Capture threshold of bipolar and unipolar pacing of left ventricle via coronary sinus branch: longitudinal study
J. Stritecky, J. Kremlacek, J. Hanus, L. Haman, H. Stritecka, J. Simka, P. Parizek
Status not-indexed Language English Country Switzerland
Document type Journal Article
NLK
Directory of Open Access Journals
from 2014
PubMed Central
from 2014
Europe PubMed Central
from 2014
Open Access Digital Library
from 2014-01-01
Open Access Digital Library
from 2014-01-01
ROAD: Directory of Open Access Scholarly Resources
from 2014
- Publication type
- Journal Article MeSH
INTRODUCTION: The aim of this paper is to first monitor the changes in the capture threshold of endovascularly placed leads for left ventricle pacing, second to compare the pacing configurations, and third to verify the effect of Steroid elution for endovascular leads. SAMPLE AND METHOD: The study included 202 consecutive single centre patients implanted with the QuartetTM lead (St. Jude Medical). The capture threshold and related lead parameters were tested during implantation, on the day of the patient's discharge, and 3, 9, and 15 months after implantation. The electrical energy corresponding to the threshold values for inducing ventricular contraction was recorded for subgroups of patients with bipolar and pseudo-unipolar pacing vectors and electrodes equipped with and without a slow-eluting steroids. The best setting for the resynchronization effect was generally chosen. Capture threshold was taken as a selection criterion only if there were multiple options with (expected) similar resynchronization effect. RESULTS AND DISCUSSION: The measurements showed that the ratio of threshold energies of UNI vs. BI was 5× higher (p < 0.001) at implantation. At the end of the follow-up, it dropped to 2.6 (p = 0.012). The steroid effect in BI vectors was caused by a double capture threshold in the NSE group compared to the SE group (p < 0.001), increased by approximately 2.5 times (p < 0.001). The study concludes that after a larger initial increase in the capture threshold, the leads showed a gradual increase in the entire set. As a result, the bipolar threshold energies increase, and the pseudo-unipolar energies decrease. Since bipolar vectors require a significantly lower pacing energy, battery life of the implanted device would improve. When evaluating the steroid elution of bipolar vectors, we observe a significant positive effect of a gradual increase of the threshold energy.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc23009717
- 003
- CZ-PrNML
- 005
- 20230721095502.0
- 007
- ta
- 008
- 230707s2023 sz f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.3389/fcvm.2023.1096538 $2 doi
- 035 __
- $a (PubMed)37288262
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a sz
- 100 1_
- $a Stritecky, Jakub $u Department of Medical Biophysics, Faculty of Medicine in Hradec Kralove, Charles University, Hradec Králové, Czech Republic $u 1st Department of Internal Medicine - Cardioangiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic
- 245 10
- $a Capture threshold of bipolar and unipolar pacing of left ventricle via coronary sinus branch: longitudinal study / $c J. Stritecky, J. Kremlacek, J. Hanus, L. Haman, H. Stritecka, J. Simka, P. Parizek
- 520 9_
- $a INTRODUCTION: The aim of this paper is to first monitor the changes in the capture threshold of endovascularly placed leads for left ventricle pacing, second to compare the pacing configurations, and third to verify the effect of Steroid elution for endovascular leads. SAMPLE AND METHOD: The study included 202 consecutive single centre patients implanted with the QuartetTM lead (St. Jude Medical). The capture threshold and related lead parameters were tested during implantation, on the day of the patient's discharge, and 3, 9, and 15 months after implantation. The electrical energy corresponding to the threshold values for inducing ventricular contraction was recorded for subgroups of patients with bipolar and pseudo-unipolar pacing vectors and electrodes equipped with and without a slow-eluting steroids. The best setting for the resynchronization effect was generally chosen. Capture threshold was taken as a selection criterion only if there were multiple options with (expected) similar resynchronization effect. RESULTS AND DISCUSSION: The measurements showed that the ratio of threshold energies of UNI vs. BI was 5× higher (p < 0.001) at implantation. At the end of the follow-up, it dropped to 2.6 (p = 0.012). The steroid effect in BI vectors was caused by a double capture threshold in the NSE group compared to the SE group (p < 0.001), increased by approximately 2.5 times (p < 0.001). The study concludes that after a larger initial increase in the capture threshold, the leads showed a gradual increase in the entire set. As a result, the bipolar threshold energies increase, and the pseudo-unipolar energies decrease. Since bipolar vectors require a significantly lower pacing energy, battery life of the implanted device would improve. When evaluating the steroid elution of bipolar vectors, we observe a significant positive effect of a gradual increase of the threshold energy.
- 590 __
- $a NEINDEXOVÁNO
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Kremlacek, Jan $u Department of Medical Biophysics, Faculty of Medicine in Hradec Kralove, Charles University, Hradec Králové, Czech Republic
- 700 1_
- $a Hanus, Josef $u Department of Medical Biophysics, Faculty of Medicine in Hradec Kralove, Charles University, Hradec Králové, Czech Republic
- 700 1_
- $a Haman, Ludek $u 1st Department of Internal Medicine - Cardioangiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic
- 700 1_
- $a Stritecka, Hana $u Department of Military Internal Medicine and Military Hygiene, Faculty of Health Sciences, University of Defence, Hradec Králové, Czech Republic
- 700 1_
- $a Simka, Jakub $u 1st Department of Internal Medicine - Cardioangiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic
- 700 1_
- $a Parizek, Petr $u 1st Department of Internal Medicine - Cardioangiology, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic
- 773 0_
- $w MED00198704 $t Frontiers in cardiovascular medicine $x 2297-055X $g Roč. 10, č. - (2023), s. 1096538
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/37288262 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20230707 $b ABA008
- 991 __
- $a 20230721095455 $b ABA008
- 999 __
- $a ok $b bmc $g 1958476 $s 1195981
- BAS __
- $a 3
- BAS __
- $a PreBMC-PubMed-not-MEDLINE
- BMC __
- $a 2023 $b 10 $c - $d 1096538 $e 20230523 $i 2297-055X $m Frontiers in cardiovascular medicine $n Front Cardiovasc Med $x MED00198704
- LZP __
- $a Pubmed-20230707