Je něco špatně v tomto záznamu ?
Evaluation the loosening of abutment screws in fluid contamination: an in vitro study
F. Sun, W. Cheng, BH. Zhao, GQ. Song, Z. Lin
Jazyk angličtina Země Velká Británie
Typ dokumentu časopisecké články
Grantová podpora
51775096
National Natural Science Foundation of China
[2007]006
Chinese Academy of Sciences-WEGO Research Development Plan
No. N2003009
Fundamental Research Funds for the Central Universities
NLK
Directory of Open Access Journals
od 2011
Free Medical Journals
od 2011
Nature Open Access
od 2011-12-01
PubMed Central
od 2011
Europe PubMed Central
od 2011
ProQuest Central
od 2011-01-01
Open Access Digital Library
od 2011-01-01
Open Access Digital Library
od 2011-01-01
Health & Medicine (ProQuest)
od 2011-01-01
ROAD: Directory of Open Access Scholarly Resources
od 2011
Springer Nature OA/Free Journals
od 2011-12-01
- MeSH
- analýza metodou konečných prvků MeSH
- kostní šrouby * MeSH
- lubrikace MeSH
- točivý moment MeSH
- tření MeSH
- zubní implantáty * MeSH
- Publikační typ
- časopisecké články MeSH
Screw loosening is one of the most common clinical problems of dental implants. Research on the influencing factors of screw loosening is very important to prevent screw loosening. The purpose of this in vitro study was to evaluate the influence of liquid contamination on the screw loosening. According to the contamination condition, forty-five abutment screws were divided into three groups (n = 15): no contamination, artificial saliva contamination, and mouthwash contamination. The preload and friction coefficient of the abutment screws were recorded. Then, the reverse torque values (RTVs) and settlement were measured after 3.0 × 105 and 6.0 × 105 cycles. The surface wear of the screws was analyzed. Finally, the stress distribution of the abutment screws was calculated by finite element analysis (FEA). The results showed that fluid contamination reduced the friction coefficient, increased the preload, decrease the settlement, improved resistance to screw loosening, and reduced wear on the thread surface. Appropriate antimicrobial lubrication may improve the anti-loosening performance of abutment screws and prevent excessive wear on the threaded surface.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22017952
- 003
- CZ-PrNML
- 005
- 20220804134458.0
- 007
- ta
- 008
- 220720s2022 xxk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1038/s41598-022-14791-w $2 doi
- 035 __
- $a (PubMed)35750776
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxk
- 100 1_
- $a Sun, Fei $u Key Laboratory of Implant Device and Interface Science of Liaoning Province, School of Mechanical Engineering and Automation, Northeastern University, Shenyang, 110819, China
- 245 10
- $a Evaluation the loosening of abutment screws in fluid contamination: an in vitro study / $c F. Sun, W. Cheng, BH. Zhao, GQ. Song, Z. Lin
- 520 9_
- $a Screw loosening is one of the most common clinical problems of dental implants. Research on the influencing factors of screw loosening is very important to prevent screw loosening. The purpose of this in vitro study was to evaluate the influence of liquid contamination on the screw loosening. According to the contamination condition, forty-five abutment screws were divided into three groups (n = 15): no contamination, artificial saliva contamination, and mouthwash contamination. The preload and friction coefficient of the abutment screws were recorded. Then, the reverse torque values (RTVs) and settlement were measured after 3.0 × 105 and 6.0 × 105 cycles. The surface wear of the screws was analyzed. Finally, the stress distribution of the abutment screws was calculated by finite element analysis (FEA). The results showed that fluid contamination reduced the friction coefficient, increased the preload, decrease the settlement, improved resistance to screw loosening, and reduced wear on the thread surface. Appropriate antimicrobial lubrication may improve the anti-loosening performance of abutment screws and prevent excessive wear on the threaded surface.
- 650 12
- $a kostní šrouby $7 D001863
- 650 12
- $a zubní implantáty $7 D015921
- 650 _2
- $a analýza metodou konečných prvků $7 D020342
- 650 _2
- $a tření $7 D017276
- 650 _2
- $a lubrikace $7 D008153
- 650 _2
- $a točivý moment $7 D019415
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Cheng, Wei $u Key Laboratory of Implant Device and Interface Science of Liaoning Province, School of Mechanical Engineering and Automation, Northeastern University, Shenyang, 110819, China
- 700 1_
- $a Zhao, Bao-Hong $u Key Laboratory of Oral Diseases of Liaoning Province, School and Hospital of Stomatology, China Medical University, Shenyang, 110002, China
- 700 1_
- $a Song, Gui-Qiu $u Key Laboratory of Implant Device and Interface Science of Liaoning Province, School of Mechanical Engineering and Automation, Northeastern University, Shenyang, 110819, China
- 700 1_
- $a Lin, Zeng $u Key Laboratory of Implant Device and Interface Science of Liaoning Province, School of Mechanical Engineering and Automation, Northeastern University, Shenyang, 110819, China. zlin@mail.neu.edu.cn
- 773 0_
- $w MED00182195 $t Scientific reports $x 2045-2322 $g Roč. 12, č. 1 (2022), s. 10797
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/35750776 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20220720 $b ABA008
- 991 __
- $a 20220804134452 $b ABA008
- 999 __
- $a ok $b bmc $g 1821848 $s 1169195
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2022 $b 12 $c 1 $d 10797 $e 20220624 $i 2045-2322 $m Scientific reports $n Sci Rep $x MED00182195
- GRA __
- $a 51775096 $p National Natural Science Foundation of China
- GRA __
- $a [2007]006 $p Chinese Academy of Sciences-WEGO Research Development Plan
- GRA __
- $a No. N2003009 $p Fundamental Research Funds for the Central Universities
- LZP __
- $a Pubmed-20220720