Je něco špatně v tomto záznamu ?
Dlouhodobé výsledky totální náhrady kolenního kloubu P.F.C. Sigma s celopolyetylenovou tibiální komponentou
[Long-Term Results of Total Knee Replacement Using P.F.C. Sigma System with an All-Polyethylene Tibial Component]
L. Nachtnebl, T. Tomáš, V. Apostolopoulos, L. Pazourek, M. Mahdal
Jazyk čeština Země Česko
Typ dokumentu časopisecké články
- MeSH
- lidé MeSH
- následné studie MeSH
- polyethylen MeSH
- protézy - design MeSH
- protézy kolene * MeSH
- reoperace MeSH
- selhání protézy MeSH
- senioři MeSH
- tibie chirurgie MeSH
- totální endoprotéza kolene * škodlivé účinky MeSH
- Check Tag
- lidé MeSH
- senioři MeSH
- Publikační typ
- časopisecké články MeSH
PURPOSE OF THE STUDY Use of an all-polyethylene (all-poly, AP) tibial component in primary total knee arthroplasty is still an attractive option considering the durability of replacement, the elimination of backside wear and the lower cost compared to modular metalbacked tibia. The purpose of the study was to evaluate the long-term results of the total knee replacement using the P.F.C. Sigma system with a monoblock all-polyethylene tibial component implanted at the 1st Orthopedic Department of the St. Anne's University Hospital and Masaryk University Brno in the period 1999-2010. MATERIAL AND METHODS In the monitored period, 911 total knee replacements using P.F.C. Sigma with an all-polyethylene tibial component were performed. This cohort of patients was evaluated at least 10 years after the primary replacement surgery. Altogether 323 knee replacements in 289 patients were evaluated. The clinical outcomes were assessed according to the Knee Society Clinical Rating System (KSS) and the Total Knee Arthroplasty Roentgenographic Evaluation and Scoring System. The replacement survival was evaluated using the Kaplan-Meier survival analysis. RESULTS In the evaluated group, the average KS was 82.6 points and the average functional score was 74.7 points. Excellent results were recorded in 213 cases, good results in 78 and satisfactory in 23 replacements. Poor clinical results were reported in 9 patients. The average range of motion was 105.8°. In 5 knees (1.5%) revision surgery was performed for infection complications, 3 cases were complicated due to periprosthetic fractures, 4 replacements were revised due to anterior knee pain and 8 knees showed a limited range of motion. Only 1 replacement was revised for aseptic loosening, with no need for reimplantation of the component. The survivorship of the implant was 98.5 % at the follow-up of 12.8 years, including the reasons of reimplantation. DISCUSSION Total knee replacement with an all-polyethylene tibial component is not so popular as the metal-backed modular implants. The all-polyethylene tibial component has certain advantages: lower unit costs, no backside wear, no liner dislocation, possibility of conservative bone resection. On the other hand, this type of implant is not modular and augmentations or longer stems cannot be used. Even though this type of endoprosthesis is recommended exclusively for elderly patients with a lower level of activity, the current mid- and long-term results show that indications for all-poly implants are much broader. CONCLUSIONS The total knee replacement with an all-polyethylene tibial component shows very good long-term clinical outcomes with an excellent survival rate. The all-poly design is cost-effective and may be used not only in elderly patients. Key words: total knee replacement, all-polyethylene tibial component, long-term results.
