Assessing the Test-Retest Reliability of MyotonPRO for Measuring Achilles Tendon Stiffness
K. Volesky, J. Novak, M. Janek, J. Katolicky, JJ. Tufano, M. Steffl, J. Courel-Ibáñez, T. Vetrovsky
Status neindexováno Jazyk angličtina Země Švýcarsko
Typ dokumentu časopisecké články
Grantová podpora
94622
Grant Agency of Charles University
NLK
Directory of Open Access Journals
od 2016
PubMed Central
od 2018
ProQuest Central
od 2021-01-01
Nursing & Allied Health Database (ProQuest)
od 2021-01-01
Health & Medicine (ProQuest)
od 2021-01-01
ROAD: Directory of Open Access Scholarly Resources
od 2016
PubMed
40137335
DOI
10.3390/jfmk10010083
Knihovny.cz E-zdroje
- Publikační typ
- časopisecké články MeSH
Objectives: This study evaluates the test-retest reliability and inter-rater reliability of the MyotonPRO for measuring Achilles tendon stiffness at two standardized sites over various time frames and settings. Methods: Eight healthy participants underwent assessments by three raters over six visits. Tendon stiffness was measured at proximal (mid-portion) and distal (insertional) regions of the Achilles tendon at various time frames (10-15 s, 10-15 min, 24 h, and 14 days apart). Measurements included participant repositioning and two activity stimuli (daily living and sport). Reliability was calculated using the intraclass correlation coefficient (ICC), its 95% confidence interval, coefficient of variation, standard error of measurement, and minimal detectable change. Results: Short-term reliability (10-15 min) was excellent, with an ICC of 0.956 (0.929-0.974). Between days reliability (24 h) was good, with an ICC of 0.889 (0.802-0.938). Between weeks reliability (2 weeks) was good with an ICC of 0.886 (0.811-0.931). Short-term reliability with the simulation of activity of daily living was good, with an ICC of 0.917 (0.875-0.945). Short-term reliability with the simulation of sport was good with an ICC of 0.933 (0.891-0.96). Between days reliability with the simulation of sport was good, with an ICC of 0.920 (0.859-0.955). Conclusions: When used in a standardized position, the MyotonPRO demonstrates reliable repeated measurements of Achilles tendon stiffness. This protocol provides a foundation for clinical research and rehabilitation by clarifying expected reliability across minutes, days, and weeks, thus aiding clinicians and researchers in monitoring tendon adaptations and making evidence-based decisions.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc25008327
- 003
- CZ-PrNML
- 005
- 20250422095807.0
- 007
- ta
- 008
- 250408s2025 sz f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.3390/jfmk10010083 $2 doi
- 035 __
- $a (PubMed)40137335
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a sz
- 100 1_
- $a Volesky, Krystof $u Faculty of Physical Education and Sport, Charles University, 162 52 Prague, Czech Republic $1 https://orcid.org/0000000305425430
- 245 10
- $a Assessing the Test-Retest Reliability of MyotonPRO for Measuring Achilles Tendon Stiffness / $c K. Volesky, J. Novak, M. Janek, J. Katolicky, JJ. Tufano, M. Steffl, J. Courel-Ibáñez, T. Vetrovsky
- 520 9_
- $a Objectives: This study evaluates the test-retest reliability and inter-rater reliability of the MyotonPRO for measuring Achilles tendon stiffness at two standardized sites over various time frames and settings. Methods: Eight healthy participants underwent assessments by three raters over six visits. Tendon stiffness was measured at proximal (mid-portion) and distal (insertional) regions of the Achilles tendon at various time frames (10-15 s, 10-15 min, 24 h, and 14 days apart). Measurements included participant repositioning and two activity stimuli (daily living and sport). Reliability was calculated using the intraclass correlation coefficient (ICC), its 95% confidence interval, coefficient of variation, standard error of measurement, and minimal detectable change. Results: Short-term reliability (10-15 min) was excellent, with an ICC of 0.956 (0.929-0.974). Between days reliability (24 h) was good, with an ICC of 0.889 (0.802-0.938). Between weeks reliability (2 weeks) was good with an ICC of 0.886 (0.811-0.931). Short-term reliability with the simulation of activity of daily living was good, with an ICC of 0.917 (0.875-0.945). Short-term reliability with the simulation of sport was good with an ICC of 0.933 (0.891-0.96). Between days reliability with the simulation of sport was good, with an ICC of 0.920 (0.859-0.955). Conclusions: When used in a standardized position, the MyotonPRO demonstrates reliable repeated measurements of Achilles tendon stiffness. This protocol provides a foundation for clinical research and rehabilitation by clarifying expected reliability across minutes, days, and weeks, thus aiding clinicians and researchers in monitoring tendon adaptations and making evidence-based decisions.
- 590 __
- $a NEINDEXOVÁNO
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Novak, Jan $u Faculty of Physical Education and Sport, Charles University, 162 52 Prague, Czech Republic
- 700 1_
- $a Janek, Michael $u Faculty of Physical Education and Sport, Charles University, 162 52 Prague, Czech Republic
- 700 1_
- $a Katolicky, Jakub $u Department of Rehabilitation and Sports Medicine, Second Faculty of Medicine, University Hospital in Motol, Charles University, 150 06 Prague, Czech Republic $1 https://orcid.org/0000000185291975
- 700 1_
- $a Tufano, James J $u Faculty of Physical Education and Sport, Charles University, 162 52 Prague, Czech Republic
- 700 1_
- $a Steffl, Michal $u Faculty of Physical Education and Sport, Charles University, 162 52 Prague, Czech Republic $1 https://orcid.org/0000000172978145
- 700 1_
- $a Courel-Ibáñez, Javier $u Department of Physical Education and Sport, Faculty of Sport Sciences, University of Granada, 520 71 Melilla, Spain $1 https://orcid.org/0000000324461875
- 700 1_
- $a Vetrovsky, Tomas $u Faculty of Physical Education and Sport, Charles University, 162 52 Prague, Czech Republic $1 https://orcid.org/0000000325297069 $7 xx0205902
- 773 0_
- $w MED00209369 $t Journal of functional morphology and kinesiology $x 2411-5142 $g Roč. 10, č. 1 (2025)
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/40137335 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20250408 $b ABA008
- 991 __
- $a 20250422095808 $b ABA008
- 999 __
- $a ok $b bmc $g 2306341 $s 1245402
- BAS __
- $a 3
- BAS __
- $a PreBMC-PubMed-not-MEDLINE
- BMC __
- $a 2025 $b 10 $c 1 $e 20250228 $i 2411-5142 $m Journal of functional morphology and kinesiology $n J Funct Morphol Kinesiol $x MED00209369
- GRA __
- $a 94622 $p Grant Agency of Charles University
- LZP __
- $a Pubmed-20250408