were achieved in 95.8% and 100% of patients, using 4.9 ± 1.9 and 5.9 ± 3.1 applications for 0.4 ± 0.16 and 0.5 ± 0.24 RFA min, respectively. At 3 months, there were no deaths, strokes, tamponade, or atrioesophageal fistula. CONCLUSIONS: This first-in-human study demonstrated clinical feasibility and safety for rapid high-current, temperature-controlled point-by-point PVI and linear ablation.
Department of Cardiology Homolka Hospital Prague Czech Republic
Department of Cardiovascular Medicine Toyohashi Heart Center Aichi Japan
Institute Klinicke a Experimentalni Mediciny Department of Cardiology Prague Czech Republic
Texas Cardiac Arrhythmia Institute at St David's Medical Center Austin Texas USA
University of Oklahoma Health Sciences Center Oklahoma City Oklahoma USA
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21026720
- 003
- CZ-PrNML
- 005
- 20211026132656.0
- 007
- ta
- 008
- 211013s2020 xxu f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.jacep.2019.12.015 $2 doi
- 035 __
- $a (PubMed)32439034
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxu
- 100 1_
- $a Anter, Elad $u Harvard-Thorndike Electrophysiology Institute, Cardiovascular Division, Department of Medicine, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts, USA
- 245 12
- $a A Lattice-Tip Temperature-Controlled Radiofrequency Ablation Catheter for Wide Thermal Lesions: First-in-Human Experience With Atrial Fibrillation / $c E. Anter, P. Neužil, G. Rackauskas, P. Peichl, A. Aidietis, J. Kautzner, H. Nakagawa, WM. Jackman, A. Natale, VY. Reddy
- 520 9_
- $a OBJECTIVES: This study sought to evaluate the safety and acute performance of the lattice tip for the treatment of atrial flutter and fibrillation (AF). BACKGROUND: A novel catheter using an expandable lattice structure with a wide thermal footprint incorporating multiple surface thermocouples/mini-electrodes has been designed for high-resolution mapping and high-current, temperature-controlled radiofrequency ablation (RFA). METHODS: Patients with typical right atrial flutter or AF were prospectively enrolled in a single-arm study at 3 centers. Patients with atrial flutter underwent cavotricuspid isthmus (CTI) ablation. Patients with paroxysmal AF underwent pulmonary vein isolation (PVI) and CTI if desired, and for patients with persistent AF, mitral isthmus and left atrial roof lines were also permitted. Mapping was performed with the lattice (Sphere-9) catheter and a novel compatible electroanatomic mapping system (Prism-1). RFA was performed in a point-by-point fashion (Tmax, 73°C to 80°C; range 2 to 7 s). Patients were followed for 3 months. RESULTS: A total of 71 patients underwent ablation: 65 PVI (38% with persistent AF) and 22 mitral isthmus, 24 roof, and 48 CTI lines. PVI was achieved in 64 of 65 (98.5%) by using the lattice alone and required a mean of 2.7 ± 0.70 RFA min. Mitral block was achieved in 100% by using 11.5 ± 10.7 applications and 1.0 ± 0.92 RFA min; only 1 patient required adjunctive epicardial coronary sinus ablation. Roof line and CTI block were achieved in 95.8% and 100% of patients, using 4.9 ± 1.9 and 5.9 ± 3.1 applications for 0.4 ± 0.16 and 0.5 ± 0.24 RFA min, respectively. At 3 months, there were no deaths, strokes, tamponade, or atrioesophageal fistula. CONCLUSIONS: This first-in-human study demonstrated clinical feasibility and safety for rapid high-current, temperature-controlled point-by-point PVI and linear ablation.
- 650 12
- $a fibrilace síní $x chirurgie $7 D001281
- 650 12
- $a katetrizační ablace $7 D017115
- 650 _2
- $a katétry $7 D057785
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a venae pulmonales $x chirurgie $7 D011667
- 650 _2
- $a teplota $7 D013696
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Neužil, Petr $u Department of Cardiology, Homolka Hospital, Prague, Czech Republic
- 700 1_
- $a Rackauskas, Gediminas $u Centre for Cardiology and Angiology, Department of Cardiovascular Diseases, Vilnius University, Vilnius, Lithuania
- 700 1_
- $a Peichl, Petr $u Institute Klinicke a Experimentalni Mediciny, Department of Cardiology, Prague, Czech Republic
- 700 1_
- $a Aidietis, Audrius $u Centre for Cardiology and Angiology, Department of Cardiovascular Diseases, Vilnius University, Vilnius, Lithuania
- 700 1_
- $a Kautzner, Josef $u Institute Klinicke a Experimentalni Mediciny, Department of Cardiology, Prague, Czech Republic
- 700 1_
- $a Nakagawa, Hiroshi $u Department of Cardiovascular Medicine, Toyohashi Heart Center, Aichi, Japan
- 700 1_
- $a Jackman, Warren M $u University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma, USA
- 700 1_
- $a Natale, Andrea $u Texas Cardiac Arrhythmia Institute at St. David's Medical Center, Austin, Texas, USA
- 700 1_
- $a Reddy, Vivek Y $u Department of Cardiology, Homolka Hospital, Prague, Czech Republic; Department of Electrophysiology, Division of Cardiology, Icahn School of Medicine at Mount Sinai, New York, New York, USA. Electronic address: vivek.reddy@mountsinai.org
- 773 0_
- $w MED00193518 $t JACC. Clinical electrophysiology $x 2405-5018 $g Roč. 6, č. 5 (2020), s. 507-519
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/32439034 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20211013 $b ABA008
- 991 __
- $a 20211026132702 $b ABA008
- 999 __
- $a ok $b bmc $g 1715456 $s 1147227
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2020 $b 6 $c 5 $d 507-519 $e 20200226 $i 2405-5018 $m JACC. Clinical electrophysiology $n JACC Clin Electrophysiol $x MED00193518
- LZP __
- $a Pubmed-20211013