Long-Term Results of Total Knee Replacement Using P.F.C. Sigma System with an All-Polyethylene Tibial Component
Literatura
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22005928
- 003
- CZ-PrNML
- 005
- 20220224144211.0
- 007
- ta
- 008
- 220207s2021 xr ad f 000 0|cze||
- 009
- AR
- 024 7_
- $2 doi $a 10.55095/achot2021/061
- 035 __
- $a (PubMed)34998443
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a cze $b eng
- 044 __
- $a xr
- 100 1_
- $a Nachtnebl, Luboš $7 xx0106708 $u I. ortopedická klinika Fakultní nemocnice U sv. Anny v Brně, Lékařská fakulta Masarykovy univerzity, Brno
- 245 10
- $a Dlouhodobé výsledky totální náhrady kolenního kloubu P.F.C. Sigma s celopolyetylenovou tibiální komponentou / $c L. Nachtnebl, T. Tomáš, V. Apostolopoulos, L. Pazourek, M. Mahdal
- 246 31
- $a Long-Term Results of Total Knee Replacement Using P.F.C. Sigma System with an All-Polyethylene Tibial Component
- 504 __
- $a Literatura
- 520 9_
- $a PURPOSE OF THE STUDY Use of an all-polyethylene (all-poly, AP) tibial component in primary total knee arthroplasty is still an attractive option considering the durability of replacement, the elimination of backside wear and the lower cost compared to modular metalbacked tibia. The purpose of the study was to evaluate the long-term results of the total knee replacement using the P.F.C. Sigma system with a monoblock all-polyethylene tibial component implanted at the 1st Orthopedic Department of the St. Anne's University Hospital and Masaryk University Brno in the period 1999-2010. MATERIAL AND METHODS In the monitored period, 911 total knee replacements using P.F.C. Sigma with an all-polyethylene tibial component were performed. This cohort of patients was evaluated at least 10 years after the primary replacement surgery. Altogether 323 knee replacements in 289 patients were evaluated. The clinical outcomes were assessed according to the Knee Society Clinical Rating System (KSS) and the Total Knee Arthroplasty Roentgenographic Evaluation and Scoring System. The replacement survival was evaluated using the Kaplan-Meier survival analysis. RESULTS In the evaluated group, the average KS was 82.6 points and the average functional score was 74.7 points. Excellent results were recorded in 213 cases, good results in 78 and satisfactory in 23 replacements. Poor clinical results were reported in 9 patients. The average range of motion was 105.8°. In 5 knees (1.5%) revision surgery was performed for infection complications, 3 cases were complicated due to periprosthetic fractures, 4 replacements were revised due to anterior knee pain and 8 knees showed a limited range of motion. Only 1 replacement was revised for aseptic loosening, with no need for reimplantation of the component. The survivorship of the implant was 98.5 % at the follow-up of 12.8 years, including the reasons of reimplantation. DISCUSSION Total knee replacement with an all-polyethylene tibial component is not so popular as the metal-backed modular implants. The all-polyethylene tibial component has certain advantages: lower unit costs, no backside wear, no liner dislocation, possibility of conservative bone resection. On the other hand, this type of implant is not modular and augmentations or longer stems cannot be used. Even though this type of endoprosthesis is recommended exclusively for elderly patients with a lower level of activity, the current mid- and long-term results show that indications for all-poly implants are much broader. CONCLUSIONS The total knee replacement with an all-polyethylene tibial component shows very good long-term clinical outcomes with an excellent survival rate. The all-poly design is cost-effective and may be used not only in elderly patients. Key words: total knee replacement, all-polyethylene tibial component, long-term results.
- 650 _2
- $a senioři $7 D000368
- 650 12
- $a totální endoprotéza kolene $x škodlivé účinky $7 D019645
- 650 _2
- $a následné studie $7 D005500
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a protézy kolene $7 D007720
- 650 _2
- $a polyethylen $7 D020959
- 650 _2
- $a protézy - design $7 D011474
- 650 _2
- $a selhání protézy $7 D011475
- 650 _2
- $a reoperace $7 D012086
- 650 _2
- $a tibie $x chirurgie $7 D013977
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Tomáš, Tomáš $7 ola2006319238 $u I. ortopedická klinika Fakultní nemocnice U sv. Anny v Brně, Lékařská fakulta Masarykovy univerzity, Brno
- 700 1_
- $a Apostolopoulos, Vasileios. $7 xx0269825 $u I. ortopedická klinika Fakultní nemocnice U sv. Anny v Brně, Lékařská fakulta Masarykovy univerzity, Brno
- 700 1_
- $a Pazourek, Lukáš $7 xx0123254 $u I. ortopedická klinika Fakultní nemocnice U sv. Anny v Brně, Lékařská fakulta Masarykovy univerzity, Brno
- 700 1_
- $a Mahdal, Michal $7 xx0254094 $u I. ortopedická klinika Fakultní nemocnice U sv. Anny v Brně, Lékařská fakulta Masarykovy univerzity, Brno
- 773 0_
- $w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Cechoslovaca $x 0001-5415 $g Roč. 88, č. 6 (2021), s. 412-417
- 856 41
- $u https://achot.cz/pdfs/ach/2021/06/02.pdf $y plný text volně přístupný
- 910 __
- $a ABA008 $b A 8 $c 507 $y p $z 0
- 990 __
- $a 20220207 $b ABA008
- 991 __
- $a 20220224144159 $b ABA008
- 999 __
- $a ok $b bmc $g 1763610 $s 1157080
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 88 $c 6 $d 412-417 $e 20211220 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
- LZP __
- $b NLK118 $a Pubmed-20220207