Fetal Dose Evaluation for Pregnant Patients on Leksell Gamma Knife Perfexion/Icon

L. Hamáčková, J. Novotný, G. Šimonová, R. Liščák, T. Chytka

. 2024 ; 102 (2) : 65-73. [pub] 20240110

Language English Country Switzerland

Document type Case Reports

INTRODUCTION: It is a normal procedure to avoid the application of ionizing radiation during pregnancy. In very rare occasions, treatment can be performed, but doses to the fetus must be evaluated and reported, and the patient must sign informed consent. There can occur two types of damage caused by ionizing radiation - deterministic and stochastic effects. Deterministic effects may occur after reaching a certain threshold (100 mGy for this study); meanwhile, stochastic effects have no limit and their probability rises with dose. This study focuses on deterministic effects. CASE PRESENTATIONS: This study compares the dose measured on phantom for the area of the pelvis and the dose measured on 3 patients with dosimeters positioned on the pelvis irradiated on Leksell Gamma Knife Perfexion/Icon. The mean dose for measurement on phantom for the pelvis was 0.73 ± 0.76 mGy, and for the patients, it was 1.28 mGy, 0.493 mGy, and 0.549 mGy which is 80 times lower, 200 times lower, and 180 times lower than the threshold for deterministic effects, respectively. CONCLUSION: The measurement carried on phantom served as the base for drafting informed consent and provided initial proof that treatment can be safely delivered. Measurements performed on patients only confirmed that irradiation of pregnant patients on Leksell Gamma Knife Perfexion/Icon is safe relative to the deterministic effects. Nevertheless, pregnant patients should be treated with ionizing radiation only in very extraordinary situations.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc24014601
003      
CZ-PrNML
005      
20240905133549.0
007      
ta
008      
240725s2024 sz f 000 0|eng||
009      
AR
024    7_
$a 10.1159/000535366 $2 doi
035    __
$a (PubMed)38198760
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a sz
100    1_
$a Hamáčková, Lucie $u Department of Medical Physics, Na Homolce Hospital, Prague, Czechia
245    10
$a Fetal Dose Evaluation for Pregnant Patients on Leksell Gamma Knife Perfexion/Icon / $c L. Hamáčková, J. Novotný, G. Šimonová, R. Liščák, T. Chytka
520    9_
$a INTRODUCTION: It is a normal procedure to avoid the application of ionizing radiation during pregnancy. In very rare occasions, treatment can be performed, but doses to the fetus must be evaluated and reported, and the patient must sign informed consent. There can occur two types of damage caused by ionizing radiation - deterministic and stochastic effects. Deterministic effects may occur after reaching a certain threshold (100 mGy for this study); meanwhile, stochastic effects have no limit and their probability rises with dose. This study focuses on deterministic effects. CASE PRESENTATIONS: This study compares the dose measured on phantom for the area of the pelvis and the dose measured on 3 patients with dosimeters positioned on the pelvis irradiated on Leksell Gamma Knife Perfexion/Icon. The mean dose for measurement on phantom for the pelvis was 0.73 ± 0.76 mGy, and for the patients, it was 1.28 mGy, 0.493 mGy, and 0.549 mGy which is 80 times lower, 200 times lower, and 180 times lower than the threshold for deterministic effects, respectively. CONCLUSION: The measurement carried on phantom served as the base for drafting informed consent and provided initial proof that treatment can be safely delivered. Measurements performed on patients only confirmed that irradiation of pregnant patients on Leksell Gamma Knife Perfexion/Icon is safe relative to the deterministic effects. Nevertheless, pregnant patients should be treated with ionizing radiation only in very extraordinary situations.
650    _2
$a lidé $7 D006801
650    _2
$a těhotenství $7 D011247
650    _2
$a ženské pohlaví $7 D005260
650    12
$a radiochirurgie $x metody $7 D016634
650    _2
$a fantomy radiodiagnostické $7 D019047
655    _2
$a kazuistiky $7 D002363
700    1_
$a Novotný, Josef $u Department of Medical Physics, Na Homolce Hospital, Prague, Czechia
700    1_
$a Šimonová, Gabriela $u Department of Stereotactic and Radiation Neurosurgery, Prague, Czechia
700    1_
$a Liščák, Roman $u Department of Stereotactic and Radiation Neurosurgery, Prague, Czechia
700    1_
$a Chytka, Tomáš $u Department of Stereotactic and Radiation Neurosurgery, Prague, Czechia
773    0_
$w MED00004437 $t Stereotactic and functional neurosurgery $x 1423-0372 $g Roč. 102, č. 2 (2024), s. 65-73
856    41
$u https://pubmed.ncbi.nlm.nih.gov/38198760 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y - $z 0
990    __
$a 20240725 $b ABA008
991    __
$a 20240905133543 $b ABA008
999    __
$a ok $b bmc $g 2144020 $s 1226467
BAS    __
$a 3
BAS    __
$a PreBMC-MEDLINE
BMC    __
$a 2024 $b 102 $c 2 $d 65-73 $e 20240110 $i 1423-0372 $m Stereotactic and functional neurosurgery $n Stereotact Funct Neurosurg $x MED00004437
LZP    __
$a Pubmed-20240725

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...