Detail
Článek
Článek online
FT
Medvik - BMČ
  • Je něco špatně v tomto záznamu ?

Preclinical assessment of the feasibility, safety and lesion durability of a novel 'single-shot' pulsed field ablation catheter for pulmonary vein isolation

J. Koruth, I. Kawamura, SR. Dukkipati, P. Neuzil, VY. Reddy

. 2023 ; 25 (4) : 1369-1378. [pub] 2023Apr15

Jazyk angličtina Země Anglie, Velká Británie

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc23011772

Grantová podpora
Affera

AIMS: Single-shot pulmonary vein isolation can improve procedural efficiency. To assess the capability of a novel, expandable lattice-shaped catheter to rapidly isolate thoracic veins using pulsed field ablation (PFA) in healthy swine. METHODS AND RESULTS: The study catheter (SpherePVI; Affera Inc) was used to isolate thoracic veins in two cohorts of swine survived for 1 and 5 weeks. In Experiment 1, an initial dose (PULSE2) was used to isolate the superior vena cava (SVC) and the right superior pulmonary vein (RSPV) in six swine and the SVC only in two swine. In Experiment 2, a final dose (PULSE3) was used for SVC, RSPV, and left superior pulmonary vein (LSPV) in five swine. Baseline and follow-up maps, ostial diameters, and phrenic nerve were assessed. Pulsed field ablation was delivered atop the oesophagus in three swine. All tissues were submitted for pathology. In Experiment 1, all 14/14 veins were isolated acutely with durable isolation demonstrated in 6/6 RSPVs and 6/8 SVC. Both reconnections occurred when only one application/vein was used. Fifty-two and 32 sections from the RSPVs and SVC revealed transmural lesions in 100% with a mean depth of 4.0 ± 2.0 mm. In Experiment 2, 15/15 veins were isolated acutely with 14/15 veins (5/5 SVC, 5/5 RSPV, and 4/5 LSPV) durably isolated. Right superior pulmonary vein (31) and SVC (34) sections had 100% transmural, circumferential ablation with minimal inflammation. Viable vessels and nerves were noted without evidence of venous stenosis, phrenic palsy, or oesophageal injury. CONCLUSION: This novel expandable lattice PFA catheter can achieve durable isolation with transmurality and safety.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc23011772
003      
CZ-PrNML
005      
20230801133329.0
007      
ta
008      
230718s2023 enk f 000 0|eng||
009      
AR
024    7_
$a 10.1093/europace/euad030 $2 doi
035    __
$a (PubMed)36794699
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a enk
100    1_
$a Koruth, Jacob $u Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, One Gustave L. Levy Place, PO Box 1030, New York, NY 10029, USA $1 https://orcid.org/0000000222218372
245    10
$a Preclinical assessment of the feasibility, safety and lesion durability of a novel 'single-shot' pulsed field ablation catheter for pulmonary vein isolation / $c J. Koruth, I. Kawamura, SR. Dukkipati, P. Neuzil, VY. Reddy
520    9_
$a AIMS: Single-shot pulmonary vein isolation can improve procedural efficiency. To assess the capability of a novel, expandable lattice-shaped catheter to rapidly isolate thoracic veins using pulsed field ablation (PFA) in healthy swine. METHODS AND RESULTS: The study catheter (SpherePVI; Affera Inc) was used to isolate thoracic veins in two cohorts of swine survived for 1 and 5 weeks. In Experiment 1, an initial dose (PULSE2) was used to isolate the superior vena cava (SVC) and the right superior pulmonary vein (RSPV) in six swine and the SVC only in two swine. In Experiment 2, a final dose (PULSE3) was used for SVC, RSPV, and left superior pulmonary vein (LSPV) in five swine. Baseline and follow-up maps, ostial diameters, and phrenic nerve were assessed. Pulsed field ablation was delivered atop the oesophagus in three swine. All tissues were submitted for pathology. In Experiment 1, all 14/14 veins were isolated acutely with durable isolation demonstrated in 6/6 RSPVs and 6/8 SVC. Both reconnections occurred when only one application/vein was used. Fifty-two and 32 sections from the RSPVs and SVC revealed transmural lesions in 100% with a mean depth of 4.0 ± 2.0 mm. In Experiment 2, 15/15 veins were isolated acutely with 14/15 veins (5/5 SVC, 5/5 RSPV, and 4/5 LSPV) durably isolated. Right superior pulmonary vein (31) and SVC (34) sections had 100% transmural, circumferential ablation with minimal inflammation. Viable vessels and nerves were noted without evidence of venous stenosis, phrenic palsy, or oesophageal injury. CONCLUSION: This novel expandable lattice PFA catheter can achieve durable isolation with transmurality and safety.
650    _2
$a prasata $7 D013552
650    _2
$a zvířata $7 D000818
650    12
$a fibrilace síní $x diagnóza $x chirurgie $x patologie $7 D001281
650    12
$a venae pulmonales $x chirurgie $x patologie $7 D011667
650    _2
$a vena cava superior $x chirurgie $7 D014683
650    _2
$a studie proveditelnosti $7 D005240
650    12
$a katetrizační ablace $x škodlivé účinky $x metody $7 D017115
650    _2
$a katétry $7 D057785
650    _2
$a výsledek terapie $7 D016896
655    _2
$a časopisecké články $7 D016428
700    1_
$a Kawamura, Iwanari $u Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, One Gustave L. Levy Place, PO Box 1030, New York, NY 10029, USA $1 https://orcid.org/0000000264319278
700    1_
$a Dukkipati, Srinivas R $u Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, One Gustave L. Levy Place, PO Box 1030, New York, NY 10029, USA
700    1_
$a Neuzil, Petr $u Homolka Hospital, Prague, Czech Republic $1 https://orcid.org/0000000343348165
700    1_
$a Reddy, Vivek Y $u Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, One Gustave L. Levy Place, PO Box 1030, New York, NY 10029, USA $1 https://orcid.org/0000000256384993
773    0_
$w MED00149837 $t Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology $x 1532-2092 $g Roč. 25, č. 4 (2023), s. 1369-1378
856    41
$u https://pubmed.ncbi.nlm.nih.gov/36794699 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y p $z 0
990    __
$a 20230718 $b ABA008
991    __
$a 20230801133325 $b ABA008
999    __
$a ok $b bmc $g 1963925 $s 1198037
BAS    __
$a 3
BAS    __
$a PreBMC-MEDLINE
BMC    __
$a 2023 $b 25 $c 4 $d 1369-1378 $e 2023Apr15 $i 1532-2092 $m Europace $n Europace $x MED00149837
GRA    __
$p Affera
LZP    __
$a Pubmed-20230718

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...