Validation of the Sysmex XN analyser and Blood Bank mode for the quality and safety of donor blood and transfusion products
O. Valina, I. Vankova, K. Mintzas, K. Knappova, Z. Gasova
Jazyk angličtina Země Anglie, Velká Británie
Typ dokumentu časopisecké články
PubMed
38193379
DOI
10.1111/tme.13028
Knihovny.cz E-zdroje
- MeSH
- dárci krve * MeSH
- erytrocyty MeSH
- krevní banky * MeSH
- krevní obraz metody MeSH
- lidé MeSH
- počet trombocytů MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
OBJECTIVES: Our objective was to compare the measurement of residual white blood cell (rWBC) and residual red blood cell (rRBC) counts in blood products using the XN Blood Bank mode and the laboratory standard operating procedures for manual counts. In addition, to compare the whole blood complete blood count (CBC) values of blood donors and the quality of blood products using the Sysmex XN analyser versus the XS-1000i analyser. MATERIALS AND METHODS: For blood donors, 190 samples from blood or apheresis donors were analysed on both the Sysmex XS-1000i and XN-1000 analysers and the mean values of six CBC parameters were compared: the white blood cell count (WBC), the red blood cell count (RBC), haemoglobin (HGB), haematocrit (HCT), the mean corpuscular volume (MCV), the platelet count (PLT). For blood products, 164 samples were collected: 13 Plasma products - whole blood, 9 Plasma products - apheresis, 36 RBC concentrates - whole blood, 30 PLT concentrates - buffy coats, 36 PLT concentrates - buffy coats - pooled and 55 PLT concentrates - apheresis. RESULTS: All CBC parameters of the blood donors tested showed similar performance, with excellent correlation coefficients (r) ranging from 0.821 to 0.995. The majority of the blood products did not have a quantifiable number of residual cells, meaning the number of rWBC and rRBC, if present, was below the limit of quantitation (LoQ) of the different methods. rWBC were detected by Blood Bank mode in Plasma products - whole blood with a mean rWBC of 0.012 × 109 /L and in PLT concentrates - buffy coats with a mean rWBC of 0.19 × 109 /L. The correlation coefficient in both analysers for all three parameters (HGB, HCT, RBC) in RBC concentrates - whole blood was excellent, ranging from 0.95 to 0.99. For platelet count, r ranged from 0.98 to 0.99. CONCLUSION: The XN-Series analyser, equipped with a Blood Bank mode, demonstrated reliable performance when used for blood donor evaluation, rWBC enumeration and measurement of end blood products.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc24007302
- 003
- CZ-PrNML
- 005
- 20240423155849.0
- 007
- ta
- 008
- 240412s2024 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1111/tme.13028 $2 doi
- 035 __
- $a (PubMed)38193379
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Valina, Ondrej $u Haematology Department, Sysmex Europe SE, Norderstedt, Germany
- 245 10
- $a Validation of the Sysmex XN analyser and Blood Bank mode for the quality and safety of donor blood and transfusion products / $c O. Valina, I. Vankova, K. Mintzas, K. Knappova, Z. Gasova
- 520 9_
- $a OBJECTIVES: Our objective was to compare the measurement of residual white blood cell (rWBC) and residual red blood cell (rRBC) counts in blood products using the XN Blood Bank mode and the laboratory standard operating procedures for manual counts. In addition, to compare the whole blood complete blood count (CBC) values of blood donors and the quality of blood products using the Sysmex XN analyser versus the XS-1000i analyser. MATERIALS AND METHODS: For blood donors, 190 samples from blood or apheresis donors were analysed on both the Sysmex XS-1000i and XN-1000 analysers and the mean values of six CBC parameters were compared: the white blood cell count (WBC), the red blood cell count (RBC), haemoglobin (HGB), haematocrit (HCT), the mean corpuscular volume (MCV), the platelet count (PLT). For blood products, 164 samples were collected: 13 Plasma products - whole blood, 9 Plasma products - apheresis, 36 RBC concentrates - whole blood, 30 PLT concentrates - buffy coats, 36 PLT concentrates - buffy coats - pooled and 55 PLT concentrates - apheresis. RESULTS: All CBC parameters of the blood donors tested showed similar performance, with excellent correlation coefficients (r) ranging from 0.821 to 0.995. The majority of the blood products did not have a quantifiable number of residual cells, meaning the number of rWBC and rRBC, if present, was below the limit of quantitation (LoQ) of the different methods. rWBC were detected by Blood Bank mode in Plasma products - whole blood with a mean rWBC of 0.012 × 109 /L and in PLT concentrates - buffy coats with a mean rWBC of 0.19 × 109 /L. The correlation coefficient in both analysers for all three parameters (HGB, HCT, RBC) in RBC concentrates - whole blood was excellent, ranging from 0.95 to 0.99. For platelet count, r ranged from 0.98 to 0.99. CONCLUSION: The XN-Series analyser, equipped with a Blood Bank mode, demonstrated reliable performance when used for blood donor evaluation, rWBC enumeration and measurement of end blood products.
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a dárci krve $7 D001782
- 650 12
- $a krevní banky $7 D001771
- 650 _2
- $a krevní obraz $x metody $7 D001772
- 650 _2
- $a počet trombocytů $7 D010976
- 650 _2
- $a erytrocyty $7 D004912
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Vankova, Ida $u Apheresis and Transfusion Departments, Institute of Haematology and Blood Transfusion, Prague, Czech Republic
- 700 1_
- $a Mintzas, Konstantinos $u Haematology Department, Sysmex Europe SE, Norderstedt, Germany
- 700 1_
- $a Knappova, Klara $u Apheresis and Transfusion Departments, Institute of Haematology and Blood Transfusion, Prague, Czech Republic
- 700 1_
- $a Gasova, Zdenka $u Apheresis and Transfusion Departments, Institute of Haematology and Blood Transfusion, Prague, Czech Republic
- 773 0_
- $w MED00004548 $t Transfusion medicine $x 1365-3148 $g Roč. 34, č. 1 (2024), s. 30-38
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/38193379 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20240412 $b ABA008
- 991 __
- $a 20240423155845 $b ABA008
- 999 __
- $a ok $b bmc $g 2081343 $s 1217069
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2024 $b 34 $c 1 $d 30-38 $e 20240109 $i 1365-3148 $m Transfusion medicine $n Transfus Med $x MED00004548
- LZP __
- $a Pubmed-20240412