Fetal Dose Evaluation for Pregnant Patients on Leksell Gamma Knife Perfexion/Icon
L. Hamáčková, J. Novotný, G. Šimonová, R. Liščák, T. Chytka
Language English Country Switzerland
Document type Case Reports
PubMed
38198760
DOI
10.1159/000535366
Knihovny.cz E-resources
- MeSH
- Phantoms, Imaging MeSH
- Humans MeSH
- Radiosurgery * methods MeSH
- Pregnancy MeSH
- Check Tag
- Humans MeSH
- Pregnancy MeSH
- Female MeSH
- Publication type
- Case Reports MeSH
INTRODUCTION: It is a normal procedure to avoid the application of ionizing radiation during pregnancy. In very rare occasions, treatment can be performed, but doses to the fetus must be evaluated and reported, and the patient must sign informed consent. There can occur two types of damage caused by ionizing radiation - deterministic and stochastic effects. Deterministic effects may occur after reaching a certain threshold (100 mGy for this study); meanwhile, stochastic effects have no limit and their probability rises with dose. This study focuses on deterministic effects. CASE PRESENTATIONS: This study compares the dose measured on phantom for the area of the pelvis and the dose measured on 3 patients with dosimeters positioned on the pelvis irradiated on Leksell Gamma Knife Perfexion/Icon. The mean dose for measurement on phantom for the pelvis was 0.73 ± 0.76 mGy, and for the patients, it was 1.28 mGy, 0.493 mGy, and 0.549 mGy which is 80 times lower, 200 times lower, and 180 times lower than the threshold for deterministic effects, respectively. CONCLUSION: The measurement carried on phantom served as the base for drafting informed consent and provided initial proof that treatment can be safely delivered. Measurements performed on patients only confirmed that irradiation of pregnant patients on Leksell Gamma Knife Perfexion/Icon is safe relative to the deterministic effects. Nevertheless, pregnant patients should be treated with ionizing radiation only in very extraordinary situations.
Department of Medical Physics Na Homolce Hospital Prague Czechia
Department of Stereotactic and Radiation Neurosurgery Prague Czechia
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc24014601
- 003
- CZ-PrNML
- 005
- 20240905133549.0
- 007
- ta
- 008
- 240725s2024 sz f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1159/000535366 $2 doi
- 035 __
- $a (PubMed)38198760
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a sz
- 100 1_
- $a Hamáčková, Lucie $u Department of Medical Physics, Na Homolce Hospital, Prague, Czechia
- 245 10
- $a Fetal Dose Evaluation for Pregnant Patients on Leksell Gamma Knife Perfexion/Icon / $c L. Hamáčková, J. Novotný, G. Šimonová, R. Liščák, T. Chytka
- 520 9_
- $a INTRODUCTION: It is a normal procedure to avoid the application of ionizing radiation during pregnancy. In very rare occasions, treatment can be performed, but doses to the fetus must be evaluated and reported, and the patient must sign informed consent. There can occur two types of damage caused by ionizing radiation - deterministic and stochastic effects. Deterministic effects may occur after reaching a certain threshold (100 mGy for this study); meanwhile, stochastic effects have no limit and their probability rises with dose. This study focuses on deterministic effects. CASE PRESENTATIONS: This study compares the dose measured on phantom for the area of the pelvis and the dose measured on 3 patients with dosimeters positioned on the pelvis irradiated on Leksell Gamma Knife Perfexion/Icon. The mean dose for measurement on phantom for the pelvis was 0.73 ± 0.76 mGy, and for the patients, it was 1.28 mGy, 0.493 mGy, and 0.549 mGy which is 80 times lower, 200 times lower, and 180 times lower than the threshold for deterministic effects, respectively. CONCLUSION: The measurement carried on phantom served as the base for drafting informed consent and provided initial proof that treatment can be safely delivered. Measurements performed on patients only confirmed that irradiation of pregnant patients on Leksell Gamma Knife Perfexion/Icon is safe relative to the deterministic effects. Nevertheless, pregnant patients should be treated with ionizing radiation only in very extraordinary situations.
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a těhotenství $7 D011247
- 650 _2
- $a ženské pohlaví $7 D005260
- 650 12
- $a radiochirurgie $x metody $7 D016634
- 650 _2
- $a fantomy radiodiagnostické $7 D019047
- 655 _2
- $a kazuistiky $7 D002363
- 700 1_
- $a Novotný, Josef $u Department of Medical Physics, Na Homolce Hospital, Prague, Czechia
- 700 1_
- $a Šimonová, Gabriela $u Department of Stereotactic and Radiation Neurosurgery, Prague, Czechia
- 700 1_
- $a Liščák, Roman $u Department of Stereotactic and Radiation Neurosurgery, Prague, Czechia
- 700 1_
- $a Chytka, Tomáš $u Department of Stereotactic and Radiation Neurosurgery, Prague, Czechia
- 773 0_
- $w MED00004437 $t Stereotactic and functional neurosurgery $x 1423-0372 $g Roč. 102, č. 2 (2024), s. 65-73
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/38198760 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20240725 $b ABA008
- 991 __
- $a 20240905133543 $b ABA008
- 999 __
- $a ok $b bmc $g 2144020 $s 1226467
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2024 $b 102 $c 2 $d 65-73 $e 20240110 $i 1423-0372 $m Stereotactic and functional neurosurgery $n Stereotact Funct Neurosurg $x MED00004437
- LZP __
- $a Pubmed-20240725