. 2022 ; 22 (11) : . [pub] 20220602

Language English Country Switzerland

Document type Journal Article

Grant support
SGS20/202/OHK4/3T/17 Czech Technical University in Prague

The reliability of pulse oximetry is crucial, especially in cases of rapid changes in body oxygenation. In order to evaluate the performance of pulse oximeters during rapidly developing short periods of concurrent hypoxemia and hypercapnia, 13 healthy volunteers underwent 3 breathing phases during outdoor experiments (39 phases in total), monitored simultaneously by five different pulse oximeters. A significant incongruity in values displayed by the tested pulse oximeters was observed, even when the accuracy declared by the manufacturers were considered. In 28.2% of breathing phases, the five used devices did not show any congruent values. The longest uninterrupted congruent period formed 74.4% of total recorded time. Moreover, the congruent periods were rarely observed during the critical desaturation phase of the experiment. The time difference between the moments when the first and the last pulse oximeter showed the typical study endpoint values of SpO2 85% and 75% was 32.1 ± 23.6 s and 24.7 ± 19.3 s, respectively. These results suggest that SpO2 might not be a reliable parameter as a study endpoint, or more importantly as a safety limit in outdoor experiments. In the design of future studies, more parameters and continuous clinical assessment should be included.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc22018072
003      
CZ-PrNML
005      
20220804134533.0
007      
ta
008      
220720s2022 sz f 000 0|eng||
009      
AR
024    7_
$a 10.3390/s22114236 $2 doi
035    __
$a (PubMed)35684858
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a sz
100    1_
$a Horakova, Lenka $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, 272 01 Kladno, Czech Republic $1 https://orcid.org/0000000315108412 $7 xx0255519
245    10
$a Pulse Oximeter Performance during Rapid Desaturation / $c L. Horakova, K. Roubik
520    9_
$a The reliability of pulse oximetry is crucial, especially in cases of rapid changes in body oxygenation. In order to evaluate the performance of pulse oximeters during rapidly developing short periods of concurrent hypoxemia and hypercapnia, 13 healthy volunteers underwent 3 breathing phases during outdoor experiments (39 phases in total), monitored simultaneously by five different pulse oximeters. A significant incongruity in values displayed by the tested pulse oximeters was observed, even when the accuracy declared by the manufacturers were considered. In 28.2% of breathing phases, the five used devices did not show any congruent values. The longest uninterrupted congruent period formed 74.4% of total recorded time. Moreover, the congruent periods were rarely observed during the critical desaturation phase of the experiment. The time difference between the moments when the first and the last pulse oximeter showed the typical study endpoint values of SpO2 85% and 75% was 32.1 ± 23.6 s and 24.7 ± 19.3 s, respectively. These results suggest that SpO2 might not be a reliable parameter as a study endpoint, or more importantly as a safety limit in outdoor experiments. In the design of future studies, more parameters and continuous clinical assessment should be included.
650    _2
$a lidé $7 D006801
650    _2
$a hypoxie $x diagnóza $7 D000860
650    _2
$a monitorování fyziologických funkcí $7 D008991
650    12
$a oxymetrie $7 D010092
650    12
$a kyslík $7 D010100
650    _2
$a reprodukovatelnost výsledků $7 D015203
655    _2
$a časopisecké články $7 D016428
700    1_
$a Roubik, Karel $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, 272 01 Kladno, Czech Republic $1 https://orcid.org/0000000158136355
773    0_
$w MED00008309 $t Sensors (Basel, Switzerland) $x 1424-8220 $g Roč. 22, č. 11 (2022)
856    41
$u https://pubmed.ncbi.nlm.nih.gov/35684858 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y p $z 0
990    __
$a 20220720 $b ABA008
991    __
$a 20220804134527 $b ABA008
999    __
$a ok $b bmc $g 1821909 $s 1169315
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2022 $b 22 $c 11 $e 20220602 $i 1424-8220 $m Sensors $n Sensors Basel $x MED00008309
GRA    __
$a SGS20/202/OHK4/3T/17 $p Czech Technical University in Prague
LZP    __
$a Pubmed-20220720

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...