Validation of a single tube 3-colour immature red blood cell screening assay for the detection and enumeration of small, medium and large paroxysmal nocturnal haemoglobinuria clones by flow cytometry
I. Marinov, SJ. Richards, A. Pešek, AJ. Illingworth, DR. Sutherland
Jazyk angličtina Země Anglie, Velká Británie
Typ dokumentu časopisecké články
PubMed
35614534
DOI
10.1111/ijlh.13877
Knihovny.cz E-zdroje
- MeSH
- barva MeSH
- buněčné klony MeSH
- erytrocyty MeSH
- lidé MeSH
- paroxysmální hemoglobinurie * diagnóza MeSH
- průtoková cytometrie metody MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
INTRODUCTION: The reliable diagnosis of paroxysmal nocturnal haemoglobinuria (PNH) by flow cytometry is based on mandatory analysis of the erythroid, neutrophilic and monocytic lineages. In this study, we have evaluated the performance characteristics of a recently published immature red blood cell (iRBC) assay as a potential screening test for PNH by flow cytometry. METHODS: Intra- and inter-assay imprecision were determined in five replicates of small, medium and large PNH iRBC clones. Analytical and functional sensitivity was assessed by performing spiking tests for five replicates. Thirty healthy donors and 441 PNH patients were tested for evaluation of clinical specificity, sensitivity, positive and negative predictive values. RESULTS: Coefficients of variation (CV) for intra-/inter-assay imprecision analyses were 1.31/1.50, 3.19/2.61 and 3.99/1.58 for the big, medium and small clone sizes, respectively. Absolute values (100%) were found for both clinical specificity and sensitivity as well as for both positive and negative predictive values. The CV from 5 replicate results for 10 clustered events was 15.7%. The coefficient of determination (r2 ), Pearson's correlation coefficient (r) and Bland-Altman mean bias were 0.9436/0.9234/1.7 for PNH iRBC compared to PNH neutrophils and 0.9553/0.9387/2.1 for PNH iRBCs compared to PNH monocytes. CONCLUSION: Our results confirm very good performance characteristics, high analytical and functional sensitivity, absolute clinical specificity and sensitivity as well as favourable correlation between PNH iRBCs and both PNH neutrophils and monocytes, suggesting that this cost-effective 3-colour iRBC assay can be used as a reliable screening test for evaluation of small, medium and large PNH clones by flow cytometry.
Clinical Department Institute of Haematology and Blood Transfusion Prague Czech Republic
Dahl Chase Diagnostic Services Bangor Maine USA
Department of Laboratory Medicine Toronto General Hospital Toronto Ontario Canada
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22024280
- 003
- CZ-PrNML
- 005
- 20221031100238.0
- 007
- ta
- 008
- 221017s2022 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1111/ijlh.13877 $2 doi
- 035 __
- $a (PubMed)35614534
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Marinov, Iuri $u Clinical Department, Institute of Haematology and Blood Transfusion, Prague, Czech Republic $1 https://orcid.org/https://orcid.org/0000000300253131
- 245 10
- $a Validation of a single tube 3-colour immature red blood cell screening assay for the detection and enumeration of small, medium and large paroxysmal nocturnal haemoglobinuria clones by flow cytometry / $c I. Marinov, SJ. Richards, A. Pešek, AJ. Illingworth, DR. Sutherland
- 520 9_
- $a INTRODUCTION: The reliable diagnosis of paroxysmal nocturnal haemoglobinuria (PNH) by flow cytometry is based on mandatory analysis of the erythroid, neutrophilic and monocytic lineages. In this study, we have evaluated the performance characteristics of a recently published immature red blood cell (iRBC) assay as a potential screening test for PNH by flow cytometry. METHODS: Intra- and inter-assay imprecision were determined in five replicates of small, medium and large PNH iRBC clones. Analytical and functional sensitivity was assessed by performing spiking tests for five replicates. Thirty healthy donors and 441 PNH patients were tested for evaluation of clinical specificity, sensitivity, positive and negative predictive values. RESULTS: Coefficients of variation (CV) for intra-/inter-assay imprecision analyses were 1.31/1.50, 3.19/2.61 and 3.99/1.58 for the big, medium and small clone sizes, respectively. Absolute values (100%) were found for both clinical specificity and sensitivity as well as for both positive and negative predictive values. The CV from 5 replicate results for 10 clustered events was 15.7%. The coefficient of determination (r2 ), Pearson's correlation coefficient (r) and Bland-Altman mean bias were 0.9436/0.9234/1.7 for PNH iRBC compared to PNH neutrophils and 0.9553/0.9387/2.1 for PNH iRBCs compared to PNH monocytes. CONCLUSION: Our results confirm very good performance characteristics, high analytical and functional sensitivity, absolute clinical specificity and sensitivity as well as favourable correlation between PNH iRBCs and both PNH neutrophils and monocytes, suggesting that this cost-effective 3-colour iRBC assay can be used as a reliable screening test for evaluation of small, medium and large PNH clones by flow cytometry.
- 650 _2
- $a buněčné klony $7 D002999
- 650 _2
- $a barva $7 D003116
- 650 _2
- $a erytrocyty $7 D004912
- 650 _2
- $a průtoková cytometrie $x metody $7 D005434
- 650 12
- $a paroxysmální hemoglobinurie $x diagnóza $7 D006457
- 650 _2
- $a lidé $7 D006801
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Richards, Stephen J $u Division of Haematology and Immunology, Leeds Institute of Medical Research at St James's, University of Leeds, Leeds, UK
- 700 1_
- $a Pešek, Adam $u Clinical Department, Institute of Haematology and Blood Transfusion, Prague, Czech Republic
- 700 1_
- $a Illingworth, Andrea J $u Dahl-Chase Diagnostic Services, Bangor, Maine, USA $1 https://orcid.org/https://orcid.org/0000000302373591
- 700 1_
- $a Sutherland, D Robert $u Department of Laboratory Medicine, Toronto General Hospital, Toronto, Ontario, Canada
- 773 0_
- $w MED00153487 $t International journal of laboratory hematology $x 1751-553X $g Roč. 44, č. 5 (2022), s. 868-874
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/35614534 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20221017 $b ABA008
- 991 __
- $a 20221031100236 $b ABA008
- 999 __
- $a ok $b bmc $g 1854162 $s 1175570
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2022 $b 44 $c 5 $d 868-874 $e 20220525 $i 1751-553X $m International journal of laboratory hematology $n Int J Lab Hematol $x MED00153487
- LZP __
- $a Pubmed-20221017