Combining Dipole and Loop Coil Elements for 7 T Magnetic Resonance Studies of the Human Calf Muscle
V. Cap, VR. Rocha Dos Santos, K. Repnin, D. Červený, E. Laistler, M. Meyerspeer, R. Frass-Kriegl
Jazyk angličtina Země Švýcarsko
Typ dokumentu časopisecké články
Grantová podpora
P 35305-B
FWF Austrian Science Fund
NLK
Directory of Open Access Journals
od 2001
PubMed Central
od 2003
Europe PubMed Central
od 2003
ProQuest Central
od 2001-01-01
Open Access Digital Library
od 2001-01-01
Open Access Digital Library
od 2003-01-01
Health & Medicine (ProQuest)
od 2001-01-01
ROAD: Directory of Open Access Scholarly Resources
od 2001
PubMed
38894105
DOI
10.3390/s24113309
Knihovny.cz E-zdroje
- MeSH
- fantomy radiodiagnostické * MeSH
- kosterní svaly * diagnostické zobrazování chemie MeSH
- kyselina mléčná chemie metabolismus MeSH
- lidé MeSH
- magnetická rezonanční spektroskopie metody MeSH
- magnetická rezonanční tomografie * metody MeSH
- poměr signál - šum MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
Combining proton and phosphorus magnetic resonance spectroscopy offers a unique opportunity to study the oxidative and glycolytic components of metabolism in working muscle. This paper presents a 7 T proton calf coil design that combines dipole and loop elements to achieve the high performance necessary for detecting metabolites with low abundance and restricted visibility, specifically lactate, while including the option of adding a phosphorus array. We investigated the transmit, receive, and parallel imaging performance of three transceiver dipoles with six pair-wise overlap-decoupled standard or twisted pair receive-only coils. With a higher SNR and more efficient transmission decoupling, standard loops outperformed twisted pair coils. The dipoles with standard loops provided a four-fold-higher image SNR than a multinuclear reference coil comprising two proton channels and 32% more than a commercially available 28-channel proton knee coil. The setup enabled up to three-fold acceleration in the right-left direction, with acceptable g-factors and no visible aliasing artefacts. Spectroscopic phantom measurements revealed a higher spectral SNR for lactate with the developed setup than with either reference coil and fewer restrictions in voxel placement due to improved transmit homogeneity. This paper presents a new use case for dipoles and highlights their advantages for the integration in multinuclear calf coils.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc24013882
- 003
- CZ-PrNML
- 005
- 20240905133928.0
- 007
- ta
- 008
- 240725s2024 sz f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.3390/s24113309 $2 doi
- 035 __
- $a (PubMed)38894105
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a sz
- 100 1_
- $a Cap, Veronika $u High Field MR Center, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, 1090 Vienna, Austria $1 https://orcid.org/0000000223214138
- 245 10
- $a Combining Dipole and Loop Coil Elements for 7 T Magnetic Resonance Studies of the Human Calf Muscle / $c V. Cap, VR. Rocha Dos Santos, K. Repnin, D. Červený, E. Laistler, M. Meyerspeer, R. Frass-Kriegl
- 520 9_
- $a Combining proton and phosphorus magnetic resonance spectroscopy offers a unique opportunity to study the oxidative and glycolytic components of metabolism in working muscle. This paper presents a 7 T proton calf coil design that combines dipole and loop elements to achieve the high performance necessary for detecting metabolites with low abundance and restricted visibility, specifically lactate, while including the option of adding a phosphorus array. We investigated the transmit, receive, and parallel imaging performance of three transceiver dipoles with six pair-wise overlap-decoupled standard or twisted pair receive-only coils. With a higher SNR and more efficient transmission decoupling, standard loops outperformed twisted pair coils. The dipoles with standard loops provided a four-fold-higher image SNR than a multinuclear reference coil comprising two proton channels and 32% more than a commercially available 28-channel proton knee coil. The setup enabled up to three-fold acceleration in the right-left direction, with acceptable g-factors and no visible aliasing artefacts. Spectroscopic phantom measurements revealed a higher spectral SNR for lactate with the developed setup than with either reference coil and fewer restrictions in voxel placement due to improved transmit homogeneity. This paper presents a new use case for dipoles and highlights their advantages for the integration in multinuclear calf coils.
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a kosterní svaly $x diagnostické zobrazování $x chemie $7 D018482
- 650 12
- $a fantomy radiodiagnostické $7 D019047
- 650 12
- $a magnetická rezonanční tomografie $x metody $7 D008279
- 650 _2
- $a magnetická rezonanční spektroskopie $x metody $7 D009682
- 650 _2
- $a poměr signál - šum $7 D059629
- 650 _2
- $a kyselina mléčná $x chemie $x metabolismus $7 D019344
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Rocha Dos Santos, Vasco Rafael $u High Field MR Center, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, 1090 Vienna, Austria
- 700 1_
- $a Repnin, Kostiantyn $u High Field MR Center, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, 1090 Vienna, Austria
- 700 1_
- $a Červený, David $u High Field MR Center, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, 1090 Vienna, Austria $u Institute for Clinical and Experimental Medicine, 140 21 Prague, Czech Republic $u Institute of Biophysics and Informatics, First Faculty of Medicine, Charles University, 121 08 Prague, Czech Republic $1 https://orcid.org/0000000258120510
- 700 1_
- $a Laistler, Elmar $u High Field MR Center, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, 1090 Vienna, Austria $1 https://orcid.org/0000000283448307
- 700 1_
- $a Meyerspeer, Martin $u High Field MR Center, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, 1090 Vienna, Austria
- 700 1_
- $a Frass-Kriegl, Roberta $u High Field MR Center, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, 1090 Vienna, Austria $1 https://orcid.org/0000000288404602
- 773 0_
- $w MED00008309 $t Sensors $x 1424-8220 $g Roč. 24, č. 11 (2024)
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/38894105 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20240725 $b ABA008
- 991 __
- $a 20240905133922 $b ABA008
- 999 __
- $a ok $b bmc $g 2143600 $s 1225748
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2024 $b 24 $c 11 $e 20240522 $i 1424-8220 $m Sensors $n Sensors (Basel) $x MED00008309
- GRA __
- $a P 35305-B $p FWF Austrian Science Fund
- LZP __
- $a Pubmed-20240725